Glenohumeral kinematics following total shoulder arthroplasty: A finite element investigation

被引:22
作者
Hopkins, Andrew R.
Hansen, Ulrich N.
Amis, Andrew A.
Taylor, Mark
Emery, Roger J.
机构
[1] Univ London Imperial Coll Sci Technol & Med, Dept Engn Mech, Biomech Sect, London SW7 2AZ, England
[2] Imperial Coll London, Musculoskeletal Surg Dept, London, England
[3] Univ Southampton, Bioengn Res Grp, Southampton SO9 5NH, Hants, England
[4] St Marys Hosp, NHS Trust, London, England
关键词
shoulder; kinematics; TSA; finite element; arthroplasty;
D O I
10.1002/jor.20290
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The osseous geometry of the glenohumeral joint is naturally nonconforming and minimally constrained, and the joint's stability is maintained by action of the rotator cuff muscles. Damage to these muscles is often associated with joint degeneration, and a variety of glenoid prostheses have been developed to impart varying degrees of stability postoperatively. The issues of conformity and constraint within the artificial shoulder have been addressed through in vivo and in vitro studies, although few computational models have been presented. The current investigation presents the results of three-dimensional finite element analyses of the total shoulder joint, and the effects of design parameters upon glenohumeral interaction. Conformity was shown not to influence the loads required to destabilize the joint, although it was the principal factor determining the magnitude of humeral head translation. Constraint was found to correlate linearly with the forces required to dislocate the humeral head, with higher constraint leading to slightly greater humeral migration at the point of joint instability. The model predicts that patients with a dysfunctional supraspinatus would experience frequent eccentric loading of the glenoid, especially in the superior direction, which would likely lead to increased fixation stresses, and hence, a greater chance of loosening. For candidates with an intact rotator cuff, the models developed in this study predict that angular constraints of at least 14 degrees and 6.5 degrees in the superoinferior and anteroposterior axes are required to provide stable unloaded abduction of the humerus, with larger constraints of 18 degrees and 10 degrees necessitated by a dysfunctional supraspinatus. The tools developed during this study can be used to determine the capacity for different implant designs to provide resistance to excessive glenohumeral translations and reduce the potential for instability of the joint, allowing surgeons to optimize postoperative functional gains on a patient, by patient basis. (c) 2006 Orthopaedic Research Society. Published by Wiley Periodicals, Inc.
引用
收藏
页码:108 / 115
页数:8
相关论文
共 41 条
[1]   THE DANA TOTAL SHOULDER ARTHROPLASTY [J].
AMSTUTZ, HC ;
THOMAS, BJ ;
KABO, JM ;
JINNAH, RH ;
DOREY, FJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (08) :1174-1282
[2]   Glenohumeral contact forces [J].
Anglin, C ;
Wyss, UP ;
Pichora, DR .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2000, 214 (H6) :637-644
[3]  
Anglin C, 2000, J SHOULDER ELB SURG, V9, P104
[4]   Reimplantation of a glenoid component after component removal and allograft bone grafting: A report of 3 cases [J].
Antuna, S ;
Sperling, JW ;
Cofield, RH .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2002, 11 (06) :637-641
[5]   Experimental investigation of reaction forces at the glenohumeral joint during active abduction [J].
Apreleva, M ;
Parsons, IM ;
Warner, JJP ;
Fu, FH ;
Woo, SLY .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2000, 9 (05) :409-417
[6]   SURVIVORSHIP OF UNCONSTRAINED TOTAL SHOULDER ARTHROPLASTY [J].
BRENNER, BC ;
FERLIC, DC ;
CLAYTON, ML ;
DENNIS, DA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1989, 71A (09) :1289-1296
[7]   Fit of current glenoid component designs: An anatomic cadaver study [J].
Checroun, AJ ;
Hawkins, C ;
Kummer, FJ ;
Zuckerman, JD .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2002, 11 (06) :614-617
[8]  
COFIELD RH, 1977, I MECH ENG C JOINT R
[9]  
ELLMAN H, 1986, ORTHOP T, V10, P217
[10]  
ENGELBRECHT E, 1987, SHOULDER REPLACEMENT, P234