Synthetic Aperture Sequential Beamforming

被引:35
作者
Kortbek, Jacob [1 ]
Jensen, Jorgen Arendt [2 ]
Gammelmark, Kim Lokke [1 ]
机构
[1] BK Med, Milepk 34, DK-2730 Herlev, Denmark
[2] Tech Univ Denmark, Ctr Fast Ultrasound Imaging, Lyngby, Denmark
来源
2008 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-4 AND APPENDIX | 2008年
关键词
D O I
10.1109/ULTSYM.2008.0233
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
A synthetic aperture focusing (SAF) technique denoted Synthetic Aperture Sequential Beamforming (SASB) suitable for 2D and 3D imaging is presented. The technique differ from prior art of SAF in the sense that SAF is performed on pre-beamformed data contrary to channel data. The objective is to improve and obtain a more range independent lateral resolution compared to conventional dynamic receive focusing (DRIP) without compromising frame rate. SASB is a two-stage procedure using two separate beamformers. First a set of B-mode image lines using a single focal point in both transmit and receive is stored. The second stage applies the focused. image lines from the first stage as input data. The SASB method has been investigated using simulations in Field II and by off-line processing of data acquired with a commercial scanner. The performance of SASB with a static image object is compared with DRE For the lateral resolution the improvement in FWHM equals a factor of 2 and the improvement at -40 dB equals a factor of 3. With SASB the resolution is almost constant throughout the range. The resolution in the near field is slightly better for DRF. A decrease in performance at the transducer edges occur for both DRF and SASB, but is more profound for SASB.
引用
收藏
页码:966 / +
页数:2
相关论文
共 14 条
[1]   A study of synthetic-aperture imaging with virtual source elements in B-mode ultrasound imaging systems [J].
Bae, MH ;
Jeong, MK .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2000, 47 (06) :1510-1519
[2]   Synthetic aperture techniques with a virtual source element [J].
Frazier, CH ;
O'Brien, WD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1998, 45 (01) :196-207
[3]  
Gammelmark KL, 2003, ULTRASON, P1569
[4]  
Hazard CR, 1999, ULTRASON, P1221, DOI 10.1109/ULTSYM.1999.849217
[5]  
Jeong JS, 2000, ULTRASON, P1759, DOI 10.1109/ULTSYM.2000.921662
[6]   Adaptive multi-element synthetic aperture imaging with motion and phase aberration correction [J].
Karaman, M ;
Bilge, HS ;
O'Donnell, M .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1998, 45 (04) :1077-1087
[7]  
KORTBEK J, 2007, P IEEE ULTR IN PRESS
[8]  
Nikolov SI, 2003, ULTRASON, P1824
[9]   Virtual ultrasound sources in high-resolution ultrasound imaging. [J].
Nikolov, SI ;
Jensen, JA .
MEDICAL IMAGE 2002: ULTRASONIC IMAGING AND SIGNAL PROCESSING, 2002, 4687 :395-405
[10]  
Nikolov SI, 2000, ULTRASON, P1743, DOI 10.1109/ULTSYM.2000.921659