The effect of sliding onto the metal-electrolyte interface: Studying model parameter modifications by means of EIS

被引:4
作者
Cassar, Josianne [1 ]
Mallia, Bertram [1 ]
Karl, Andreas [2 ]
Buhagiar, Joseph [1 ]
机构
[1] Univ Malta, Dept Met & Mat Engn, MSD-2080 Msida, Malta
[2] Bodycote Specialist Technol GrnbH, Landsberg, Germany
来源
MATERIALS SCIENCE AND ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS | 2017年 / 75卷
关键词
EIS; Tribocorrosion; CoCrMo alloy; S-phase; Carburizing; Expanded austenite; Cobalt-chromium alloy; ASTM F1537; Biomedical; Orthopaedic; SIMULATED BODY-FLUIDS; TICXOY THIN-FILMS; TRIBOCORROSION BEHAVIOR; S-PHASE; CO-CR; CORROSION; ALLOYS;
D O I
10.1016/j.msec.2017.03.048
中图分类号
TB3 [工程材料学]; R318.08 [生物材料学];
学科分类号
0805 ; 080501 ; 080502 ;
摘要
Several problems are associated with corrosion-wear occurring on metal-on-metal hip implants made out of cobalt-chromium based alloys. Low temperature carburizing, a process that creates a hard and corrosion resistant diffused layer in Cobalt-Chromium-Molybdenum (CoCrMo) alloys, known as S-phase, may be a possible solution towards mitigating these problems. In this work, static- and tribo-corrosion testing involving an alumina versus CoCrMo (untreated and carburized) were conducted in Ringer's solution. Electrochemical impedance spectroscopy was used to compare impedance plots attained before and after sliding so as to understand how the metal electrolyte interface is affected by rubbing. Both untreated and carburized CoCrMo experienced extensive reduction in corrosion resistance following sliding wear damage such that one should expect a considerably deteriorated performance of both surfaces in a tribocorrosion application. The structure of the interface was relatively unaffected after sliding at the equilibrium and passive potentials. This implies that the layers making up the interface before sliding were still present after sliding. However, their properties changed- the interface's real resistance dropped while its capacitance increased. The former was linked to a weaker, damaged passive film while the latter was linked to accumulation of wear debris and corrosion products. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:1366 / 1375
页数:10
相关论文
共 27 条
[21]  
Orazem M. E., 2008, ELECTROCHEMICAL IMPE
[22]   Failure rates of stemmed metal-on-metal hip replacements: analysis of data from the National Joint Registry of England and Wales [J].
Smith, Alison J. ;
Dieppe, Paul ;
Vernon, Kelly ;
Porter, Martyn ;
Blom, Ashley W. .
LANCET, 2012, 379 (9822) :1199-1204
[23]  
Tait W S, 1994, An Introduction to Electrochemical Corrosion Testing for Practicing Engineers and Scientists
[24]   Effect of physico-chemical properties of simulated body fluids on the electrochemical behaviour of CoCrMo alloy [J].
Valero Vidal, C. ;
Igual Munoz, A. .
ELECTROCHIMICA ACTA, 2011, 56 (24) :8239-8248
[25]  
Vidal CV, 2011, BIOMEDICAL ENGINEERING, TRENDS IN MATERIALS SCIENCE, P283
[26]   Electrochemical characterisation of biomedical alloys for surgical implants in simulated body fluids [J].
Vidal, C. Valero ;
Munoz, A. Igual .
CORROSION SCIENCE, 2008, 50 (07) :1954-1961
[27]   Effect of thermal treatment and applied potential on the electrochemical behaviour of CoCrMo biomedical alloy [J].
Vidal, C. Valero ;
Munoz, A. Igual .
ELECTROCHIMICA ACTA, 2009, 54 (06) :1798-1809