Continuous glucose monitoring in interstitial fluid using glucose oxidase-based sensor compared to established blood glucose measurement in rats

被引:27
|
作者
Woderer, Stefanie
Henninger, Nadja
Garthe, Claus-Dieter
Kloetzer, Hans Martin
Hajnsek, Martin
Kamecke, Ulrike
Gretz, Norbert
Kraenzlin, Bettina
Pill, Johannes
机构
[1] Roche Diagnost GmbH, D-68305 Mannheim, Germany
[2] Heidelberg Univ, Univ Hosp Mannheim, Med Res Ctr, D-68167 Mannheim, Germany
[3] Roche Diagnost GmbH, A-8020 Graz, Austria
[4] Metecon GmbH, D-68165 Mannheim, Germany
关键词
glucose biosensor; continuous glucose monitoring; glucose oxidase; glucose in interstitial fluid; blood glucose; rat;
D O I
10.1016/j.aca.2006.08.016
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Glucose monitoring is of importance for success of complex therapeutic interventions in diabetic patients. Its impact on treatment and glycemic control is demonstrated in large clinical trials. Up to eight blood glucose measurements per day are recommended. Notwithstanding, a substantial number of diabetic patients cannot or will not monitor their blood glucose appropriately. Considerable progress in control of disturbed metabolism in diabetic patients can be expected by continuous glucose monitoring. The aim of the study was to evaluate the performance of a new amperometric glucose oxidase-based glucose sensor in vitro and in vivo after subcutaneous implantation into rats. For in vitro testing current output of sensors was measured by exposure to increasing and decreasing glucose concentrations up to 472 mg dL(-1) over a time period of 7 days. After subcutaneous implantation of sensors into interscapular region of male rats glucose in interstitial fluid was evaluated and compared to glucose in arterial blood up to 7 days. Hyper- and hypoglycaemia were induced by intravenous application of glucose and insulin, respectively. Current of each implanted sensor was converted into glucose concentration using the first blood glucose measurement only. A change of current with glucose of 0.35 nA mg(-1) dL(-1) indicates high sensitivity of the sensor in vitro. The response time (90% of steady state) was calculated by approximately 60 s. Test strips for blood glucose measurement as reference for sensor readings was found as an appropriate and rapidly available method in rats by comparison with established hexokinase method in an automated lab analyzer with limits of agreement of +32.8 and -25.7 mg dL(-1) in Bland-Altman analysis. In normo- and hypoglycaemic range sensor readings in interstitial fluid correlated well with blood glucose measurements whereas hyperglycaemia was not reflected by the sensor completely when blood glucose was changing rapidly. The data given characterize a sensor with high sensitivity, long term stability and short response time. A single calibration of the sensor is required only in measurement periods up to 7 days. The findings demonstrate that the sensor is a highly promising candidate for assessment in humans. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:7 / 12
页数:6
相关论文
共 50 条
  • [1] Continuous Glucose Monitoring by Real Time Sensor in Interstitial Fluid
    Honsova, Sarka
    Navratil, Tomas
    XXXIII MODERNI ELEKTROCHEMICKE METODY, 2013, : 63 - 67
  • [2] Glucose measurement in interstitial fluid by microdialysis for the calibration of minimally invasive blood glucose monitoring
    Li, Dachao
    Wang, Ridong
    Chong, Hao
    Liu, Yu
    Xu, Kexin
    ADVANCED BIOMEDICAL AND CLINICAL DIAGNOSTIC SYSTEMS XI, 2013, 8572
  • [3] LOW NOISE MEASUREMENT OF PHOTOCURRENT FOR CONTINUOUS GLUCOSE MONITORING Low Noise Measurement System Enables Continuous Monitoring of Glucose in Subcutaneous Interstitial Fluid
    Cooley, Daniel W.
    Andersen, David R.
    BIODEVICES 2010: PROCEEDINGS OF THE THIRD INTERNATIONAL CONFERENCE ON BIOMEDICAL ELECTRONICS AND DEVICES, 2010, : 195 - 200
  • [4] Impedance-based polymer microneedle patch sensor for continuous interstitial fluid glucose monitoring
    Piao, Honglin
    Choi, Yong-Ho
    Kim, Jaehyun
    Park, Daerl
    Lee, Jia
    Khang, Dahl-Young
    Choi, Heon-Jin
    BIOSENSORS & BIOELECTRONICS, 2024, 247
  • [5] A glucose sensor protein for continuous glucose monitoring
    Veetil, Jithesh V.
    Jin, Sha
    Ye, Kaiming
    BIOSENSORS & BIOELECTRONICS, 2010, 26 (04) : 1650 - 1655
  • [6] Glucose Oxidase-Based Glucose-Sensitive Drug Delivery for Diabetes Treatment
    Zhao, Li
    Wang, Liyan
    Zhang, Yuhan
    Xiao, Shanshan
    Bi, Fei
    Zhao, Jianyu
    Gai, Guangqing
    Ding, Jianxun
    POLYMERS, 2017, 9 (07):
  • [7] An Interstitial Fluid Transdermal Extraction System for Continuous Glucose Monitoring
    Yu, Haixia
    Li, Dachao
    Roberts, Robert C.
    Xu, Kexin
    Tien, Norman C.
    JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2012, 21 (04) : 917 - 925
  • [8] Flap Monitoring Using Interstitial Fluid Glucose Measurements
    Kiuchi, Tomoki
    Ishii, Naohiro
    Uno, Takahiro
    Uoya, Yuichiro
    Sakai, Shigeki
    Matsuzaki, Kyoichi
    Kishi, Kazuo
    PLASTIC SURGERY, 2024, 32 (02) : 299 - 304
  • [9] Continuous monitoring of blood glucose
    Robert, JJ
    HORMONE RESEARCH, 2002, 57 : 81 - 84
  • [10] An Interstitial Fluid Transdermal Extraction Chip with Vacuum Generator and Volume Sensor for Continuous Glucose Monitoring
    Yu, Haixia
    Li, Dachao
    Ji, Yongjie
    Zhang, Xiaoli
    Xu, Kexin
    MICRO-NANO TECHNOLOGY XIV, PTS 1-4, 2013, 562-565 : 571 - 575