Using a Real-Time Location System for Assessment of Patient Ambulation in a Hospital Setting

被引:16
|
作者
Jeong, In Cheol [1 ]
Bychkov, David [1 ]
Hiser, Stephanie [2 ]
Kreif, Julie D. [2 ]
Klein, Lisa M. [2 ]
Hoyer, Erik H. [3 ]
Searson, Peter C. [1 ]
机构
[1] Johns Hopkins Univ, Johns Hopkins Individualized Hlth Initiat, Measurement Corps, 100 Croft Hall,3400 N Charles St, Baltimore, MD 21218 USA
[2] Johns Hopkins Univ Hosp, Baltimore, MD 21287 USA
[3] Johns Hopkins Univ, Dept Phys Med & Rehabil, Baltimore, MD USA
来源
关键词
Rehabilitation; Walk test; Walking; INTENSIVE-CARE-UNIT; 2-MINUTE WALK TEST; FUNCTIONAL AMBULATION; MOBILIZING PATIENTS; VALIDITY; 6-MINUTE; TESTS; RESPONSIVENESS; RELIABILITY; CAPACITY;
D O I
10.1016/j.apmr.2017.02.006
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
Objective: To assess the feasibility of using an infrared-based Real-Time Location System (RTLS) for measuring patient ambulation in a 2-minute walk test (2MWT) by comparing the distance walked and the Johns Hopkins Highest Level of Mobility (JH-HLM) score to clinician observation as a criterion standard. Design: Criterion standard validation study. Setting: Inpatient, university hospital. Participants: Patients (N=25) in an adult neuroscience/brain rescue unit. Interventions: Not applicable. Main Outcome Measures: RTLS and clinician-reported ambulation distance in feet, and JH-HLM score on an 8-point ordinal scale. Results: The RTLS ambulation distance for the 25 patients in the 2MWT was between 68 and 516ft. The mean difference between clinician reported and RTLS ambulation distance was 8.4 +/- 11.7ft (2.7% +/- 4.6%). The correlation between clinician-reported and RTLS ambulation distance was 97.9% (P<.01). The clinician-reported ambulation distance for 2 patients was +100ft and -99ft compared with the RTLS distance, implying clinician error in counting the number of laps (98ft). The correlation between the RTLS distance and clinician-reported distance excluding these 2 patients is 99.8% (P<.01). The accuracy of the RTLS for assessment of JH-HLM score for all 25 patients was 96%. The average patient speed obtained from RTLS data varied between 0.4 and 3.0mph. Conclusions: The RTLS is able to accurately measure patient ambulation and calculate JH-HLM for a 2MWT when compared with clinician observation as the criterion standard. (C) 2016 by the American Congress of Rehabilitation Medicine
引用
收藏
页码:1366 / 1373
页数:8
相关论文
共 50 条
  • [1] Real-Time Safety Risk Assessment Based on a Real-Time Location System for Hydropower Construction Sites
    Jiang, Hanchen
    Lin, Peng
    Fan, Qixiang
    Qiang, Maoshan
    SCIENTIFIC WORLD JOURNAL, 2014,
  • [2] Using a real-time location system to measure patient flow in a radiation oncology outpatient clinic
    Conley, Kevin
    Chambers, Chester
    Elnahal, Shereef
    Choflet, Amanda
    Williams, Kayode
    DeWeese, Theodore
    Herman, Joseph
    Dada, Maqbool
    PRACTICAL RADIATION ONCOLOGY, 2018, 8 (05) : 317 - 323
  • [3] Real-Time Monitoring System Using Location Based Service
    You, Jae-Hwe
    Lee, Young-Gu
    Jun, Moon-Seog
    UBIQUITOUS COMPUTING AND MULTIMEDIA APPLICATIONS, PT I, 2011, 150 : 369 - 379
  • [4] Contact tracing using real-time location system (RTLS): a simulation exercise in a tertiary hospital in Singapore
    Ng, Guan Yee
    Ong, Biauw Chi
    BMJ OPEN, 2022, 12 (10):
  • [5] RFID-based hospital real-time patient management system
    Chowdhury, Belal
    Khosla, Rajiv
    6TH IEEE/ACIS INTERNATIONAL CONFERENCE ON COMPUTER AND INFORMATION SCIENCE, PROCEEDINGS, 2007, : 363 - +
  • [6] Real-Time Location Systems for Hospital Emergency Response
    D'Souza, Ivor
    Ma, Wei
    Notobartolo, Cindy
    IT PROFESSIONAL, 2011, 13 (02) : 37 - 43
  • [7] Evaluation of real-time location systems in their hospital contexts
    Fisher, Jill A.
    Monahan, Torin
    INTERNATIONAL JOURNAL OF MEDICAL INFORMATICS, 2012, 81 (10) : 705 - 712
  • [8] TraceMe - Indoor Real-Time Location System
    Silva, Pedro M. Mestre A.
    Paralta, Maximino
    Caldeirinha, Rafael
    Rodrigues, Jorge
    Serodio, Carlos M. J. A.
    IECON: 2009 35TH ANNUAL CONFERENCE OF IEEE INDUSTRIAL ELECTRONICS, VOLS 1-6, 2009, : 2573 - +
  • [9] TraceMe - Indoor real-time location system
    Mestre A Silva, Pedro M.
    Paralta, Maximino
    Caldeirinha, Rafael
    Rodrigues, Jorge
    Serôdio, Carlos M.J.A.
    IECON Proceedings (Industrial Electronics Conference), 2009, : 2721 - 2725
  • [10] Characterization of emergency department abandonment using a real-time location system
    Geers, Jenna M.
    Pasupathy, Kalyan S.
    Lovik, Kimberly K.
    Finley, Janet L.
    Hellmich, Thomas R.
    Marisamy, Gomathi
    Nestler, David M.
    Sadosty, Annie T.
    Sir, Mustafa Y.
    Heaton, Heather A.
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2020, 38 (04): : 759 - 762