Wearable Impedance Monitoring system for dialysis patients

被引:0
作者
Bonnet, S. [1 ,2 ]
Bourgerette, A. [3 ]
Gharbi, S. [3 ]
Rubeck, C. [3 ]
Arkouche, W. [4 ]
Massot, B. [5 ]
McAdams, E. [5 ]
Montalibet, A. [5 ]
Jallon, P. [3 ]
机构
[1] Univ Grenoble Alpes, F-38000 Grenoble, France
[2] CEA LETI MlNATEC Campus, F-38054 Grenoble, France
[3] CEA LETI, Grenoble, France
[4] AURAL, F-69008 Lyon, France
[5] INSA Lyon, INL UMR5270, Inst Nanotechnol Lyon, F-69100 Lyon, France
来源
2016 38TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC) | 2016年
关键词
BIOIMPEDANCE ANALYSIS; WHOLE-BODY;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper describes the development and the validation of a prototype wearable miniaturized impedance monitoring system for remote monitoring in home-based dialysis patients. This device is intended to assess the hydration status of dialysis patients using calf impedance measurements. The system is based on the low-power AD8302 component. The impedance calibration procedure is described together with the Cole parameter estimation and the hydric volume estimation. Results are given on a test cell to validate the design and on preliminary calf measurements showing Cole parameter variations during hemodialysis.
引用
收藏
页码:5196 / 5199
页数:4
相关论文
共 50 条
[11]   Achieving Euvolemia in Peritoneal Dialysis Patients: A Surprisingly Difficult Proposition [J].
Tan, Boon Kay ;
Chan, Cian ;
Davies, Simon J. .
SEMINARS IN DIALYSIS, 2010, 23 (05) :456-461
[12]   The effect of dietary salt restriction on hypertension in peritoneal dialysis patients [J].
Inal, Salih ;
Erten, Yasemin ;
Tek, Nilufer ;
Ulusal Okyay, Gulay ;
Onec, Kursad ;
Akbulut, Gamze ;
Sanlier, Nevin .
TURKISH JOURNAL OF MEDICAL SCIENCES, 2014, 44 (05) :814-819
[13]   Body fluid overload and bioelectrical impedance analysis in renal patients [J].
Piccoli, A ;
Rossi, B ;
Pillon, L ;
Bucciante, G .
MINERAL AND ELECTROLYTE METABOLISM, 1996, 22 (1-3) :76-78
[14]   Bioelectrical Impedance Spectroscopy for Monitoring Mammalian Cells and Tissues under Different Frequency Domains: A Review [J].
Abasi, Sara ;
Aggas, John R. ;
Garayar-Leyva, Guillermo G. ;
Walther, Brandon K. ;
Guiseppi-Elie, Anthony .
ACS MEASUREMENT SCIENCE AU, 2022, 2 (06) :495-516
[15]   Association Between Peritonitis and Low Muscle Mass in Peritoneal Dialysis Patients [J].
Young, Jun Do ;
Kang, Seok Hui .
JOURNAL OF RENAL NUTRITION, 2020, 30 (04) :341-346
[16]   Effects of volume status on body composition in incident peritoneal dialysis patients [J].
Kang, Seok Hui ;
Do, Jun Young .
EUROPEAN JOURNAL OF CLINICAL NUTRITION, 2020, 74 (04) :633-641
[17]   NORMALIZED BIOIMPEDANCE INDICES ARE BETTER PREDICTORS OF OUTCOME IN PERITONEAL DIALYSIS PATIENTS [J].
Koh, Keng-Hee ;
Wong, Hin-Seng ;
Go, Kuan-Weng ;
Morad, Zaki .
PERITONEAL DIALYSIS INTERNATIONAL, 2011, 31 (05) :574-582
[18]   THE ASSOCIATION BETWEEN ARTERIAL STIFFNESS AND FLUID STATUS IN PERITONEAL DIALYSIS PATIENTS [J].
Kocyigit, Ismail ;
Sipahioglu, Murat Hayri ;
Orscelik, Ozcan ;
Unal, Aydin ;
Celik, Ahmet ;
Abbas, Samer R. ;
Zhu, Fansan ;
Tokgoz, Bulent ;
Dogan, Ali ;
Oymak, Oktay ;
Kotanko, Peter ;
Levin, Nathan W. .
PERITONEAL DIALYSIS INTERNATIONAL, 2014, 34 (07) :781-790
[19]   The comparative evaluation of patients' body dry weight under hemodialysis using two methods: Bioelectrical impedance analysis and conventional method [J].
Alijanian, Neda ;
Naini, Afsoon Emami ;
Shahidi, Shahrzad ;
Liaghat, Lida ;
Samani, Rahil Riahi .
JOURNAL OF RESEARCH IN MEDICAL SCIENCES, 2012, 17 (10) :923-927
[20]   Utility of bio-electrical impedance vector analysis for monitoring treatment of severe acute malnutrition in children [J].
Girma, Tsinuel ;
Kaestel, Pernille ;
Molgaard, Christian ;
Ritz, Christian ;
Andersen, Gregers S. ;
Michaelsen, Kim F. ;
Friis, Henrik ;
Wells, Jonathan C. K. .
CLINICAL NUTRITION, 2021, 40 (02) :624-631