Influence of the orientation of the Osstell® transducer during measurement of dental implant stability using resonance frequency analysis: A numerical approach

被引:20
作者
Capek, Lukas [1 ]
Simunek, Antonin [2 ]
Slezak, Radovan [2 ]
Dzan, Ladislav [3 ]
机构
[1] Tech Univ Liberec, Dept Mech Engn, Liberec 46117 1, Czech Republic
[2] Charles Univ Prague, Fac Med Hradec Kralove, Dept Dent, Hradec Kralove 50001, Czech Republic
[3] Reg Hosp Liberec, Dept Maxillofacial Surg, Liberec 46001, Czech Republic
关键词
Resonance frequency analysis; Dental implants; Stability; Finite element analysis; Resonance frequency; Eigenfrequency; DAMPING CAPACITY ASSESSMENT; IN-VITRO; INTERFACE; BEHAVIOR; SURGERY; SITES; MODEL; TIBIA;
D O I
10.1016/j.medengphy.2009.02.003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Stability of dental implants is measured by means of the Osstell (R) device using a method of resonance frequency analysis. The aim of the study was to evaluate the role of direction-dependence of the Osstell (R) transducer. For this purpose, a set of parametrical finite element analyses has been used. When rotating the transducer around the vertical axis the crucial phenomenon is the change in behavior of the system, which occurs in positions between 30 degrees and 80 degrees to the long axis of the alveolar crest. It seems from the presented results that, when measuring approximately perpendicularly to this long axis, the deviation from the ideal perpendicular position must not exceed 30 degrees. In this case, the first resonance frequency is recorded. When measuring in the approximately parallel position to the long axis of the alveolar crest, the second resonance frequency is recorded. The deviation from the ideal parallel position must not exceed 10 degrees. These values are not comparable with the previous ones. (C) 2009 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:764 / 769
页数:6
相关论文
共 29 条
  • [1] Comparative histomorphometry and resonance frequency analysis of implants with moderately rough surfaces in a loaded animal model
    Al-Nawas, B.
    Groetz, K. A.
    Goetz, H.
    Duschner, H.
    Wagner, W.
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2008, 19 (01) : 1 - 8
  • [2] Validity and clinical significance of biomechanical testing of implant/bone interface
    Aparicio, Carlos
    Lang, Niklaus R.
    Rangert, Bo
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2006, 17 : 2 - 7
  • [3] The design of a vibration transducer to monitor the integrity of dental implants
    Cawley, P
    Pavlakovic, B
    Alleyne, DN
    George, R
    Back, T
    Meredith, N
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1998, 212 (H4) : 265 - 272
  • [4] Evaluation of factors influencing resonance frequency analysis values, at insertion surgery, of implants placed in sinus-augmented and nongrafted sites
    Degidi, Marco
    Daprile, Giuseppe
    Piattelli, Adriano
    Carinci, Francesco
    [J]. CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2007, 9 (03) : 144 - 149
  • [5] DeTolla D H, 2000, J Oral Implantol, V26, P77, DOI 10.1563/1548-1336(2000)026<0077:TROTFE>2.3.CO
  • [6] 2
  • [7] Resonance frequency analysis in relation to jawbone characteristics and during early healing of implant installation
    Huwiler, M. A.
    Pjetursson, B. E.
    Bosshardt, D. D.
    Salvi, G. E.
    Lang, N. P.
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2007, 18 (03) : 275 - 280
  • [8] *INT DIAGN AB, OSST MENT US MAN
  • [9] *INT DIAGN AB, OSST US MAN
  • [10] Relevance of resonance frequency analysis to evaluate dental implant stability: simulation and histomorphometrical animal experiments
    Ito, Yoshihiro
    Sato, Daisuke
    Yoneda, Satoshi
    Ito, Daisuke
    Kondo, Hisatomo
    Kasugai, Shohei
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2008, 19 (01) : 9 - 14