Time resolved speckle contrast optical spectroscopy at quasi-null source-detector separation for non-invasive measurement of microvascular blood flow

被引:8
作者
Pagliazzi, Marco [1 ]
Colombo, Lorenzo [2 ]
Vidal-Rosas, Ernesto E. [1 ]
Dragojevic, Tanja [1 ]
Parfentyeva, Veronika [1 ]
Culver, Joseph P. [3 ]
Sekar, Santhana Konugolu Venkata [2 ]
Sieno, Laura Di [2 ]
Contini, Davide [2 ]
Torricelli, Alessandro [2 ,4 ]
Pifferi, Antonio [2 ,4 ]
Mora, Alberto Dalla [2 ]
Durduran, Turgut [1 ,5 ]
机构
[1] Barcelona Inst Sci & Technol, ICFO Inst Ciencies Foton, Barcelona 08860, Spain
[2] Politecn Milan, Dipartimento Fis, I-20133 Milan, Italy
[3] Washington Univ, Dept Biomed Engn, St Louis, MO 63130 USA
[4] CNR, Ist Foton & Nanotecnol, I-20133 Milan, Italy
[5] Inst Catalana Recerca & Estudis Avancats ICREA, Barcelona 08015, Spain
基金
欧盟地平线“2020”;
关键词
DIFFUSE CORRELATION SPECTROSCOPY; WAVE SPECTROSCOPY; CORRELATION TOMOGRAPHY; IN-VIVO; SCATTERING; ACCURACY; DEVICE; BRAIN;
D O I
10.1364/BOE.418882
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Time (or path length) resolved speckle contrast optical spectroscopy (TD-SCOS) at quasi-null (2.85 mm) source-detector separation was developed and demonstrated. The method was illustrated by in vivo studies on the forearm muscle of an adult subject. The results have shown that selecting longer photon path lengths results in higher hyperemic blood flow change and a faster return to baseline by a factor of two after arterial cuff occlusion when compared to SCOS without time resolution. This indicates higher sensitivity to the deeper muscle tissue. In the long run, this approach may allow the use of simpler and cheaper detector arrays compared to time resolved diffuse correlation spectroscopy that are based on readily available technologies. Hence, TD-SCOS may increase the performance and decrease cost of devices for continuous non-invasive, deep tissue blood flow monitoring. (c) 2021 Optical Society of America under the terms of the OSA Open Access Publishing Agreement
引用
收藏
页码:1499 / 1511
页数:13
相关论文
共 53 条
  • [51] Noncontact Speckle Contrast Diffuse Correlation Tomography of Blood Flow Distributions in Burn Wounds: A Preliminary Study
    Zhao, Mingjun
    Mazdeyasna, Siavash
    Huang, Chong
    Agochukwu-Nwubah, Nneamaka
    Bonaroti, Alisha
    Wong, Lesley
    Yu, Guoqiang
    [J]. MILITARY MEDICINE, 2020, 185 : 82 - 87
  • [52] Diffuse optical correlation tomography of cerebral blood flow during cortical spreading depression in rat brain
    Zhou, C
    Yu, GQ
    Furuya, D
    Greenberg, JH
    Yodh, AG
    Durduran, T
    [J]. OPTICS EXPRESS, 2006, 14 (03) : 1125 - 1144
  • [53] Highly parallel, interferometric diffusing wave spectroscopy for monitoring cerebral blood flow dynamics
    Zhou, Wenjun
    Kholiqov, Oybek
    Chong, Shau Poh
    Srinivasan, Vivek J.
    [J]. OPTICA, 2018, 5 (05): : 518 - 527