Performance of Al2O3:C optically stimulated luminescence dosimeters for clinical radiation therapy applications

被引:15
作者
Hu, B. [1 ,2 ]
Wang, Y. [1 ,2 ]
Zealey, W. [1 ]
机构
[1] Univ Wollongong, Sch Engn Phys, Ctr Med Radiat Phys, Sydney, NSW, Australia
[2] Radiat Associate Oncol, St Vincents Clin, Sydney, NSW, Australia
关键词
Optical Stimulated Luminescence (OSL); Al2O3:C; in vivo dosimetry; radiation therapy; RADIOTHERAPY; OSL; ONCOLOGY;
D O I
10.1007/BF03179243
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A commercial Optical Stimulated Luminescence (OSL) dosimetry system developed by Landauer was tested to analyse the possibility of using OSL dosimetry for external beam radiotherapy planning checks. Experiments were performed to determine signal sensitivity, dose response range, beam type/energy dependency, reproducibility and linearity. Optical annealing processes to test OSL material reusability were also studied. In each case the measurements were converted into absorbed dose. The experimental results show that OSL dosimetry provides a wide dose response range, good linearity and reproducibility for the doses up to 800cGy. The OSL output is linear with dose up to 600cGy range showing a maximum deviation from linearity of 2.0% for the doses above 600cGy. The standard deviation in response of 20 dosimeters was 3.0%. After optical annealing using incandescent light, the readout intensity decreased by approximately 98% in the first 30 minutes. The readout intensity, I, decreased after repeated optical annealing as a power law, given by I proportional to t(-1.3). This study concludes that OSL dosimetry can provide an alternative dosimetry technique for use in in-vivo dosimetry if rigorous measurement protocols are established.
引用
收藏
页码:226 / 232
页数:7
相关论文
共 20 条
[1]  
AKSEKROD MS, 1993, RADIAT PROT DOSIM, V33, P159
[2]  
AKSELROD MS, 1990, RADIAT PROT DOSIM, V33, P119
[3]  
Antonov-Romanovsky V.V., 1956, C AC SCI USSR PEAC 1, P239
[4]   Real-time optical-fibre luminescence dosimetry for radiotherapy:: physical characteristics and applications in photon beams [J].
Aznar, MC ;
Andersen, CE ;
Botter-Jensen, L ;
Bäck, SÅJ ;
Mattsson, S ;
Kjær-Kristoffersen, F ;
Medin, J .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (09) :1655-1669
[5]  
BotterJensen L, 1996, RADIAT PROT DOSIM, V65, P273, DOI 10.1093/oxfordjournals.rpd.a031640
[6]  
BRAUULICH P, 1967, P 1 INT C LUM DOS ST
[7]   CW-OSL measurement protocols using optical fibre Al2O3:C dosemeters [J].
Edmund, J. M. ;
Andersen, C. E. ;
Marckmann, C. J. ;
Aznar, M. C. ;
Akselrod, M. S. ;
Botter-Jensen, L. .
RADIATION PROTECTION DOSIMETRY, 2006, 119 (1-4) :368-374
[8]  
*ICRU, 1976, ICRU PUBL, V24
[9]   Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements [J].
Jursinic, Paul A. .
MEDICAL PHYSICS, 2007, 34 (12) :4594-4604
[10]   COMPREHENSIVE QA FOR RADIATION ONCOLOGY - REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK-GROUP-40 [J].
KUTCHER, GJ ;
COIA, L ;
GILLIN, M ;
HANSON, WF ;
LEIBEL, S ;
MORTON, RJ ;
PALTA, JR ;
PURDY, JA ;
REINSTEIN, LE ;
SVENSSON, GK ;
WELLER, M ;
WINGFIELD, L .
MEDICAL PHYSICS, 1994, 21 (04) :581-618