Electrochemical investigations of corticosteroid isomers-testosterone and epitestosterone and their simultaneous determination in human urine

被引:43
作者
Goyal, Rajendra N. [1 ]
Gupta, Vinod K. [1 ,2 ]
Chatterjee, Sanghamitra [1 ]
机构
[1] Indian Inst Technol, Dept Chem, Roorkee 247667, Uttar Pradesh, India
[2] King Fahd Univ Petr & Minerals, Dhahran 31261, Saudi Arabia
关键词
Doping; Single-wall carbon nanotubes; Edge plane pyrolytic graphite electrode; Testosterone; Epitestosterone; Human urine; CHROMATOGRAPHY-MASS SPECTROMETRY; LIQUID-CHROMATOGRAPHY; VOLTAMMETRIC DETERMINATION; BIOLOGICAL-FLUIDS; ANABOLIC-STEROIDS; QUANTIFICATION; ELECTRODE; SAMPLES; RATIO; ADENOSINE;
D O I
10.1016/j.aca.2009.10.035
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Voltammetric investigation of two corticoid isomers-testosterone and epitestosterone has been carried out at bare and single-wall carbon nanotubes (SWNT)-modified edge plane pyrolytic graphite electrode (EPPGE). Square wave voltammetry (OSWV) has been used for the simultaneous determination of isomeric steroids. The reduction of the two isomers occurred in a pH dependent, 2e, 2H(+) process and well-defined voltammetric peaks were observed. Under the optimum experimental conditions, linear calibration curves are obtained within the concentration range 5-1000 nM for both the steroids with the limit of detection 2.8 x 10(-9) and 4.1 x 10(-9) M for testosterone and epitestosterone respectively. The developed protocol is successfully implemented for the analysis of both the compounds in the urine samples of normal subjects as well as in patients undergoing treatment with testosterone. The results obtained from the proposed voltammetric method were also compared with HPLC analysis and found to be similar. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:147 / 153
页数:7
相关论文
共 34 条
[1]  
Aguilera R, 2002, CLIN CHEM, V48, P629
[2]   Analysis of anabolic steroids by partial filling micellar electrokinetic capillary chromatography and electrospray mass spectrometry [J].
Amundsen, LK ;
Kokkonen, JT ;
Rovio, S ;
Sirén, H .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1040 (01) :123-131
[3]   Carbon nanotubes contain metal impurities which are responsible for the "electrocatalysis" seen at some nanotube-modified electrodes [J].
Banks, CE ;
Crossley, A ;
Salter, C ;
Wilkins, SJ ;
Compton, RG .
ANGEWANDTE CHEMIE-INTERNATIONAL EDITION, 2006, 45 (16) :2533-2537
[4]  
BIRRELL SN, 2007, Patent No. 200758
[5]   MEASUREMENT OF STEROID-HORMONES IN PLASMA BY ISOCRATIC HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY COUPLED TO RADIOIMMUNOASSAY [J].
BOSCHI, S ;
DEIASIO, R ;
MESINI, P ;
BOLELLI, GF ;
SCIAJNO, R ;
PASQUALI, R ;
CAPELLI, M .
CLINICA CHIMICA ACTA, 1994, 231 (01) :107-113
[6]  
CATLIN DH, 1987, CLIN CHEM, V33, P319
[7]   Quantification of testosterone and epitestosterone in biological samples by capillary electrophoresis with immunoaffinity extraction [J].
Chen, Hong-Xu ;
Deng, Qin-Pei ;
Zhang, Li-Wei ;
Zhang, Xin-Xiang .
TALANTA, 2009, 78 (02) :464-470
[8]   Determination of testosterone : epitestosterone ratio after pentafluorophenyldimethylsilyl-trimethyls derivatisation using gas chromatography-mass spectrometry in equine urine [J].
Choi, MH ;
Kim, JY ;
Chung, BC .
ANALYST, 1999, 124 (05) :675-678
[9]  
Christian G.D., 1962, JELECTROANAL CHEM, V3, P363
[10]   Serum inhibin B, FSH, LH and testosterone levels before and after human chorionic gonadotropin stimulation in prepubertal boys with cryptorchidism [J].
Christiansen, P ;
Andersson, AM ;
Skakkebæk, NE ;
Juul, A .
EUROPEAN JOURNAL OF ENDOCRINOLOGY, 2002, 147 (01) :95-101