DOSIMETRIC IMPACT OF DAILY SETUP VARIATIONS DURING TREATMENT OF CANINE NASAL TUMORS USING INTENSITY-MODULATED RADIATION THERAPY

被引:17
作者
Deveau, Michael A. [1 ]
Gutierrez, Alonso N. [2 ]
Mackie, Thomas R. [1 ]
Tome, Wolfgang A. [1 ]
Forrest, Lisa J. [1 ]
机构
[1] Univ Wisconsin, Sch Med & Publ Hlth, Sch Vet Med, Madison, WI 53792 USA
[2] Univ Texas Hlth Sci Ctr San Antonio, Dept Radiat Oncol, Sch Med, Canc Therapy & Res Ctr, San Antonio, TX 78229 USA
关键词
canine nasal tumors; intensity-modulated radiation therapy; setup variations; tomotherapy; SIMULTANEOUSLY INTEGRATED BOOST; HELICAL TOMOTHERAPY; RADIOTHERAPY; IMRT; HEAD; ERRORS;
D O I
10.1111/j.1740-8261.2009.01629.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Intensity-modulated radiation therapy (IMRT) can be employed to yield precise dose distributions that tightly conform to targets and reduce high doses to normal structures by generating steep dose gradients. Because of these sharp gradients, daily setup variations may have an adverse effect on clinical outcome such that an adjacent normal structure may be overdosed and/or the target may be underdosed. This study provides a detailed analysis of the impact of daily setup variations on optimized IMRT canine nasal tumor treatment plans when variations are not accounted for due to the lack of image guidance. Setup histories of ten patients with nasal tumors previously treated using helical tomotherapy were replanned retrospectively to study the impact of daily setup variations on IMRT dose distributions. Daily setup shifts were applied to IMRT plans on a fraction-by-fraction basis. Using mattress immobilization and laser alignment, mean setup error magnitude in any single dimension was at least 2.5 mm (0-10.0 mm). With inclusions of all three translational coordinates, mean composite offset vector was 5.9 +/- 3.3 mm. Due to variations, a loss of equivalent uniform dose for target volumes of up to 5.6% was noted which corresponded to a potential loss in tumor control probability of 39.5%. Overdosing of eyes and brain was noted by increases in mean normalized total dose and highest normalized dose given to 2% of the volume. Findings suggest that successful implementation of canine nasal IMRT requires daily image guidance to ensure accurate delivery of precise IMRT distributions when non-rigid immobilization techniques are utilized. Unrecognized geographical misses may result in tumor recurrence and/or radiation toxicities to the eyes and brain.
引用
收藏
页码:90 / 96
页数:7
相关论文
共 29 条
  • [1] An accelerated technique for irradiation of malignant canine nasal and paranasal sinus tumors
    Adams, WM
    Miller, PE
    Vail, DM
    Forrest, LJ
    MacEwen, EG
    [J]. VETERINARY RADIOLOGY & ULTRASOUND, 1998, 39 (05) : 475 - 481
  • [2] [Anonymous], 1993, J ICRU
  • [3] Assessment of a radiotherapy patient immobilization device using single plane port radiographs and a remote computed tomography scanner
    Bley, CR
    Blattmann, H
    Roos, M
    Sumova, A
    Kaser-Hotz, B
    [J]. VETERINARY RADIOLOGY & ULTRASOUND, 2003, 44 (04) : 470 - 475
  • [4] IMRT: a review and preview
    Bortfeld, Thomas
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (13) : R363 - R379
  • [5] The university of Florida frameless high-precision stereotactic radiotherapy system
    Bova, FJ
    Buatti, JM
    Friedman, WA
    Mendenhall, WM
    Yang, CC
    Liu, C
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04): : 875 - 882
  • [6] OPTIMIZATION OF STATIONARY AND MOVING BEAM RADIATION-THERAPY TECHNIQUES
    BRAHME, A
    [J]. RADIOTHERAPY AND ONCOLOGY, 1988, 12 (02) : 129 - 140
  • [7] Intensity-modulated radiotherapy of head-and-neck cancer: Encouraging early results
    Eisbruch, A
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (01): : 1 - 3
  • [8] Ocular sparing using image-guided helical tomotherapy (IGHT) in spontaneous sino-nasal tumors in dogs
    Forrest, L. J.
    Lawrence, J. A.
    Miller, P. E.
    Juradat, H. A.
    Mackie, T. R.
    Dubielzig, R. R.
    Bentzen, S. M.
    Mehta, M. P.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (03): : S425 - S425
  • [9] Evaluation of dosimetric margins in prostate IMRT treatment plans
    Gordon, J. J.
    Siebers, J. V.
    [J]. MEDICAL PHYSICS, 2008, 35 (02) : 569 - 575
  • [10] A vacuum-formable mattress for veterinary radiotherapy positioning: Comparison with conventional methods
    Green, EM
    Forrest, LJ
    Adams, WM
    [J]. VETERINARY RADIOLOGY & ULTRASOUND, 2003, 44 (04) : 476 - 479