Denoising PET images for proton therapy using a residual U-net

被引:14
|
作者
Sano, Akira [1 ,2 ]
Nishio, Teiji [1 ,3 ]
Masuda, Takamitsu [1 ]
Karasawa, Kumiko [4 ]
机构
[1] Tokyo Womens Med Univ, Grad Sch Med, Dept Med Phys, Shinjuku Ku, 8-1 Kawadacho, Tokyo 1628666, Japan
[2] Mizuho Informat & Res Inst Inc, Chiyoda Ku, 2-3 Kanda Nishikicho, Tokyo 1018443, Japan
[3] Osaka Univ, Grad Sch Med, Div Hlth Sci, Med Phys Lab, 1-7 Yamadaoka, Suita, Osaka 5650871, Japan
[4] Tokyo Womens Med Univ, Sch Med, Dept Radiat Oncol, Shinjuku Ku, 8-1 Kawadacho, Tokyo 1628666, Japan
关键词
proton therapy; PET imaging; range verification; denoising; convolutional neural network; VERIFICATION;
D O I
10.1088/2057-1976/abe33c
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The use of proton therapy has the advantage of high dose concentration as it is possible to concentrate the dose on the tumor while suppressing damage to the surrounding normal organs. However, the range uncertainty significantly affects the actual dose distribution in the vicinity of the proton range, limiting the benefit of proton therapy for reducing the dose to normal organs. By measuring the annihilation gamma rays from the produced positron emitters, it is possible to obtain a proton induced positron emission tomography (pPET) image according to the irradiation region of the proton beam. Smoothing with a Gaussian filter is generally used to denoise PET images; however, this approach lowers the spatial resolution. Furthermore, other conventional smoothing processing methods may deteriorate the steep region of the pPET images. In this study, we proposed a denoising method based on a Residual U-Net for pPET images. We conducted the Monte Carlo simulation and irradiation experiment on a human phantom to obtain pPET data. The accuracy of the range estimation and the image similarity were evaluated for pPET images using the Residual U-Net, a Gaussian filter, a median filter, the block-matching and 3D-filtering (BM3D), and a total variation (TV) filter. Usage of the Residual U-Net yielded effective results corresponding to the range estimation; however, the results of peak-signal-to-noise ratio were identical to those for the Gaussian filter, median filter, BM3D, and TV filter. The proposed method can contribute to improving the accuracy of treatment verification and shortening the PET measurement time.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] A Residual Dense U-Net Neural Network for Image Denoising
    Gurrola-Ramos, Javier
    Dalmau, Oscar
    Alarcon, Teresa E.
    IEEE ACCESS, 2021, 9 : 31742 - 31754
  • [2] Elevator vibration signal denoising by deep residual U-Net
    Xie, Pengdong
    Zhang, Linxuan
    Li, Minghong
    Lau, Shing Fung Sean
    Huang, Jinhui
    MEASUREMENT, 2024, 225
  • [3] Comparing U-Net Based Models for Denoising Color Images
    Komatsu, Rina
    Gonsalves, Tad
    AI, 2020, 1 (04) : 465 - 486
  • [4] RatUNet: residual U-Net based on attention mechanism for image denoising
    Zhang, Huibin
    Lian, Qiusheng
    Zhao, Jianmin
    Wang, Yining
    Yang, Yuchi
    Feng, Suqin
    PEERJ COMPUTER SCIENCE, 2022, 8
  • [5] Image denoising based on Swin Transformer Residual Conv U-Net
    Gan, Yong
    Zhou, Shaohui
    Chen, Haonan
    Wang, Yuefeng
    27TH IEEE/ACIS INTERNATIONAL SUMMER CONFERENCE ON SOFTWARE ENGINEERING ARTIFICIAL INTELLIGENCE NETWORKING AND PARALLEL/DISTRIBUTED COMPUTING, SNPD 2024-SUMMER, 2024, : 166 - 170
  • [6] Fault Detection on Seismic Structural Images Using a Nested Residual U-Net
    Gao, Kai
    Huang, Lianjie
    Zheng, Yingcai
    IEEE TRANSACTIONS ON GEOSCIENCE AND REMOTE SENSING, 2022, 60
  • [7] An Automatic Nuclei Segmentation on Microscopic Images using Deep Residual U-Net
    Shree, H. P. Ramya
    Minavathi
    Dinesh, M. S.
    INTERNATIONAL JOURNAL OF ADVANCED COMPUTER SCIENCE AND APPLICATIONS, 2023, 14 (10) : 571 - 577
  • [8] Fault Detection on Seismic Structural Images Using a Nested Residual U-Net
    Gao, Kai
    Huang, Lianjie
    Zheng, Yingcai
    IEEE Transactions on Geoscience and Remote Sensing, 2022, 60
  • [9] U-Net for SPECT Image Denoising
    Reymann, Maximilian P.
    Wuerfl, Tobias
    Ritt, Philipp
    Stimpel, Bernhard
    Cachovan, Michal
    Vija, A. Hans
    Maier, Andreas
    2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,
  • [10] Mosaic Images Segmentation using U-net
    Fenu, Gianfranco
    Medvet, Eric
    Panfilo, Daniele
    Pellegrino, Felice Andrea
    ICPRAM: PROCEEDINGS OF THE 9TH INTERNATIONAL CONFERENCE ON PATTERN RECOGNITION APPLICATIONS AND METHODS, 2020, : 485 - 492