Repeatability and Reproducibility for Intraocular Pressure Measurement by Dynamic Contour, Ocular Response Analyzer, and Goldmann Applanation Tonometry

被引:37
作者
Sullivan-Mee, Michael [1 ]
Gerhardt, Gretchen
Halverson, Kathy D.
Qualls, Clifford [2 ]
机构
[1] Albuquerque VA Med Ctr, San Pedro SE, Albuquerque, NM USA
[2] Univ New Mexico, Sch Med, Dept Biostat, Albuquerque, NM 87131 USA
关键词
intraocular pressure; repeatability; reproducibility; Goldmann applanation tonometry; Pascal dynamic contour tonometry; ocular response analyzer tonometry; CENTRAL CORNEAL THICKNESS; BIOMECHANICAL PROPERTIES; HYSTERESIS; GLAUCOMA; PARAMETERS; AGREEMENT; EYES; AGE;
D O I
10.1097/IJG.0b013e31819c487d
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate and compare the intraocular pressure measurement variability between Goldmann applanation tonometry (GAT), Pascal dynamic contour tonometry (DCT), and ocular response analyzer (ORA) tonometry. Methods: Subjects were prospectively recruited from consecutive Albuquerque VA Medical Center eye clinic patients that were previously diagnosed with ocular hypertension, glaucoma suspect. primary open-angle, or normal pressure glaucoma. Two sets of intraocular pressure measurements (3-4 ORA, 2 DCT, and 2 GAT) were obtained approximately 15 minutes apart. Each set was obtained by 1 of 2 examiners using random examiner sequences. ORA was measured first in both eyes, followed by either DCT or GAT, which were obtained in random order. Repeatability was assessed by examining variability of consecutive measurements by the same examiner, and reproducibility was examined by assessing variability between the first and second measurement sets. Results: One hundred and twenty eyes of 60 subjects were included. Mean age was 64.1 +/- 9.6 years and 58/60 (97%) were male. Intraobserver repeatability was highest for GAT, followed closely by DCT, and then ORA. Intersession reproducibility was similar for all methods, although a tonographic effect may have corrupted GAT and DCT reproducibility results. We found no repeatability or reproducibility differences between eyes, between examiners. or between measurement sets. Conclusions: Although some intermethod variability differences were identified, all 3 methods in this study demonstrated clinically acceptable measurement repeatability and reproducibility. This result, in conjunction with the finding that variability was not different between eyes, examiners, or measurement sets, suggests that DCT and ORA are reliable enough to be clinically useful.
引用
收藏
页码:666 / 673
页数:8
相关论文
共 37 条
[1]  
Bland JM, 1996, BRIT MED J, V313, P744
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Lower corneal hysteresis in glaucoma patients with acquired pit of the optic nerve (APON) [J].
Bochmann, Frank ;
Ang, Ghee S. ;
Azuara-Blanco, Augusto .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2008, 246 (05) :735-738
[4]   Dynamic contour tonometry in comparison to intracameral IOP measurements [J].
Boehm, Andreas G. ;
Weber, Anja ;
Pillunat, Lutz E. ;
Koch, Rainer ;
Spoerl, Eberhard .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2008, 49 (06) :2472-2477
[5]   Central corneal thickness and measured IOP response to topical ocular hypotensive medication in the ocular hypertension treatment study [J].
Brandt, JD ;
Beiser, JA ;
Gordon, MO ;
Kass, MA .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2004, 138 (05) :717-722
[6]   Influence of corneal structure, corneal responsiveness, and other ocular parameters on tonometric measurement of Intraocular pressure [J].
Broman, Aimee Teo ;
Congdon, Nathan G. ;
Bandeen-Roche, Karen ;
Quigley, Harry A. .
JOURNAL OF GLAUCOMA, 2007, 16 (07) :581-588
[7]  
CHENG NF, SAS MACRO COMPUTE LI
[8]   Central corneal thickness and corneal hysteresis associated with glaucoma damage [J].
Congdon, NG ;
Broman, AT ;
Bandeen-Roche, K ;
Grover, D ;
Quigley, HA .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2006, 141 (05) :868-875
[9]   RELIABILITY OF INTRAOCULAR-PRESSURE MEASUREMENT WITH THE GOLDMANN APPLANATION TONOMETER IN EPIDEMIOLOGIC STUDIES [J].
DIELEMANS, I ;
VINGERLING, JR ;
HOFMAN, A ;
GROBBEE, DE ;
DEJONG, PTVM .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1994, 232 (03) :141-144
[10]   Comparison of dynamic contour tonometry with Goldman applanation tonometry over a wide range of central corneal thickness [J].
Doyle, A ;
Lachkar, Y .
JOURNAL OF GLAUCOMA, 2005, 14 (04) :288-292