Characteristics and quality assurance of a dedicated open 0.23 T MRI for radiation therapy simulation

被引:35
作者
Mah, D
Steckner, M
Palacio, E
Mitra, R
Richardson, T
Hanks, GE
机构
[1] Fox Chase Canc Ctr, Dept Radiat Oncol, Philadelphia, PA 19111 USA
[2] Philips Med Syst, Cleveland, OH 44143 USA
关键词
quality assurance; open MRI; treatment planning; MRI simulation; distortions;
D O I
10.1118/1.1513991
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A commercially available open MRI unit is under routine use for radiation therapy simulation. The effects of a gradient distortion correction (GDC) program used to post process the images were assessed by comparison with the known geometry of a phantom. The GDC reduced the magnitude of the distortions at the periphery of the axial images from 12 mm to 2 mm horizontally along the central axis and distortions exceeding 20 mm were reduced to as little as 2 mm at the image periphery. Coronal and sagittal scans produced similar results. Coalescing these data into distortion as a function of radial distance, we found that for radial distances of <10 cm, the distortion after GDC was <2 mm and for radial distances up to 20 cm, the distortion was <5 mm. The dosimetric errors resulting from homogeneous dose calculations with this level of distortion of the external contour is <2%. A set of triangulation lasers has been added to establish a virtual isocenter for convenient setup and marking of patients and phantoms. Repeated measurements of geometric phantoms over several months showed variations in position between the virtual isocenter and the magnetic isocenter were constrained to <2 mm. Additionally, the interscan variations of 12 randomly selected points in space defined by a rectangular grid phantom was found to be within the intraobserver error of similar to1 mm in the coronal, sagittal, and transverse planes. Thus, the open MRI has sufficient geometric accuracy for most radiation therapy planning and is temporally stable. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:2541 / 2547
页数:7
相关论文
共 26 条
[1]   Magnetic resonance imaging system for three-dimensional conformal radiotherapy and its impact on gross tumor volume delineation of central nervous system tumors [J].
Aoyama, H ;
Shirato, H ;
Nishioka, T ;
Hashimoto, S ;
Tsuchiya, K ;
Kagei, K ;
Onimaru, R ;
Watanabe, Y ;
Miyasaka, K .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 50 (03) :821-827
[2]   ANALYSIS OF MACHINE-DEPENDENT AND OBJECT-INDUCED GEOMETRIC DISTORTION IN 2DFT MR IMAGING [J].
BAKKER, CJG ;
MOERLAND, MA ;
BHAGWANDIEN, R ;
BEERSMA, R .
MAGNETIC RESONANCE IMAGING, 1992, 10 (04) :597-608
[3]   The contribution of magnetic resonance imaging to the three-dimensional treatment planning of localized prostate cancer [J].
Debois, M ;
Oyen, R ;
Maes, F ;
Verswijvel, G ;
Gatti, G ;
Bosmans, H ;
Feron, M ;
Bellon, E ;
Kutcher, G ;
Van Poppel, H ;
Vanuytsel, L .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (04) :857-865
[4]   American Association of Physicists in Medicine radiation therapy committee task group 53: Quality assurance for clinical radiotherapy treatment planning [J].
Fraass, B ;
Doppke, K ;
Hunt, M ;
Kutcher, G ;
Starkschall, G ;
Stern, R ;
Van Dyke, J .
MEDICAL PHYSICS, 1998, 25 (10) :1773-1829
[5]   INTEGRATION OF MAGNETIC-RESONANCE-IMAGING INTO RADIATION-THERAPY TREATMENT PLANNING .1. TECHNICAL CONSIDERATIONS [J].
FRAASS, BA ;
MCSHAN, DL ;
DIAZ, RF ;
TENHAKEN, RK ;
AISEN, A ;
GEBARSKI, S ;
GLAZER, G ;
LICHTER, AS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1987, 13 (12) :1897-1908
[6]  
Fransson A., 2000, Radiotherapy and Oncology, V56, pS17
[7]   Aspects of MR image distortions in radiotherapy treatment planning [J].
Fransson, A ;
Andreo, P ;
Pötter, R .
STRAHLENTHERAPIE UND ONKOLOGIE, 2001, 177 (02) :59-73
[8]  
Freedman G. M., 2001, International Journal of Radiation Oncology Biology Physics, V51, P301, DOI 10.1016/S0360-3016(01)02377-X
[9]   Measurement of patient positioning errors in three-dimensional conformal radiotherapy of the prostate [J].
Hanley, J ;
Lumley, MA ;
Mageras, GS ;
Sun, J ;
Zelefsky, MJ ;
Leibel, SA ;
Fuks, Z ;
Kutcher, GJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 37 (02) :435-444
[10]   ACCURATE FRAMELESS REGISTRATION OF MR AND CT IMAGES OF THE HEAD - APPLICATIONS IN PLANNING SURGERY AND RADIATION-THERAPY [J].
HILL, DLG ;
HAWKES, DJ ;
GLEESON, MJ ;
COX, TCS ;
STRONG, AJ ;
WONG, WL ;
RUFF, CF ;
KITCHEN, ND ;
THOMAS, DGT ;
SOFAT, A ;
CROSSMAN, JE ;
STUDHOLME, C ;
GANDHE, AJ ;
GREEN, SEM ;
ROBINSON, GP .
RADIOLOGY, 1994, 191 (02) :447-454