Continuous glucose monitoring in interstitial fluid using glucose oxidase-based sensor compared to established blood glucose measurement in rats

被引:27
作者
Woderer, Stefanie
Henninger, Nadja
Garthe, Claus-Dieter
Kloetzer, Hans Martin
Hajnsek, Martin
Kamecke, Ulrike
Gretz, Norbert
Kraenzlin, Bettina
Pill, Johannes
机构
[1] Roche Diagnost GmbH, D-68305 Mannheim, Germany
[2] Heidelberg Univ, Univ Hosp Mannheim, Med Res Ctr, D-68167 Mannheim, Germany
[3] Roche Diagnost GmbH, A-8020 Graz, Austria
[4] Metecon GmbH, D-68165 Mannheim, Germany
关键词
glucose biosensor; continuous glucose monitoring; glucose oxidase; glucose in interstitial fluid; blood glucose; rat;
D O I
10.1016/j.aca.2006.08.016
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Glucose monitoring is of importance for success of complex therapeutic interventions in diabetic patients. Its impact on treatment and glycemic control is demonstrated in large clinical trials. Up to eight blood glucose measurements per day are recommended. Notwithstanding, a substantial number of diabetic patients cannot or will not monitor their blood glucose appropriately. Considerable progress in control of disturbed metabolism in diabetic patients can be expected by continuous glucose monitoring. The aim of the study was to evaluate the performance of a new amperometric glucose oxidase-based glucose sensor in vitro and in vivo after subcutaneous implantation into rats. For in vitro testing current output of sensors was measured by exposure to increasing and decreasing glucose concentrations up to 472 mg dL(-1) over a time period of 7 days. After subcutaneous implantation of sensors into interscapular region of male rats glucose in interstitial fluid was evaluated and compared to glucose in arterial blood up to 7 days. Hyper- and hypoglycaemia were induced by intravenous application of glucose and insulin, respectively. Current of each implanted sensor was converted into glucose concentration using the first blood glucose measurement only. A change of current with glucose of 0.35 nA mg(-1) dL(-1) indicates high sensitivity of the sensor in vitro. The response time (90% of steady state) was calculated by approximately 60 s. Test strips for blood glucose measurement as reference for sensor readings was found as an appropriate and rapidly available method in rats by comparison with established hexokinase method in an automated lab analyzer with limits of agreement of +32.8 and -25.7 mg dL(-1) in Bland-Altman analysis. In normo- and hypoglycaemic range sensor readings in interstitial fluid correlated well with blood glucose measurements whereas hyperglycaemia was not reflected by the sensor completely when blood glucose was changing rapidly. The data given characterize a sensor with high sensitivity, long term stability and short response time. A single calibration of the sensor is required only in measurement periods up to 7 days. The findings demonstrate that the sensor is a highly promising candidate for assessment in humans. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:7 / 12
页数:6
相关论文
共 50 条
  • [41] Needle-shaped glucose sensor based on polypyrrole doped with glucose oxidase
    Zhang, Bao Li
    Jin, Xuan
    Sun, Long Hua
    Guo, Xin Dong
    MICROCHEMICAL JOURNAL, 2020, 158
  • [42] Glucose oxidase immobilized macro porous silicon based conductive glucose sensor
    Sarkar, Tanusree
    Mukherjee, Nandini
    Das, Jayoti
    APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING, 2022, 128 (04):
  • [43] Comparison of Glucose Area Under the Curve Measured Using Minimally Invasive Interstitial Fluid Extraction Technology with Continuous Glucose Monitoring System in Diabetic Patients
    Uemura, Mei
    Yano, Yutaka
    Suzuki, Toshinari
    Yasuma, Taro
    Sato, Toshiyuki
    Morimoto, Aya
    Hosoya, Samiko
    Suminaka, Chihiro
    Nakajima, Hiromu
    Gabazza, Esteban C.
    Takei, Yoshiyuki
    DIABETES & METABOLISM JOURNAL, 2017, 41 (04) : 265 - 274
  • [44] Evaluation of blood glucose fluctuation in Japanese patients with type 1 diabetes mellitus by self-monitoring of blood glucose and continuous glucose monitoring
    Kusunoki, Yoshiki
    Katsuno, Tomoyuki
    Nakae, Rie
    Watanabe, Kahori
    Akagami, Takafumi
    Ochi, Fumihiro
    Tokuda, Masaru
    Murai, Kazuki
    Miuchi, Masayuki
    Miyagawa, Jun-ichiro
    Namba, Mitsuyoshi
    DIABETES RESEARCH AND CLINICAL PRACTICE, 2015, 108 (02) : 342 - 349
  • [45] A long lifetime and highly sensitive wearable microneedle sensor for the continuous real-time monitoring of glucose in interstitial fluid
    Yin, Sijie
    Yu, Zhiqiang
    Song, Ningning
    Guo, Zhanjun
    Li, Wei
    Ma, Jinhang
    Wang, Xiaoyan
    Liu, Junyao
    Liang, Minmin
    BIOSENSORS & BIOELECTRONICS, 2024, 244
  • [46] Correspondence of continuous interstitial glucose measurement against arterialised and capillary glucose following an oral glucose tolerance test in healthy volunteers
    Dye, Louise
    Mansfield, Michael
    Lasikiewicz, Nicola
    Mahawish, Lena
    Schnell, Rainer
    Talbot, Duncan
    Chauhan, Hitesh
    Croden, Fiona
    Lawton, Clare
    BRITISH JOURNAL OF NUTRITION, 2010, 103 (01) : 134 - 140
  • [47] Miniaturization of an Osmotic Pressure-Based Glucose Sensor for Continuous Intraperitoneal and Subcutaneous Glucose Monitoring by Means of Nanotechnology
    Pfuetzner, Andreas
    Tencer, Barbora
    Stamm, Boris
    Mehta, Mandar
    Sharma, Preeti
    Gilyazev, Rustam
    Jensch, Hendrick
    Thome, Nicole
    Huth, Michael
    SENSORS, 2023, 23 (09)
  • [48] Activated platinum electrodes as transducer for a glucose sensor using glucose oxidase in a photopolymer membrane
    D. Wilke
    H. Müller
    N. Kolytsheva
    Fresenius' Journal of Analytical Chemistry, 1997, 357 : 534 - 538
  • [49] Low Levels of Serum 1,5-Anhydroglucitol during Pregnancy Compared to Blood Glucose Measured by Continuous Glucose Monitoring
    Moriya, Tatsumi
    Kishihara, Eriko
    Matsubara, Madoka
    Mori, Hiroko
    Okada, Yosuke
    Koga, Masafumi
    ANNALS OF CLINICAL AND LABORATORY SCIENCE, 2016, 46 (02) : 227 - 228
  • [50] Glucose Concentrations from Continuous Glucose Monitoring Devices Compared to Those from Blood Plasma during an Oral Glucose Tolerance Test in Healthy Young Adults
    Kontou, Thomas G.
    Sargent, Charli
    Roach, Gregory D.
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2021, 18 (24)