Creatinine determination in urine from the point of view of reference values

被引:6
作者
Kratochvila, Josef
Friedecky, Bedrich
Budina, Marek
Sperlingova, Ilona [1 ]
机构
[1] Natl Inst Publ Hlth, Prague, Czech Republic
[2] SEKK EQA Syst Ltd Pardubice, Pardubice, Czech Republic
关键词
creatinine; urine; homogeneity; certification; uncertainty; EQA;
D O I
10.1007/s00769-006-0195-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this paper, the creation of a certified reference material for urinary creatinine is described. We used the Jaffe method and HPLC method for establishment of the certified value. Homogeneity tests are also described. We obtained material with sufficient homogeneity, stability, and with certified value (expanded uncertainty, k=2 for CI 95%) (7.77 +/- 0.27) mmol.L-1. This material was consequently used for the interlaboratory comparison (EQA Czech Republic for clinical chemistry). Twenty-nine percent of the participants obtained measurement results within the interval of the certified value +/- expanded uncertainty, while 85% of the participants obtained values inside the interval of the certified value +/- target measurement uncertainty. Direct use of the certified reference materials for method evaluation in EQA programs means a significant advance for monitoring and documentation traceability of results in routine measurements.
引用
收藏
页码:146 / 150
页数:5
相关论文
共 11 条
[1]  
BARDODEJ Z, 1980, EXPOSURE TESTS IND T, P39
[2]  
ISO, ISO 17511: 2003
[3]  
Jaffe M., 1886, PHYSIOL CHEM PHYS M, V10, P391, DOI DOI 10.1515/BCHM1.1886.10.5.391
[4]  
NARAYANAN S, 1980, CLIN CHEM, V26, P1119
[5]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF CREATININE IN SERUM, AND A CORRELATION OF THE RESULTS WITH THOSE OF THE JAFFE AND ENZYMATIC METHODS [J].
SCHNEIDERKA, P ;
PACAKOVA, V ;
STULIK, K ;
KLOUDOVA, M ;
JELINKOVA, K .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1993, 614 (02) :221-226
[6]  
SIEKMANN L, 1985, J CLIN CHEM CLIN BIO, V23, P137
[7]  
Stockl D, 1996, EUR J CLIN CHEM CLIN, V34, P319
[8]   Development of a liquid chromatography-mass spectrometry method for the high-accuracy determination of creatinine in serum [J].
Stokes, P ;
O'Connor, G .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 794 (01) :125-136
[9]  
SUCHANEK M, 2004, KVALIMETRIE
[10]  
VANNEVEL L, 2002, IMEP17 1