Femoral head material loss at the head-neck junction in total hip arthroplasty: the effect of head size, stem material and stem offset

被引:8
作者
Valente, Giuseppe [1 ]
Lanting, Brent [1 ]
MacDonald, Steven [1 ]
Teeter, Matthew G. [2 ,3 ]
Van Citters, Douglas [4 ]
Howard, James [1 ]
机构
[1] Western Univ, London Hlth Sci Ctr, Schulich Sch Med & Dent, Div Orthopaed Surg,Dept Surg,Univ Hosp, London, ON, Canada
[2] Western Univ, Schulich Sch Med & Dent, Dept Med Biophys, London, ON, Canada
[3] Lawson Hlth Res Inst, Surg Innovat Program, London, ON, Canada
[4] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
关键词
Corrosion; metal-on-polyethylene; taper; total hip arthroplasty; trunnion; trunnionosis; TAPER JUNCTION; CORROSION; METAL; TRUNNIONOSIS; COMPONENTS; INTERFACE;
D O I
10.1177/1120700018814473
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: Material loss at the head-neck junction in total hip arthroplasty may cause adverse clinical symptoms and implant failure. The purpose of this study was to quantitatively examine the effects of head size, stem material and stem offset on material loss of the head-neck taper interface of a single trunnion design in retrieval implants of metal on polyethylene bearing surfaces. Methods: A retrieval study was performed to identify all 28-mm and 32-mm femoral heads from a single implant/taper design implanted for >2 years. This included n = 56 of the 28-mm heads and n = 23 of the 32-mm heads. The 28-mm femoral heads were matched to 32-mm femoral heads based on time in vivo and head length. A coordinate measuring machine was used to determine maximum linear corrosion depth (MLD). Differences in MLD for head diameter, stem material, and stem offset were determined. Results: There were no differences between groups for age, gender, BMI, or implantation time. There was no difference in MLD between 28 mm and 32 mm matched paired head diameters (p = 0.59). There was also no difference in MLD between titanium or cobalt-chromium stems (p = 0.79), and regular or high-offset stems (p = 0.95). Conclusion: There is no statistical difference in femoral head MLD at the head-neck junction in THA between 28-mm and 32-mm matched paired femoral heads, similar or mixed alloy coupled femoral head stem constructs, and regular or high offset stems.
引用
收藏
页码:647 / 651
页数:5
相关论文
共 50 条
  • [21] Spontaneous reduction in both dislocated head and dissociation of head-neck in a modular total hip arthroplasty
    Raviraj, A.
    Anand, Ashish
    Chakravarthy, Murali
    Pai, Sanjay
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2012, 22 : 127 - 130
  • [22] Head-Neck Taper Corrosion in Hip Arthroplasty
    Hussenbocus, S.
    Kosuge, D.
    Solomon, L. B.
    Howie, D. W.
    Oskouei, R. H.
    BIOMED RESEARCH INTERNATIONAL, 2015, 2015
  • [23] Understanding the role of head size and neck length in micromotion generation at the taper junction in total hip arthroplasty
    Bologna, Federico A.
    Putame, Giovanni
    Audenino, Alberto L.
    Terzini, Mara
    SCIENTIFIC REPORTS, 2024, 14 (01)
  • [24] Contact conditions for total hip head-neck modular taper junctions with microgrooved stem tapers
    Bechstedt, Maren
    Gustafson, Jonathan A.
    Mell, Steven P.
    Guehrs, Julian
    Morlock, Michael M.
    Levine, Brett R.
    Lundberg, Hannah J.
    JOURNAL OF BIOMECHANICS, 2020, 103
  • [25] Damage Patterns at the Head-Stem Taper Junction Helps Understand the Mechanisms of Material Loss
    Hothi, Harry S.
    Panagiotopoulos, Andreas C.
    Whittaker, Robert K.
    Bills, Paul J.
    McMillan, Rebecca A.
    Skinner, John A.
    Hart, Alister J.
    JOURNAL OF ARTHROPLASTY, 2017, 32 (01) : 291 - 295
  • [26] A novel method to improve femoral head and stem taper stability intraoperatively in total hip arthroplasty - a proof of concept study
    Dun, Shouchen
    Lim, Boon Him
    Swope, Stephen W.
    Whitaker, Dustin R.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2022, 108 (08)
  • [27] Short-stem hip arthroplasty in osteonecrosis of the femoral head
    Floerkemeier, Thilo
    Budde, Stefan
    Gronewold, Jens
    Radtke, Kerstin
    Ettinger, Max
    Windhagen, Henning
    von Lewinski, Gabriela
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2015, 135 (05) : 715 - 722
  • [28] Short-stem hip arthroplasty in osteonecrosis of the femoral head
    Thilo Floerkemeier
    Stefan Budde
    Jens Gronewold
    Kerstin Radtke
    Max Ettinger
    Henning Windhagen
    Gabriela von Lewinski
    Archives of Orthopaedic and Trauma Surgery, 2015, 135 : 715 - 722
  • [29] Determining material loss from the femoral stem trunnion in hip arthroplasty using a coordinate measuring machine
    Bone, Martin C.
    Sidaginamale, Raghavendra P.
    Lord, James K.
    Scholes, Susan C.
    Joyce, Thomas J.
    Nargol, Anthony V. F.
    Langton, David J.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2015, 229 (01) : 69 - 76
  • [30] Material Loss at the Taper Junction of Retrieved Large Head Metal-on-Metal Total Hip Replacements
    Matthies, Ashley K.
    Racasan, Radu
    Bills, Paul
    Blunt, Liam
    Cro, Suzie
    Panagiotidou, Anna
    Blunn, Gordon
    Skinner, John
    Hart, Alister J.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2013, 31 (11) : 1677 - 1685