Temperature monitoring during thermal ablation on ex-vivo organs by Fiber Bragg gratings

被引:0
|
作者
Palumbo, Giovanna [1 ]
Iadicicco, Agostino [1 ]
Campopiano, Stefania [1 ]
Tosi, Daniele [2 ]
Verze, Paolo [3 ,4 ]
Carlomagno, Nicola [3 ,4 ]
Tammaro, Vincenzo [3 ,4 ]
Ippolito, Juliet [3 ,4 ]
机构
[1] Univ Naples Parthenope, Dept Engn, Naples, Italy
[2] Nazarbavev Univ, Sch Engn, Astana, Kazakhstan
[3] Univ Naples Federico II, Dept Urol, Naples, Italy
[4] Univ Naples Federico II, Dept Gen Surg, Naples, Italy
来源
关键词
optical fiber sensors; Fiber Bragg grating; radiofrequency thermal ablation; temperature sensor; RADIOFREQUENCY;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this paper we report on the application of Fiber Bragg gratings (FBGs) for temperature monitoring during radiofrequency ablation (RFA) for tumor treatment. A proper configuration with several FBG sensors deployed close to the region reached by RF discharges has been developed to monitor the necrotized area. Our ex-vivo experiments on animal kidney and liver confirm that we were able to monitor the temperature with a resolution of 0.1 degrees C during a series of different and consecutive RF discharges with the laparoscopic bipolar radiofrequency device Habib 4x.
引用
收藏
页数:3
相关论文
共 50 条
  • [31] Evaluation of Temperature Measurement Errors due to FBG Sensors During Laser Ablation of Ex-Vivo Porcine Liver
    Gassino, R.
    Vallan, A.
    Perrone, G.
    2018 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC): DISCOVERING NEW HORIZONS IN INSTRUMENTATION AND MEASUREMENT, 2018, : 1665 - 1669
  • [32] Simultaneous monitoring of strain and temperature in concrete structures with embedded fiber Bragg gratings
    Wang, Y
    Tijn, SC
    Sun, X
    Lim, TK
    Moyo, P
    Brownjohn, J
    SECOND INTERNATIONAL CONFERENCE ON EXPERIMENTAL MECHANICS, 2001, 4317 : 540 - 545
  • [33] Multipoint Temperature Monitoring of Microwave Thermal Ablation in Bones through Fiber Bragg Grating Sensor Arrays
    De Vita, Elena
    Zaltieri, Martina
    De Tommasi, Francesca
    Massaroni, Carlo
    Faiella, Eliodoro
    Zobel, Bruno Beomonte
    Iadicicco, Agostino
    Schena, Emiliano
    Grasso, Rosario Francesco
    Campopiano, Stefania
    SENSORS, 2020, 20 (11)
  • [34] Optical system for monitoring groundwater pressure and temperature using fiber Bragg gratings
    Ho, Yen-Te
    Wang, Yu-Li
    Chang, Liang-Cheng
    Wang, Tzu-Pin
    Tsai, Jui-Pin
    OPTICS EXPRESS, 2021, 29 (11): : 16032 - 16045
  • [35] Ultrasound properties and images of ex-vivo liver during thermal therapy
    Gertner, MR
    Wilson, BC
    Worthington, AE
    Sherar, MD
    1997 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 & 2, 1997, : 1353 - 1356
  • [36] Fiber Bragg Grating Thermometry and Post-Treatment Ablation Size Analysis of Radiofrequency Thermal Ablation on Ex Vivo Liver, Kidney and Lung
    Korganbayev, Sanzhar
    Bianchi, Leonardo
    Girgi, Clara
    Vergantino, Elva
    Santucci, Domiziana
    Faiella, Eliodoro
    Saccomandi, Paola
    SENSORS, 2025, 25 (01)
  • [37] Fiber Optic Sensors Based on Fiber Bragg Gratings for Methanol Steam Reforming Temperature Monitoring
    Trudel, E.
    Peppley, B. A.
    Wild, P.
    CHEMICAL SENSORS 12: CHEMICAL AND BIOLOGICAL SENSORS AND ANALYTICAL SYSTEMS, 2016, 75 (16): : 265 - 276
  • [38] A package for thermal compensation of fiber Bragg gratings
    Howlett, AL
    Vandeperre, LJ
    Clegg, WJ
    ADVANCES IN ELECTRONIC PACKAGING 2003, VOL 2, 2003, : 29 - 33
  • [39] Fiber Bragg gratings with enhanced thermal stability
    Brambilla, G
    Rutt, H
    APPLIED PHYSICS LETTERS, 2002, 80 (18) : 3259 - 3261
  • [40] Fast thermal regeneration of fiber Bragg gratings
    Bueno, Antonio
    Kinet, Damien
    Megret, Patrice
    Caucheteur, Christophe