共 50 条
Mass Measurement and Top-Down HPLC/MS Analysis of Intact Monoclonal Antibodies on a Hybrid Linear Quadrupole Ion Trap-Orbitrap Mass Spectrometer
被引:120
|作者:
Bondarenko, Pavel V.
[1
]
Second, Tonya P.
[2
]
Zabrouskov, Vlad
[2
]
Makarov, Alexander A.
[3
]
Zhang, Zhongqi
机构:
[1] Amgen Inc, Dept Pharmaceut, Thousand Oaks, CA 91320 USA
[2] Thermo Fisher, San Jose, CA USA
[3] Thermo Fisher, Bremen, Germany
关键词:
LIQUID CHROMATOGRAPHY/MASS SPECTROMETRY;
IMMUNOGLOBULIN-GAMMA-ANTIBODIES;
STRUCTURAL-CHARACTERIZATION;
PROTEIN;
SITES;
PHARMACEUTICS;
DEAMIDATION;
STABILITY;
ISOFORMS;
RECEPTOR;
D O I:
10.1016/j.jasms.2009.03.020
中图分类号:
Q5 [生物化学];
学科分类号:
071010 ;
081704 ;
摘要:
Mass and top-down analyses of 150-kDa monoclonal immunoglobulin gamma (IgG) antibodies were performed on an Orbitrap analyzer. Three different sample delivery methods were tested including (1) infusion of an off-line desalted IgG sample using nano-electrospray; (2) on-line desalting followed by a step elution with a high percentage of organic solvent; and (3) reversed-phase HPLC separation and on-line mass and top-down analyses of disulfide isoforms of an IgG2 antibody. The accuracy of mass measurements of intact antibody was within +/-2 Da (15 ppm). The glycoforms of intact IgG antibodies separated by 162 Da were baseline resolved. In-source fragmentation of the intact antibodies produced mainly 115 residue fragments including N-terminal variable domains of heavy and light chains. The sequence coverage (the number of cleavages) was greatly increased after reduction of disulfide bonds and HPLC/MS/MS analysis of light and heavy chains using collision-induced dissociation in the ion trap of the LTQ-Orbitrap. This is an attractive alternative to peptide mapping for characterization and monitoring of post-translational modifications attributed to minimal sample preparation, high speed of the mass/top-down analysis, and relatively minor method-induced sample modifications. (J Am Soc Mass Spectrom 2009, 20, 1415-1424) (C) 2009 Published by Elsevier Inc. on behalf of American Society for Mass Spectrometry
引用
收藏
页码:1415 / 1424
页数:10