Investigations on exponential lifetime measurements for fluorescence thermometry

被引:20
作者
Fernicola, VC [1 ]
Rosso, L
Galleano, R
Sun, T
Zhang, ZY
Grattan, KTV
机构
[1] CNR, Ist Metrol G Colonnetti, I-10126 Turin, Italy
[2] City Univ London, London EC1V 0MB, England
关键词
D O I
10.1063/1.1150714
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Lifetime-based methods have been, on the whole, one of the most successful schemes for fiber optic temperature sensing, using fluorescent materials whose response is intensity independent. Several approaches for determining the fluorescence lifetime, and with that the measurand, have been investigated. An experimental comparison of direct and indirect measurement methods, i.e., involving actual signals from representative optical media instead of simply using Monte Carlo simulations, has been carried out. Direct fitting methods, including Marquardt, log-fit and Prony, were used to estimate the fluorescence lifetime of a Cr3+:YAG-based sensor system and the results were compared. An agreement to better than 0.5% between Marquardt and log-fit algorithms and an agreement of about 1.5% between Marquardt and Prony approaches was found. Thus, a temperature reproducibility, of 0.5 and 1.2 degrees C, respectively, can be obtained with the Cr3+:YAG sensor system. An indirect measurement approach based on a phase-locked (analog-to-digital signal processor) (A-DSP) was also tested. It was found that when the A-DSP output is used to estimate the lifetime, it performs only slightly better than using direct fitting methods. On the contrary, when the whole A-DSP sensor system was directly calibrated against temperature, the measurement accuracy improves by at least a factor of 10. (C) 2000 American Institute of Physics. [S0034-6748(00)01807-4].
引用
收藏
页码:2938 / 2943
页数:6
相关论文
共 9 条
[1]   PRECISION LIMITS OF LIFETIME ESTIMATION ALGORITHMS AS DETERMINED BY MONTE-CARLO SIMULATION - A COMPARISON OF THEORY AND EXPERIMENT [J].
DOWELL, LJ ;
GILLIES, GT .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (08) :1310-1315
[2]   ERRORS CAUSED BY BASE-LINE OFFSET AND NOISE IN THE ESTIMATION OF EXPONENTIAL LIFETIMES [J].
DOWELL, LJ ;
GILLIES, GT .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (01) :242-243
[3]  
FERNICOLA V, 1993, P SOC PHOTO-OPT INS, V2070, P472
[4]   DIGITAL SIGNAL-PROCESSING FOR FIBEROPTIC THERMOMETERS [J].
FERNICOLA, VC ;
CROVINI, L .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1995, 44 (02) :447-450
[5]  
Grattan K. T. V., 1995, FIBER OPTIC FLUORESC
[6]   Exponential analysis in physical phenomena [J].
Istratov, AA ;
Vyvenko, OF .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (02) :1233-1257
[7]   AN ALGORITHM FOR LEAST-SQUARES ESTIMATION OF NONLINEAR PARAMETERS [J].
MARQUARDT, DW .
JOURNAL OF THE SOCIETY FOR INDUSTRIAL AND APPLIED MATHEMATICS, 1963, 11 (02) :431-441
[8]   Prony's method for exponential lifetime estimations in fluorescence-based thermometers [J].
Zhang, ZY ;
Grattan, KTV ;
Hu, YL ;
Palmer, AW ;
Meggitt, BT .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1996, 67 (07) :2590-2594
[9]   Spectral characteristics and effects of heat treatment on intrinsic Nd-doped fiber thermometer probes [J].
Zhang, ZY ;
Grattan, KTV ;
Palmer, AW ;
Meggitt, BT .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (01) :139-145