Detection issues in two-dimensional on-line chromatography

被引:32
作者
Horvath, Krisztian [1 ,2 ]
Fairchild, Jacob N. [1 ]
Guiochon, Georges [1 ]
机构
[1] Univ Tennessee, Dept Chem, Knoxville, TN 37996 USA
[2] Univ Pannonia, Dept Analyt Chem, H-8200 Veszprem, Hungary
关键词
Two-dimensional chromatography; On-line two-dimensional chromatography; Detection; Detection limit; Dilution factor; LIQUID-CHROMATOGRAPHY; GRADIENT ELUTION; LIMIT; SEPARATIONS;
D O I
10.1016/j.chroma.2009.09.016
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The dilution of analytes during their migration through two-dimensional liquid chromatographic systems was investigated. Simplified equations for the calculation of the dilution factors in the first and second dimension columns were derived considering the variance of the injected sample, the flow rates, the split ratio, the volume of the intermediate sample loop and the sampling time of the first-dimension effluent. The net dilution factor is the product of the dilution factors in the two single dimensions. It is between 200 anti 300 in typical two-dimensional liquid chromatographic (2D-LC) separations. These values are less pessimistic than those reported previously [1]. It was shown that the fraction of organic modifier or stronger eluent component in the injected sample has a significant effect on the widths and heights of the eluted peaks. Analytes with high retention factors and high molecular weights are less sensitive toward this effect than those having small retention factors. These results suggest that the optimization of the experimental conditions of two dimension separations must be made from the points of view of both the retention factor and the detector response factor. The frequency of sampling of the eluent of the first-dimension separation does not have a significant effect on the detection limit of a 2D system if the split ratio is adjusted in the ratio of the frequency of sample collection. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:7785 / 7792
页数:8
相关论文
共 15 条
[1]  
Dolan JW, 2006, LC GC EUR, V19, P12
[2]   Approaches to comprehensive multidimensional liquid chromatography systems [J].
Fairchild, Jacob N. ;
Horvath, Krisztian ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2009, 1216 (09) :1363-1371
[3]   CLARIFICATION OF THE LIMIT OF DETECTION IN CHROMATOGRAPHY [J].
FOLEY, JP ;
DORSEY, JG .
CHROMATOGRAPHIA, 1984, 18 (09) :503-511
[4]   Exact peak compression factor in linear gradient elution I. Theory [J].
Gritti, Fabrice ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2008, 1212 (1-2) :35-40
[5]   Implementations of two-dimensional liquid chromatography [J].
Guiochon, Georges ;
Marchetti, Nicola ;
Mriziq, Khaled ;
Shalliker, R. Andrew .
JOURNAL OF CHROMATOGRAPHY A, 2008, 1189 (1-2) :109-168
[6]  
Jandera P., 1985, Gradient Elution in Column Liquid Chromatography
[7]   ROLE OF COLUMN PARAMETERS AND INJECTION VOLUME ON DETECTION LIMITS IN LIQUID-CHROMATOGRAPHY [J].
KARGER, BL ;
MARTIN, M ;
GUIOCHON, G .
ANALYTICAL CHEMISTRY, 1974, 46 (12) :1640-1647
[8]   Comprehensive off-line, two-dimensional liquid chromatography. Application to the separation of peptide digests [J].
Marchetti, Nicola ;
Fairchild, Jacob N. ;
Guiochon, Georges .
ANALYTICAL CHEMISTRY, 2008, 80 (08) :2756-2767
[9]   OSCILLOMETRIC DETECTOR FOR ION CHROMATOGRAPHY - A NOTE ON DETECTION LIMIT AND DETECTOR SENSITIVITY [J].
PAL, F ;
PUNGOR, E ;
KOVATS, ES .
ANALYTICAL CHEMISTRY, 1988, 60 (20) :2254-2258
[10]   PEAK WIDTH IN SOLVENT-PROGRAMMED CHROMATOGRAPHY .1. GENERAL DESCRIPTION OF PEAK BROADENING IN SOLVENT-PROGRAMMED ELUTION [J].
POPPE, H ;
PAANAKKER, J ;
BRONCKHORST, M .
JOURNAL OF CHROMATOGRAPHY, 1981, 204 (JAN) :77-84