Comparison of radiation doses to patients undergoing standard radiographic examinations with conventional screen-film radiography, computed radiography and direct digital radiography

被引:33
作者
Compagnone, G.
Baleni, M. Casad
Pagan, L.
Calzolaio, F. L.
Barozzi, L.
Bergamini, C.
机构
[1] St Orsola Marcello Malpighi Hosp, Dept Med Phys, I-40138 Bologna, Italy
[2] St Orsola Marcello Malpighi Hosp, Dept Accid & Emergency, I-40138 Bologna, Italy
关键词
D O I
10.1259/bjr/57138583
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
New flat-panel direct digital radiography equipment has recently been installed in our Accident and Emergency Department; its characteristics and versatility are well suited to the work undertaken in this environment. The aim of this study was to compare radiation doses to patients undergoing standard radiographic,examinations using conventional screen-film radiography, computed radiography and direct digital radiography; entrance surface dose and effective dose were calculated for six standard examinations (a total of 10 projections) using standard patient exposure parameters for the three imaging modalities. It was found that doses for computed radiography (all examinations) were higher than the doses for the other two modalities; effective doses for direct digital radiography were similar to 29% and similar to 43% lower than those for screen-film radiography and computed radiography, respectively. The image quality met the criteria in the European guidelines for all modalities.
引用
收藏
页码:899 / 904
页数:6
相关论文
共 23 条
[1]  
[Anonymous], 1991, Annals of the ICRP, V21
[2]  
BOGUCKI TM, 1995, TECHNICAL SCI MONOGR, V6
[3]   Experiences with computed radiography: Can we afford the cost? [J].
Bragg, DG ;
Murray, KA ;
Tripp, D .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1997, 169 (04) :935-941
[4]   An assessment of the usefulness of screen-film speed classifications [J].
Brennan, PC ;
Devereux, SA .
EUROPEAN RADIOLOGY, 2002, 12 (06) :1577-1583
[5]   Local diagnostic reference levels in standard X-ray examinations [J].
Compagnone, G ;
Pagan, L ;
Bergamini, C .
RADIATION PROTECTION DOSIMETRY, 2005, 113 (01) :54-63
[6]  
COX R, 1993, DOCUMENTS NATL RADIO, V4
[7]  
*EUR COMM, 1996, 16260EN EUR EUR COMM
[8]   Chest radiographs in the emergency department: is the radiologist really necessary? [J].
Gatt, ME ;
Spectre, G ;
Paltiel, O ;
Hiller, N ;
Stalnikowicz, R .
POSTGRADUATE MEDICAL JOURNAL, 2003, 79 (930) :214-217
[9]   IMAGE QUALITY AND PATIENT DOSE FOR DIFFERENT SCREEN-FILM COMBINATIONS [J].
GUIBELALDE, E ;
FERNANDEZ, JM ;
LLORCA, A ;
RUIZ, MJ .
BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (794) :166-173
[10]   A mathematical spread sheet application for production of entrance skin exposure nomograms [J].
Harpen, MD .
MEDICAL PHYSICS, 1996, 23 (02) :241-242