Chest radiography: Optimization of X-ray spectrum for cesium iodide-amorphous silicon flat-panel detector

被引:59
|
作者
Dobbins, JT
Samei, E
Chotas, HG
Warp, RJ
Baydush, AH
Floyd, CE
Ravin, CE
机构
[1] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
[2] Duke Univ, Med Ctr, Dept Biomed Engn, Durham, NC 27710 USA
关键词
flat panel detector; radiography; digital;
D O I
10.1148/radiol.2261012023
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To ascertain the optimum x-ray spectrum for chest radiography with a cesium iodide-amorphous silicon flat-panel detector. MATERIALS AND METHODS: End points for optimization included the ratio of tissue contrast to bone contrast and a figure of merit (FOM) equal to the square of the signal-to-noise ratio of tissue divided by incident exposure to the patient. Studies were conducted with both computer spectrum modeling and experimental measurement in narrow-beam and full-field exposure conditions for four tissue thicknesses (8-32 cm). Three parameters that affect spectra were considered: the atomic number (Z) of filter material (Z = 13, 26, 29, 42, 50, 56, 64, 74, and 82), kilovoltage (from 50 to 150 kVp), and filter thickness (from 0.25 to 2.00 half-value layer [HVL]). RESULTS: Computer modeling and narrow-beam experimental data showed similar trends for the full range of parameters evaluated. Spectrum model results showed that copper filtration at 120 kVp or more was optimum for FOM. The ratio of contrasts showed a trend to be higher with higher kilovoltage and only a minor variation with filter material. Full-field experimental results, which reflect the added contribution of x-ray scatter, differed in magnitude but not trends from the narrow-beam data in all cases except the ratio of contrasts in the mediastinum. CONCLUSION: The best performance overall, including both FOM and ratio of contrasts, was at 120 kVp with 1-HVL copper filtration (0.2 mm). With this beam spectrum and an increase in tube output (ie, milliampere seconds) of about 50%, a chest radiograph can be obtained with image quality approximately equal to that with a conventional spectrum but with about 25% less patient exposure. (C) RSNA, 2002.
引用
收藏
页码:221 / 230
页数:10
相关论文
共 50 条
  • [1] Contrast-detail phantom study for X-ray spectrum optimization regarding chest radiography using a cesium iodide -: Amorphous silicon flat-panel detector
    Hamer, OW
    Völk, M
    Zorger, N
    Borisch, I
    Büittner, R
    Feuerbach, S
    Strotzer, M
    INVESTIGATIVE RADIOLOGY, 2004, 39 (10) : 610 - 618
  • [2] Imaging characteristics of an amorphous silicon flat-panel detector for digital chest radiography
    Floyd, CE
    Warp, RJ
    Dobbins, JT
    Chotas, HG
    Baydush, AH
    Vargas-Voracek, R
    Ravin, CE
    RADIOLOGY, 2001, 218 (03) : 683 - 688
  • [3] Characterization of a flat-panel amorphous silicon detector for X-ray computed tomography
    Lee, H.-R.
    Ellingson, W.
    Journal of X-Ray Science and Technology, 2001, 9 (01) : 43 - 53
  • [4] Digital chest radiography with a large-area flat-panel silicon X-ray detector: clinical comparison with conventional radiography
    Svenja P. Hennigs
    Marietta Garmer
    Horst J. Jaeger
    Reinhard Classen
    Andreas Jacobs
    Hans M. Gissler
    Andreas Christmann
    Klaus Mathias
    European Radiology, 2001, 11 : 1688 - 1696
  • [5] Digital chest radiography with a large-area flat-panel silicon X-ray detector: clinical comparison with conventional radiography
    Hennigs, SP
    Garmer, M
    Jaeger, HJ
    Classen, R
    Jacobs, A
    Gissler, HM
    Christmann, A
    Mathias, K
    EUROPEAN RADIOLOGY, 2001, 11 (09) : 1688 - 1696
  • [6] Digital radiography with a large-area, amorphous-silicon, flat-panel X-ray detector system
    Spahn, M
    Strotzer, M
    Völk, M
    Böhm, S
    Geiger, B
    Hahm, G
    Feuerbach, S
    INVESTIGATIVE RADIOLOGY, 2000, 35 (04) : 260 - 266
  • [7] Radiation dose reduction in chest radiography using a flat-panel amorphous silicon detector
    Hosch, WP
    Fink, C
    Radeleff, B
    Kampschulte, A
    Kauffmann, GW
    Hansmann, J
    CLINICAL RADIOLOGY, 2002, 57 (10) : 902 - 907
  • [8] Clinical evaluation of digital radiography based on a large-area cesium iodide-amorphous silicon flat-panel detector compared with screen-film radiography for skeletal system and abdomen
    Okamura, T
    Tanaka, S
    Koyama, K
    Norihumi, N
    Daikokuya, H
    Matsuoka, T
    Kishimoto, K
    Hatagawa, M
    Kudoh, H
    Yamada, R
    EUROPEAN RADIOLOGY, 2002, 12 (07) : 1741 - 1747
  • [9] Clinical evaluation of digital radiography based on a large-area cesium iodide-amorphous silicon flat-panel detector compared with screen-film radiography for skeletal system and abdomen
    Terue Okamura
    Saori Tanaka
    Koichi Koyama
    Nishida Norihumi
    Hideo Daikokuya
    Toshiyuki Matsuoka
    Kenji Kishimoto
    Masakatsu Hatagawa
    Hiroaki Kudoh
    Ryusaku Yamada
    European Radiology, 2002, 12 : 1741 - 1747
  • [10] Dynamic chest radiography: flat-panel detector (FPD) based functional X-ray imaging
    Tanaka R.
    Radiological Physics and Technology, 2016, 9 (2) : 139 - 153