Barricor blood collection tubes are equivalent to PST for a variety of chemistry and immunoassay analytes except for lactate dehydrogenase

被引:11
作者
Raizman, Joshua E. [1 ,2 ]
Goudreau, Bobbi Lynn [2 ]
Fuzery, Anna K. [1 ,2 ]
Cembrowski, George S. [1 ,2 ]
机构
[1] Univ Alberta, Dept Lab Med & Pathol, Edmonton, AB, Canada
[2] Alberta Publ Labs Ltd, Edmonton, AB, Canada
关键词
Blood collection; Preanalytical; Centrifugation; Lactate dehydrogenase; LITHIUM-HEPARIN; PLASMA SAMPLES; TROPONIN-I; SERUM; DEVICES;
D O I
10.1016/j.cca.2019.06.013
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Introduction: The BD Barricor tube uses a novel mechanical separator designed to eliminate gel artifacts, decrease cellular contamination, and improve stability. Here, we evaluated the Barricor tube as a possible replacement for PST using Beckman Coulter analyzers under both optimal, alternative, and suboptimal centrifugation conditions based on BD recommendations. Methods: Paired PST and Barricor samples were collected from 4 local hospitals and processed based on site-specific preanalytical systems involving automated or manual centrifugation. Centrifugation conditions ranged from 1912 x g for 10 min (suboptimal), 2060g for 10 min (alternative), and 4000 x g for 3 or 10 min (optimal). Tube volume (4.5 vs. 5.5 ml) was also assessed. Forty-three chemistry and immunochemistry analytes were measured on Beckman Coulter DxC and DxI analyzers. Results: Using an automated preanlaytical system with suboptimal spin conditions, no bias between PST and Barricor was observed for all analytes tested except lactate dehydrogenase (LD). Further investigation revealed significant increase in LD when Barricor was spun for 10 min at 1912, 2060 and 4000 x g, ranging from +7.4-19.4% vs. PST across the entire measurement interval (87-493 U/l). Smaller tube volume was also associated with higher LD. Differences in LD occurred despite no change in other hemolysis markers such as potassium, phosphate, and AST. Conclusions: LD is most sensitive to varying centrifugation conditions (time and speed) in Barricor tubes. We recommend that BD centrifugation protocols should be closely evaluated to determine if Barricor is equivalent to PST under local preanalytical configurations.
引用
收藏
页码:18 / 24
页数:7
相关论文
共 26 条
[1]   The local clinical validation of a new lithium heparin tube with a barrier: BD Vacutainer® Barricor LH Plasma tube [J].
Arslan, Fatma Demet ;
Karakoyun, Inanc ;
Basok, Banu Isbilen ;
Aksit, Merve Zeytinli ;
Baysoy, Anil ;
Ozturk, Yasemin Kilic ;
Guclu, Yusuf Adnan ;
Duman, Can .
BIOCHEMIA MEDICA, 2017, 27 (03) :535-545
[2]   Effect of blood collection tubes on the incidence of artifactual hyperkalemia on patient samples from an outreach clinic [J].
Babic, Nikolina ;
Zibrat, Steven ;
Gordon, Ilyssa O. ;
Lee, Christine C. ;
Yeo, Kiang-Teck J. .
CLINICA CHIMICA ACTA, 2012, 413 (19-20) :1454-1458
[3]   Cellular content in plasma of Becton-Dickinson lithium-heparin tubes: cause of unreliable results in the IFCC-recommended lactate dehydrogenase method [J].
Bakker, Andries J. ;
Bakker, Anneke ;
Renting-Wiering, Hennie .
ANNALS OF CLINICAL BIOCHEMISTRY, 2006, 43 :510-512
[4]   Stability of plasma electrolytes in Barricor and PST II tubes under different storage conditions [J].
Bautista Balbas, Luis Alfredo ;
Segovia Amaro, Marta ;
Gomez Rioja, Ruben ;
Alcaide Martin, Maria Jose ;
Buno Soto, Antonio .
BIOCHEMIA MEDICA, 2017, 27 (01) :225-230
[5]  
BD Life Science, 2016, CISC VIS NETW IND GL
[6]  
BD Life Sciences, 2016, BD VACC BARR LH PLAS
[7]   Blood collection tubes as medical devices: The potential to affect assays and proposed verification and validation processes for the clinical laboratory [J].
Bowen, Raffick A. R. ;
Adcock, Dorothy M. .
CLINICAL BIOCHEMISTRY, 2016, 49 (18) :1321-1330
[8]   Impact of blood collection devices on clinical chemistry assays [J].
Bowen, Raffick A. R. ;
Hortin, Glen L. ;
Csako, Gyorgy ;
Otanez, Oscar H. ;
Remaley, Alan T. .
CLINICAL BIOCHEMISTRY, 2010, 43 (1-2) :4-25
[9]   Influence of centrifugation conditions on the results of 77 routine clinical chemistry analytes using standard vacuum blood collection tubes and the new BD-Barricor tubes [J].
Cadamuro, Janne ;
Mrazek, Cornelia ;
Leichtle, Alexander B. ;
Kipman, Ulrike ;
Felder, Thomas K. ;
Wiedemann, Helmut ;
Oberkofler, Hannes ;
Fiedler, Georg M. ;
Haschke-Becher, Elisabeth .
BIOCHEMIA MEDICA, 2018, 28 (01)
[10]   What's floating on my plasma? [J].
Cadamuro, Janne ;
Wiedemann, Helmut ;
Felder, Thomas K. ;
Mrazek, Cornelia ;
Kipman, Ulrike ;
Hannes, Oberkofler ;
Haschke-Becher, Elisabeth .
BIOCHEMIA MEDICA, 2017, 27 (02) :430-433