A 15-crown-5-functionalized carbosilane dendrimer as ionophore for ammonium selective electrodes

被引:19
作者
Chandra, Sudeshna [1 ]
Buschbeck, Roy [1 ]
Lang, Heinrich [1 ]
机构
[1] Tech Univ Chemnitz, Fak Nat Wissensch, Inst Chem, Lehrstuhl Anorgan Chem, D-09111 Chemnitz, Germany
关键词
carbosilane dendrimer; crown ether; ammonium; ion selective electrode; ionophore;
D O I
10.1016/j.talanta.2006.02.042
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The synthesis of CH2=CHCH2OCH2[15-crown-5] (III) is achieved by the treatment of HOCH2-[15-crown-5] (I) with equimolar amounts of CH2=CHCH2Br (II) in the presence of KOH. The hydrosilylation of III with Si(CH2CH2CH2SiMe2H)(4) (IV) in the presence of the Karstedt catalyst affords the crown ether end-capped carbosilane dendrimer Si(CH2CH2CH2Si-Me2CH2CH2CH2OCH2[15-crown-5,])(4) (V). PVC-based membranes of V as ionophore with sodium tetraphenyl borate (NaTPB) as anion excluder and dioctyl phthalate (DOP), diphenyl ether (DPE), dibutyl amine (DBA) and dibutyl phthalate (DBP) as plasticizing solvent mediators were prepared and investigated as NH4+-selective electrode. The response of the electrode was linear with a Nernstian slope of 53.3 mV/decade over an NH4+ ion concentration range of 7.60 x 10(-6) to 1.0 x 10(-1) M and a detection limit of 3.9 x 10(-6) M. The response time to achieve a steady potential for NH4+ ions was between 6 and 10s, and the electrode is suitable for use within the pH range of 2.2-8.5. The selectivity relative to alkali, alkaline earth. and transition heavy metal ions is good. The newly developed ionophore showed higher NH4+ selectivity over K+ (log K-NH4 '+K+(Pot) = - 1.801) and Na+ (log K-NH4 '+Na+(Pot) = - 2.850). The electrode could be used for at least 45 days without considerable alteration in its potential. The electrode also shows a better working concentration range and slope in comparison to other NH4+-selective electrodes reported in literature. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:1087 / 1093
页数:7
相关论文
共 41 条
[11]  
Frey H, 2000, TOP CURR CHEM, V210, P69
[12]   EVALUATION OF AN AMMONIUM IONOPHORE FOR USE IN POLY(VINYL CHLORIDE) MEMBRANE ION-SELECTIVE ELECTRODES - SOLVENT MEDIATOR EFFECTS [J].
GHAURI, MS ;
THOMAS, JDR .
ANALYST, 1994, 119 (11) :2323-2326
[13]   MEMBRANE TECHNOLOGY AND DYNAMIC-RESPONSE OF ION-SELECTIVE LIQUID-MEMBRANE ELECTRODES [J].
HUSER, M ;
GEHRIG, PM ;
MORF, WE ;
SIMON, W ;
LINDNER, E ;
JENEY, J ;
TOTH, K ;
PUNGOR, E .
ANALYTICAL CHEMISTRY, 1991, 63 (14) :1380-1386
[14]   Functional dendrimers, hyperbranched and star polymers [J].
Inoue, K .
PROGRESS IN POLYMER SCIENCE, 2000, 25 (04) :453-571
[15]   End-capped carbosiloxane dendrimers with cholesterol and pyridine derivatives [J].
Kim, C ;
Park, J .
JOURNAL OF ORGANOMETALLIC CHEMISTRY, 2001, 629 (1-2) :194-200
[16]   Thiazole containing benzo crown ethers: A new class of ammonium selective ionophores [J].
Kim, HS ;
Park, HJ ;
Oh, HJ ;
Koh, YK ;
Choi, JH ;
Lee, DH ;
Cha, GS ;
Nam, H .
ANALYTICAL CHEMISTRY, 2000, 72 (19) :4683-4688
[17]   POTENTIOMETRIC SILICON MICROSENSOR FOR NITRATE AND AMMONIUM [J].
KNOLL, M ;
CAMMANN, K ;
DUMSCHAT, C ;
SUNDERMEIER, C ;
ESHOLD, J .
SENSORS AND ACTUATORS B-CHEMICAL, 1994, 18 (1-3) :51-55
[18]   THEORY AND APPLICATIONS OF ION-SELECTIVE ELECTRODES .8. [J].
KORYTA, J .
ANALYTICA CHIMICA ACTA, 1990, 233 (01) :1-30
[19]   POTENTIOMETRIC ION-SELECTIVE AND BIOSELECTIVE ELECTRODES BASED ON ASYMMETRIC POLYURETHANE MEMBRANES [J].
LIU, D ;
MEYERHOFF, ME ;
GOLDBERG, HD ;
BROWN, RB .
ANALYTICA CHIMICA ACTA, 1993, 274 (01) :37-46
[20]   Dendrimers containing heteroatoms (Si, P, B, Ge, or Bi) [J].
Majoral, JP ;
Caminade, AM .