Phantom-based characterization of distortion on a magnetic resonance imaging simulator for radiation oncology

被引:43
作者
Huang, Ke [1 ,2 ]
Cao, Yue [1 ]
Baharom, Umar [3 ]
Balter, James M. [1 ]
机构
[1] Univ Michigan Hosp, Dept Radiat Oncol, Ann Arbor, MI 48105 USA
[2] Georgia Regents Univ, Dept Radiat Oncol, Augusta, GA 30912 USA
[3] Integrated Med Technol, Troy, NY 12180 USA
关键词
MR image; distortion; sub-voxel accuracy; template matching; normalized cross correlation; GEOMETRIC DISTORTION; STRUCTURAL MRI; IMAGES; LOCALIZATION; REGISTRATION; ACCURACY;
D O I
10.1088/0031-9155/61/2/774
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
One of the major issues potentially limiting treatment planning with solely MR images is the possibility of geometric distortion inherent in MR images. We designed a large distortion phantom containing a 3D array of spheres and proposed a three-dimensional (3D) approach to determine the distortion of MR image volume. The approach to overcome partially filled spheres is also presented. The phantom was assembled with a 3D array of spheres filled with contrast and was scanned with a 3T MRI simulator. A 3D whole-sphere or half-sphere template is used to match the image pattern. The half-sphere template is used when the normalized cross-correlation value for the whole-sphere template is below a predetermined threshold. Procrustes method was applied to remove the shift induced by rotation and translation of the phantom. Then the distortion map was generated. Accuracy of the method was verified using CT images of a small phantom of the same design. The analysis of the small phantom showed that the method is accurate with an average offset of estimated sphere center 0.12 +/- 0.04 mm. The Procrustes analysis estimated the rotation angle to be 1.95 degrees and 0.01 degrees, respectively, when the phantom was placed at 2 degrees and 0 degrees from the ceiling laser. The analysis showed that on the central plane through the magnet center, the average displacement is less than 1 mm for all radii. At distal planes, when the radius is less than 18 cm, the average displacement is less than 1 mm. However, the average displacement is over 1 mm but still less than 1.5 mm for larger radii. A large distortion phantom was assembled and analysis software was developed to characterize distortions in MRI scans. The use of two templates helps reduce the potential impact of residual air bubbles in some of the spheres.
引用
收藏
页码:774 / 790
页数:17
相关论文
共 31 条
  • [1] MAGNETIC-RESONANCE IMAGE-DIRECTED STEREOTAXIC NEUROSURGERY - USE OF IMAGE FUSION WITH COMPUTERIZED-TOMOGRAPHY TO ENHANCE SPATIAL ACCURACY
    ALEXANDER, E
    KOOY, HM
    VANHERK, M
    SCHWARTZ, M
    BARNES, PD
    TARBELL, N
    MULKERN, RV
    HOLUPKA, EJ
    LOEFFLER, JS
    [J]. JOURNAL OF NEUROSURGERY, 1995, 83 (02) : 271 - 276
  • [2] [Anonymous], 1995, CANADIAN IMAGE PROCE
  • [3] Characterization, prediction, and correction of geometric distortion in 3 T MR images
    Baldwin, Lesley N.
    Wachowicz, Keith
    Thomas, Steven D.
    Rivest, Ryan
    Fallone, B. Gino
    [J]. MEDICAL PHYSICS, 2007, 34 (02) : 388 - 399
  • [4] Evaluation of the spatial accuracy of magnetic resonance imaging-based stereotactic target localization for gamma knife radiosurgery of functional disorders
    Bednarz, G
    Downes, B
    Corn, BW
    Curran, WJ
    Goldman, HW
    [J]. NEUROSURGERY, 1999, 45 (05) : 1156 - 1161
  • [5] Detection and correction of geometric distortion in 3D MR images
    Breeuwer, M
    Holden, M
    Zylka, W
    [J]. MEDICAL IMAGING: 2001: IMAGE PROCESSING, PTS 1-3, 2001, 4322 : 1110 - 1120
  • [6] Template matching using fast normalized cross correlation
    Briechle, K
    Hanebeck, UD
    [J]. OPTICAL PATTERN RECOGNITION XII, 2001, 4387 : 95 - 102
  • [7] A SURVEY OF IMAGE REGISTRATION TECHNIQUES
    BROWN, LG
    [J]. COMPUTING SURVEYS, 1992, 24 (04) : 325 - 376
  • [8] Evaluation of Hardware-related Geometrical Distortion in Structural MRI at 7 Tesla for Image-guided Applications in Neurosurgery
    Dammann, Philipp
    Kraff, Oliver
    Wrede, Karsten H.
    Oezkan, Neriman
    Orzada, Stephan
    Mueller, Oliver M.
    Sandalcioglu, I. Erol
    Sure, Ulrich
    Gizewski, Elke R.
    Ladd, Mark E.
    Gasser, Thomas
    [J]. ACADEMIC RADIOLOGY, 2011, 18 (07) : 910 - 916
  • [9] A complete distortion correction for MR images: I. Gradient warp correction
    Doran, SJ
    Charles-Edwards, L
    Reinsberg, SA
    Leach, MO
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (07) : 1343 - 1361
  • [10] INTEGRATION OF MAGNETIC-RESONANCE-IMAGING INTO RADIATION-THERAPY TREATMENT PLANNING .1. TECHNICAL CONSIDERATIONS
    FRAASS, BA
    MCSHAN, DL
    DIAZ, RF
    TENHAKEN, RK
    AISEN, A
    GEBARSKI, S
    GLAZER, G
    LICHTER, AS
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1987, 13 (12): : 1897 - 1908