Fluid-Structure Interaction in Problems of Patient Specific Transcatheter Aortic Valve Implantation with and Without Paravalvular Leakage Complication

被引:7
作者
Basri, Adi Azriff [1 ,6 ]
Zuber, Mohammad [2 ]
Basri, Ernnie Illyani [1 ]
Zakaria, Muhammad Shukri [5 ]
Abd Aziz, Ahmad Fazli [3 ]
Tamagawa, Masaaki [4 ]
Ahmad, Kamarul Arifin [1 ,6 ]
机构
[1] Univ Putra Malaysia, Dept Aerosp Engn, Serdang 43400, Malaysia
[2] Manipal Acad Higher Educ, Manipal Inst Technol, Dept Aeronaut & Automobile Engn, Manipal 576104, Karnataka, India
[3] Univ Putra Malaysia, Dept Med, Serdang 43400, Malaysia
[4] Kyushu Inst Technol, Dept Biol Funct & Engn, Kitakyushu, Fukuoka, Japan
[5] Univ Tekn Malaysia Melaka, Fak Kejuruteraan Mekan, Hang Tuah Jaya 76100, Malaysia
[6] Univ Putra Malaysia, Aerosp Malaysia Res Ctr, Serdang 43400, Malaysia
来源
FDMP-FLUID DYNAMICS & MATERIALS PROCESSING | 2021年 / 17卷 / 03期
关键词
Paravalvular Leakage (PVL); hemodynamics; transcatheter aortic valve implantation (TAVI); fluid-structure interaction (FSI); edward sapien valve aortic valve (ESV); aortic stenosis (AS); WALL SHEAR-STRESS; BLOOD-FLOW; STABILITY ANALYSIS; NUMERICAL-METHODS; ASCENDING AORTA; HEART-VALVES; SIMULATION; DYNAMICS; STENOSIS; ASYMMETRY;
D O I
10.32604/fdmp.2021.010925
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Paravalvular Leakage (PVL) has been recognized as one of the most dangerous complications in relation to Transcathether Aortic Valve Implantation (TAVI) activities. However, data available in the literature about Fluid Structure Interaction (FSI) for this specific problem are relatively limited. In the present study, the fluid and structure responses of the hemodynamics along the patient aorta model and the aortic wall deformation are studied with the aid of numerical simulation taking into account PVL and 100% TAVI valve opening. In particular, the aorta without valve (AWoV) is assumed as the normal condition, whereas an aorta with TAVI 26 mm for 100% Geometrical Orifice Area (GOA) is considered as the patient aorta with PVL complication. A 3D patient-specific aorta model is elaborated using the MIMICS software. Implantation of the identical TAVI valve of Edward SAPIEN XT 26 (Edwards Lifes ciences, Irvine, California) is considered. An undersized 26 mm TAVI valve with 100% valve opening is selected to mimic the presence of PVL at the aortic annulus. The present research indicates that the existence of PVL can increase the blood velocity, pressure drop and WSS in comparison to normal conditions, thereby paving the way to the development of recirculation flow, thrombus formation, aorta wall collapse, aortic rupture and damage of endothelium.
引用
收藏
页码:531 / 553
页数:23
相关论文
共 74 条
  • [41] Two-Year Outcomes after Transcatheter or Surgical Aortic-Valve Replacement
    Kodali, Susheel K.
    Williams, Mathew R.
    Smith, Craig R.
    Svensson, Lars G.
    Webb, John G.
    Makkar, Raj R.
    Fontana, Gregory P.
    Dewey, Todd M.
    Thourani, Vinod H.
    Pichard, Augusto D.
    Fischbein, Michael
    Szeto, Wilson Y.
    Lim, Scott
    Greason, Kevin L.
    Teirstein, Paul S.
    Malaisrie, S. Chris
    Douglas, Pamela S.
    Hahn, Rebecca T.
    Whisenant, Brian
    Zajarias, Alan
    Wang, Duolao
    Akin, Jodi J.
    Anderson, William N.
    Leon, Martin B.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2012, 366 (18) : 1686 - 1695
  • [42] ONE-DIMENSIONAL STEADY INVISCID FLOW THROUGH A STENOTIC COLLAPSIBLE TUBE
    KU, DN
    ZEIGLER, MN
    DOWNING, JM
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1990, 112 (04): : 444 - 450
  • [43] WALL SHEAR STRESS IN A SUBJECT SPECIFIC HUMAN AORTA - INFLUENCE OF FLUID-STRUCTURE INTERACTION
    Lantz, Jonas
    Renner, Johan
    Karlsson, Matts
    [J]. INTERNATIONAL JOURNAL OF APPLIED MECHANICS, 2011, 3 (04) : 759 - 778
  • [44] Paravalvular Aortic Leak After Transcatheter Aortic Valve Replacement Current Knowledge
    Lerakis, Stamatios
    Hayek, Salim S.
    Douglas, Pamela S.
    [J]. CIRCULATION, 2013, 127 (03) : 397 - 407
  • [45] Blood flow in the human ascending aorta: a combined MRI and CFD study
    Leuprecht, A
    Kozerke, S
    Boesiger, P
    Perktold, K
    [J]. JOURNAL OF ENGINEERING MATHEMATICS, 2003, 47 (3-4) : 387 - 404
  • [46] On the Modeling of Patient-Specific Transcatheter Aortic Valve Replacement: A Fluid-Structure Interaction Approach
    Luraghi, Giulia
    Migliavacca, Francesco
    Garcia-Gonzalez, Alberto
    Chiastra, Claudio
    Rossi, Alexia
    Cao, Davide
    Stefanini, Giulio
    Rodriguez Matas, Jose Felix
    [J]. CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2019, 10 (03) : 437 - 455
  • [47] Luu Jane, 2013, JACC Cardiovasc Interv, V6, pe6, DOI 10.1016/j.jcin.2012.08.026
  • [48] Manimaran R., 2011, WORLD ACAD SCI ENG T, V73, P804
  • [49] Fluid-Structure Interaction Study of Transcatheter Aortic Valve Dynamics Using Smoothed Particle Hydrodynamics
    Mao, Wenbin
    Li, Kewei
    Sun, Wei
    [J]. CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2016, 7 (04) : 374 - 388
  • [50] Mark A, 2011, FLUID DYN MATER PROC, V7, P241