Effective dose and image quality for intraoperative imaging with a cone-beam CT and a mobile multi-slice CT in spinal surgery: A phantom study

被引:9
作者
Casiraghi, Margherita [1 ]
Scarone, Pietro [2 ]
Bellesi, Luca [1 ]
Piliero, Maria Antonietta [1 ]
Pupillo, Francesco [1 ]
Gaudino, Diego [1 ]
Fumagalli, Giulia [1 ]
Del Grande, Filippo [3 ]
Presilla, Stefano [1 ]
机构
[1] Osped Reg Bellinzona & Valli, Med Phys Div, Imaging Inst Southern Switzerland, Ente Osped Cantonale, CH-6500 Bellinzona, Switzerland
[2] Neuroctr Southern Switzerland, Serv Neurosurg, CH-6900 Lugano, Switzerland
[3] Osped Reg Lugano, Dept Radiol, Imaging Inst Southern Switzerland, Ente Osped Cantonale, CH-6900 Lugano, Switzerland
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2021年 / 81卷
关键词
CT effective dose; Intraoperative imaging; Image quality; O-arm CBCT; Airo MSCT; COMPUTED-TOMOGRAPHY; O-ARM; DOSIMETRIC CHARACTERIZATION; RADIATION-EXPOSURE; NAVIGATION; SYSTEM; INSTRUMENTATION; ASSURANCE; GUIDANCE; PATIENT;
D O I
10.1016/j.ejmp.2020.11.006
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare the effective dose (ED) and image quality (IQ) of O-arm cone-beam CT (Medtronic, Min-neapolis, MN, USA) and Airo multi-slice CT (Brainlab AG, Munich, Germany) for intraoperative-CT (i-CT) in spinal surgery. Methods: The manufacturer-defined protocols available in the O-arm and Airo systems for three-dimensional lumbar spine imaging were compared. Organ dose was measured both with thermo-luminescent dosimeters and GafChromic films in the Alderson Radiation Therapy anthropomorphic phantom. A subjective analysis was performed by neurosurgeons to compare the clinical IQ of the anthropomorphic phantom images acquired with the different i-CT systems and imaging protocols. Image uniformity, noise, contrast-to-noise-ratio (CNR), and spatial resolution were additionally assessed with the Catphan 504 phantom. Results: O-arm i-CT caused 56% larger ED than Aim due to the high definition (HD) imaging protocol. The noise was larger for O-arm images leading to a lower CNR than that measured for Airo. Moreover, scattering and beam hardening effects were observed in the O-arm images. Better spatial resolution was measured for the O-arm system (9 lp/cm) than for Airo (4 lp/cm). For all the investigated protocols, O-arm was found to be better for identifying anatomical features important for accurate pedicle screw positioning. Conclusions: According to phantom measurements, the HD protocol of O-arm offered better clinical IQ than Airo but larger ED. The larger noise of O-arm images did not compromise the clinical IQ while the superior spatial resolution of this system allowed a better visibility of anatomical features important for pedicle screw positioning in the lumbar region.
引用
收藏
页码:9 / 19
页数:11
相关论文
共 40 条
[1]   Optimization of Radiation Exposure and Image Quality of the Cone-beam O-arm Intraoperative Imaging System in Spinal Surgery [J].
Abul-Kasim, Kasim ;
Soderberg, Marcus ;
Selariu, Eufrozina ;
Gunnarsson, Mikael ;
Kherad, Mehrsa ;
Ohlin, Acke .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2012, 25 (01) :52-58
[2]  
[Anonymous], 1977, Annals of the ICRP, V1
[3]   A quality assurance program for image quality of cone-beam CT guidance in radiation therapy [J].
Bissonnette, Jean-Pierre ;
Moseley, Douglas J. ;
Jaffray, David A. .
MEDICAL PHYSICS, 2008, 35 (05) :1807-1815
[4]   Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179 [J].
Bissonnette, Jean-Pierre ;
Balter, Peter A. ;
Dong, Lei ;
Langen, Katja M. ;
Lovelock, D. Michael ;
Miften, Moyed ;
Moseley, Douglas J. ;
Pouliot, Jean ;
Sonke, Jan-Jakob ;
Yoo, Sua .
MEDICAL PHYSICS, 2012, 39 (04) :1946-1963
[5]   Implementation of radiochromic film dosimetry protocol for volumetric dose assessments to various organs during diagnostic CT procedures [J].
Brady, Samuel ;
Yoshizumi, Terry ;
Toncheva, Greta ;
Frush, Donald .
MEDICAL PHYSICS, 2010, 37 (09) :4782-4792
[6]   Computed Tomography-Based Image-Guided System in Spinal Surgery: State of the Art Through 10 Years of Experience [J].
Costa, Francesco ;
Dorelli, Gianluigi ;
Ortolina, Alessandro ;
Cardia, Andrea ;
Attuati, Luca ;
Tomei, Massimo ;
Milani, Davide ;
Balzarini, Luca ;
Galbusera, Fabio ;
Morenghi, Emanuela ;
Fornari, Maurizio .
OPERATIVE NEUROSURGERY, 2015, 11 (01) :59-67
[7]   Spinal Navigation: Standard Preoperative Versus Intraoperative Computed Tomography Data Set Acquisition for Computer-Guidance System Radiological and Clinical Study in 100 Consecutive Patients [J].
Costa, Francesco ;
Cardia, Andrea ;
Ortolina, Alessandro ;
Fabio, Galbusera ;
Zerbi, Alberto ;
Fornari, Maurizio .
SPINE, 2011, 36 (24) :2094-2098
[8]   Spinal navigation for posterior instrumentation of C1-2 instability using a mobile intraoperative CT scanner [J].
Czabanka, Marcus ;
Haemmerli, Julien ;
Hecht, Nils ;
Foehre, Bettina ;
Arden, Klaus ;
Liebig, Thomas ;
Woitzik, Johannes ;
Vajkoczy, Peter .
JOURNAL OF NEUROSURGERY-SPINE, 2017, 27 (03) :268-275
[9]   Prospective Comparative Study in Spine Surgery Between O-Arm and Airo Systems: Efficacy and Radiation Exposure [J].
Farah, Kaissar ;
Coudert, Pierre ;
Graillon, Thomas ;
Blondel, Benjamin ;
Dufour, Henry ;
Gille, Olivier ;
Fuentes, Stephane .
WORLD NEUROSURGERY, 2018, 118 :E175-E184
[10]   Organ doses evaluation for chest computed tomography procedures with TL dosimeters: Comparison with Monte Carlo simulations [J].
Giansante, Louise ;
Martins, Juliana C. ;
Nersissian, Denise Y. ;
Kiers, Karen C. ;
Kay, Fernando U. ;
Sawamura, Marcia V. Y. ;
Lee, Choonsik ;
Gebrim, Eloisa M. M. S. ;
Costa, Paulo R. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (01) :308-320