A Comparison of the Deflecting-Tip Bronchial Blocker With a Wire-Guided Blocker or Left-Sided Double-Lumen Tube

被引:34
|
作者
Dumans-Nizard, Virginie [1 ]
Liu, Ngai [1 ]
Laloe, Pierre-Antoine [1 ]
Fischler, Marc [1 ]
机构
[1] Hop Foch, Dept Anesthesiol, F-92151 Suresnes, France
关键词
one-lung ventilation; double-lumen endotracheal tube; bronchial blocker; TORQUE CONTROL BLOCKER; ONE-LUNG VENTILATION; ENDOTRACHEAL-TUBE; CARDIAC-SURGERY;
D O I
10.1053/j.jvca.2009.02.002
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objective: To compare a new bronchial blocker, the Cohen blocker, with the Arndt blocker and a left double-lumen tube (DLT). Design: A prospective, randomized, controlled trial. Setting: University hospital. Participants: Forty-eight patients undergoing lung surgery. Intervention: Intubation with 1 of the 3 devices. Comparisons among groups included (1) time for initial positioning, (2) degree of lung collapse at pleura opening, and (3) number of intraoperative fiberoptic examinations. Measurements and Main Results: Positioning of the Cohen blocker (256 [166-341] seconds; median [interquartile range]) took no longer compared with the Arndt blocker (253 [184-305] seconds), and there was a trend toward difference between the 2 blockers and the DLT (137 [102-199] seconds) (p = 0.07). The time to place the Cohen blocker was longer in cases of left bronchus occlusion compared with a right one (340 [300-450] v 170 [124-259] seconds, p = 0.02); they were similar in the Arndt group. The degree of lung collapse was different among groups (p = 0.05), but the difference between any pair did not reach statistical significance. The number of patients who required at least 1 additional FOB examination was not statistically different (50% of patients in each blocker group v 19% in the DLT group). Conclusions: There was a trend toward a difference between times to place a bronchial blocker and the DLT. The Cohen blocker is more difficult to position in the left main bronchus than in the right one. (C) 2009 Elsevier Inc. All rights reserved
引用
收藏
页码:501 / 505
页数:5
相关论文
共 50 条
  • [1] Lung isolation in the morbidly obese patient: a comparison of a left-sided double-lumen tracheal tube with the Arndt wire-guided blocker
    Campos, J. H.
    Hallam, E. A.
    Ueda, K.
    BRITISH JOURNAL OF ANAESTHESIA, 2012, 109 (04) : 630 - 635
  • [2] The wire-guided endobronchial blocker as a solution to provide one-lung ventilation when a double-lumen endotracheal tube is malpositioned
    Kuhlman, G
    Legros, C
    Laloë, PA
    Puyo, P
    Fischler, M
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2003, 17 (05) : 636 - 637
  • [3] Con: A Bronchial Blocker Is Not a Substitute for a Double-Lumen Endobronchial Tube
    Brodsky, Jay B.
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2015, 29 (01) : 237 - 239
  • [4] Disconnection Technique With a Bronchial Blocker for Improving Lung Deflation: A Comparison With a Double-Lumen Tube and Bronchial Blocker Without Disconnection
    Yoo, Ji Young
    Kim, Dae Hee
    Choi, Ho
    Kim, Kun
    Chae, Yun Jeong
    Park, Sung Yong
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2014, 28 (04) : 904 - 907
  • [5] Kinking of a Left-Sided Double-Lumen Tube Within the Trachea
    Choi, Dae-Kee
    Hwang, Jai-Hyun
    Song, Myung-Hee
    Hahm, Kyung-Don
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2011, 25 (06) : 1119 - 1120
  • [6] Bronchial blocker and double-lumen endotracheal tube combination to facilitate lung isolation during esophagectomy
    Liang, Hong
    Thomas, Mathew
    Renew, J. Ross
    ANNALS OF CARDIAC ANAESTHESIA, 2022, 25 (02) : 214 - 216
  • [7] A Randomized Comparison of Positional Stability: The EZ-Blocker Versus Left-Sided Double-Lumen Endobronchial Tubes in Adult Patients Undergoing Thoracic Surgery
    Morris, Benjamin N.
    Fernando, Rohesh J.
    Garner, Chandrika R.
    Johnson, Sean D.
    Gardner, Jeffrey C.
    Marchant, Bryan E.
    Johnson, Kathleen N.
    Harris, Hannah M.
    Russell, Gregory B.
    Wudel, L. James, Jr.
    Templeton, T. Wesley
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2021, 35 (08) : 2319 - 2325
  • [8] Use of the modified left-sided double-lumen tube in thoracic surgeries
    Ge, Jianling
    Zhou, Guoxia
    Peng, Lichao
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2017, 10 (03): : 5222 - 5227
  • [9] Bronchial blocker compared to double-lumen tube for one-lung ventilation during thoracoscopy
    Bauer, C
    Winter, C
    Hentz, JG
    Ducrocq, X
    Steib, A
    Dupeyron, JP
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2001, 45 (02) : 250 - 254
  • [10] A randomised comparison of the efficacy of a Coopdech bronchial blocker and a double-lumen endotracheal tube for minimally invasive esophagectomy
    Zhang, Tian-Hua
    Liu, Xiao-Qing
    Cao, Long-Hui
    Fu, Jian-Hua
    Lin, Wen-Qian