Analysis of polyhexamethylene biguanide in multipurpose contact lens solutions

被引:32
作者
Lucas, Anne D. [1 ]
Gordon, Edward A. [1 ]
Stratmeyer, Melvin E. [1 ]
机构
[1] US FDA, Ctr Devices & Radiol Hlth, OSEL DB, Silver Spring, MD 20993 USA
关键词
Contact lens; Polyhexamethylene biguanide; Liquid chromatography; HYDROCHLORIDE;
D O I
10.1016/j.talanta.2009.07.031
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The objective of this study was to establish a reasonably simple and reliable method to measure very low concentrations of polyhexamethylene biguanicle (PHMB) in multipurpose contact lens solutions (MPSs). By using a weak cation exchange solid phase extraction cartridge to extract the PHMB from MPS, followed by HPLC analysis using an evaporative light scattering detector, low levels (0.1 ppm) of PHMB were detected. Application of this method to a series of off-the-shelf MPS with PHMB as the active ingredient demonstrated these solutions contain I ppm. The contact lens solution with hydrogen peroxide as the active ingredient gave no peak where the PHMB peak eluted. The Polyquad (R) contact lens solution generated a peak close to the retention time of PHMB. Recovery of PHMB from fortified hydrogen peroxide contact lens solution was good at 0.25 ppm and above; 105% with a RSD of 17% or less. The repeatability of the HPLC system ranged from 4 to 11% RSD: the reproducibility of the entire method was less than 17.5% RSD. Storage and stability studies indicated that storage of MPS with PHMB for chemical analysis are not temperature dependent, but are affected by the composition of the container in which the contact lens solution is stored. Published by Elsevier B.V.
引用
收藏
页码:1016 / 1019
页数:4
相关论文
共 15 条
[1]   Determination of chlorhexidine digluconate and polyhexamethylene biguanide in eye drops by capillary electrophoresis with contactless conductivity detection [J].
Abad-Villar, EM ;
Etter, SF ;
Thiel, MA ;
Hauser, PC .
ANALYTICA CHIMICA ACTA, 2006, 561 (1-2) :133-137
[2]   The charge of glass and silica surfaces [J].
Behrens, SH ;
Grier, DG .
JOURNAL OF CHEMICAL PHYSICS, 2001, 115 (14) :6716-6721
[3]   THE EFFECT OF BAQUACIL ON PATHOGENIC FREE-LIVING AMEBAS (PFLA) .1. IN AXENIC CONDITIONS [J].
DAWSON, MW ;
BROWN, TJ ;
TILL, DG .
NEW ZEALAND JOURNAL OF MARINE AND FRESHWATER RESEARCH, 1983, 17 (03) :305-311
[4]   Evaluation of evaporative light-scattering detection for metabolite quantification without authentic analytical standards or radiolabel [J].
Fries, HE ;
Evans, CA ;
Ward, KW .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2005, 819 (02) :339-344
[5]   SYNERGISM WITHIN POLYHEXAMETHYLENE BIGUANIDE BIOCIDE FORMULATIONS [J].
GILBERT, P ;
PEMBERTON, D ;
WILKINSON, DE .
JOURNAL OF APPLIED BACTERIOLOGY, 1990, 69 (04) :593-598
[6]   Determination of biguanide groups in polyhexamethylene biguanide hydrochloride by titrimetric methods [J].
Hattori, T ;
Nakata, Y ;
Kato, R .
ANALYTICAL SCIENCES, 2003, 19 (11) :1525-1528
[7]   Characterization of ion exchange stationary phases via pH transition profiles [J].
Lendero, Nika ;
Vidic, Jana ;
Brne, Peter ;
Frankovic, Vida ;
Strancar, Ales ;
Podgornik, Ales .
JOURNAL OF CHROMATOGRAPHY A, 2008, 1185 (01) :59-70
[8]   Potentiometric titration of polyhexamethylene biguanide hydrochloride with potassium poly(vinyl sulfate) solution using a cationic surfactant-selective electrode [J].
Masadome, Takashi ;
Yamagishi, Yuichi ;
Takano, Masaki ;
Hattori, Toshiaki .
ANALYTICAL SCIENCES, 2008, 24 (03) :415-418
[9]  
MOMAN RH, 2007, MICROBIOL HYG ALIM, V19, P20
[10]   Biodegradability of end-groups of the biocide polyhexamethylene biguanide (PHMB) assessed using model compounds [J].
O'Malley, Leon P. ;
Collins, Andrew N. ;
White, Graham F. .
JOURNAL OF INDUSTRIAL MICROBIOLOGY & BIOTECHNOLOGY, 2006, 33 (08) :677-684