Wearable Impedance Monitoring system for dialysis patients

被引:0
作者
Bonnet, S. [1 ,2 ]
Bourgerette, A. [3 ]
Gharbi, S. [3 ]
Rubeck, C. [3 ]
Arkouche, W. [4 ]
Massot, B. [5 ]
McAdams, E. [5 ]
Montalibet, A. [5 ]
Jallon, P. [3 ]
机构
[1] Univ Grenoble Alpes, F-38000 Grenoble, France
[2] CEA LETI MlNATEC Campus, F-38054 Grenoble, France
[3] CEA LETI, Grenoble, France
[4] AURAL, F-69008 Lyon, France
[5] INSA Lyon, INL UMR5270, Inst Nanotechnol Lyon, F-69100 Lyon, France
来源
2016 38TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC) | 2016年
关键词
BIOIMPEDANCE ANALYSIS; WHOLE-BODY;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper describes the development and the validation of a prototype wearable miniaturized impedance monitoring system for remote monitoring in home-based dialysis patients. This device is intended to assess the hydration status of dialysis patients using calf impedance measurements. The system is based on the low-power AD8302 component. The impedance calibration procedure is described together with the Cole parameter estimation and the hydric volume estimation. Results are given on a test cell to validate the design and on preliminary calf measurements showing Cole parameter variations during hemodialysis.
引用
收藏
页码:5196 / 5199
页数:4
相关论文
共 50 条
  • [21] Advancing Nutritional Care Through Bioelectrical Impedance Analysis in Critical Patients
    Dumitriu, Ana Maria
    Cobilinschi, Cristian
    Dumitriu, Bogdan
    Valcea, Sebastian
    Ungureanu, Raluca
    Popa, Angela
    Ene, Rcezvan
    Tincu, Radu
    Grintescu, Ioana Marina
    Mirea, Liliana
    NUTRIENTS, 2025, 17 (03)
  • [22] Validation of estimates of total body water in pediatric dialysis patients by deuterium dilution
    Mendley, SR
    Majkowski, NL
    Schoeller, DA
    KIDNEY INTERNATIONAL, 2005, 67 (05) : 2056 - 2062
  • [23] Change in Body Composition in Accordance With Residual Renal Function in Patients on Peritoneal Dialysis
    Kang, Seok Hui
    Cho, Kyu Hyang
    Park, Jong Won
    Yoon, Kyung Woo
    Do, Jun Young
    JOURNAL OF RENAL NUTRITION, 2013, 23 (06) : 438 - 444
  • [24] Monitoring of body water composition by the simultaneous use of bioelectrical impedance analysis and Crit-LineA® during hemodialysis
    Sugano, Naoki
    Yokoyama, Keitaro
    Kato, Naohiko
    Hara, Yoichiro
    Endo, Satoshi
    Mitome, Jun
    Kin, Taisei
    Tokudome, Goro
    Kuriyama, Satoru
    Hosoya, Tatsuo
    Yokoo, Takashi
    CLINICAL AND EXPERIMENTAL NEPHROLOGY, 2014, 18 (06) : 944 - 951
  • [25] Limb/Trunk Lean Mass Ratio as a Risk Factor for Mortality in Peritoneal Dialysis Patients
    Kang, Seok Hui
    Park, Jong on
    Yoo, Kyung Woo
    Do, Jun Young
    JOURNAL OF RENAL NUTRITION, 2013, 23 (04) : 315 - 323
  • [26] Relationship of ADMA Levels with Cardiovascular Parameters in Patients with Peritoneal Dialysis: A Bioimpedance Analysis Study
    Sumnu, Abdullah
    Cebeci, Egemen
    Ozturk, Savas
    Gursu, Meltem
    Kasapoglu, Ergun
    Ozkan, Oktay
    Gumus, Alper
    Gurdal, Ahmet
    Karadag, Serhat
    Kumbasar, Abdulbaki
    Kazancioglu, Rumeyza
    JOURNAL OF ACADEMIC RESEARCH IN MEDICINE-JAREM, 2018, 8 (03): : 147 - 152
  • [27] Clinical usefulness of bioimpedance analysis for assessing volume status in patients receiving maintenance dialysis
    Park, Jung Hwan
    Jo, Young-Il
    Lee, Jong-Ho
    KOREAN JOURNAL OF INTERNAL MEDICINE, 2018, 33 (04) : 660 - 669
  • [28] A Current Review of the Uses of Bioelectrical Impedance Analysis and Bioelectrical Impedance Vector Analysis in Acute and Chronic Heart Failure Patients: An Under-valued Resource?
    Thanapholsart, Jenjiratchaya
    Khan, Ehsan
    Lee, Geraldine A. A.
    BIOLOGICAL RESEARCH FOR NURSING, 2023, 25 (02) : 240 - 249
  • [29] Anthropometrical and Bioelectrical Impedance Analysis Parameters in Anorexia Nervosa Patients' Nutritional Status Assessment
    Popiolek, Joanna
    Teter, Mariusz
    Kozak, Gustaw
    Powrozek, Tomasz
    Mlak, Radoslaw
    Karakula-Juchnowicz, Hanna
    Malecka-Massalska, Teresa
    MEDICINA-LITHUANIA, 2019, 55 (10):
  • [30] Body composition by bioelectrical impedance, muscle strength, and nutritional risk in oropharyngeal dysphagia patients
    Ramos-Vazquez, Aniella G.
    Reyes-Torres, Carlos A.
    Castillo-Martinez, Lilia
    Serralde Zuniga, Aurora E.
    NUTRICION HOSPITALARIA, 2021, 38 (02) : 315 - 320