Implantation techniques and chronic lead parameters of biventricular pacing dual-chamber defibrillators

被引:67
作者
Daoud, EG [1 ]
Kalbfleisch, SJ [1 ]
Hummel, JD [1 ]
Weiss, R [1 ]
Augustini, RS [1 ]
Duff, SB [1 ]
Polsinelli, G [1 ]
Castor, J [1 ]
Meta, T [1 ]
机构
[1] Riverside Methodist Hosp, MidWest Res Fdn, Columbus, OH USA
关键词
heartfailure; pacemaker; defibrillator; resynchronization; left ventricle; lead;
D O I
10.1046/j.1540-8167.2002.00964.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: The aim of this study is to describe implantation techniques and lead performance for biventricular pacing, dual-chamber implantable cardioverter defibrillators (ICDs). Methods and Results: A dual-chamber ICD with biventricular pacing was implanted in 87 patients with congestive heart failure (ejection fraction: 0.21 +/- 0.09), prolonged QRS duration (161 +/- 22 msec), and an indication for ICD therapy. Left ventricular pacing was achieved with a thoracotomy approach (n = 21) or a nonthoracotomy approach (n = 66). With a thoracotomy, biventricular devices were implanted successfully in all patients. During follow-up (17 +/- 11 months), 9 patients died (43%), 2 underwent transplantation, and 2 required left ventricular lead revision. At last follow-up, biventricular sensing and capture threshold were 11 +/- 5 mV and 1.5 +/- 0.8 V, respectively. For nonthoracotomy procedures, two types of coronary sinus (CS) leads were implanted: an over-the-wire lead (n = 45) and a shaped lead (n = 21). The rate of successful implantation (overall: 89%) (over-the-wire 93% vs shaped 81%; P = 0.1) and durations for CS lead placement (66 +/- 50 vs 58 +/- 34 min, P = 0.6) and the procedure (133 +/- 58 vs 129 +/- 33 min, P = 0.8) were not different between the two CS leads. During follow-up (11 9 months), 9 patients died (14%), and the shaped CS lead dislodged in 3 patients (3 shaped vs 0 over-the-wire, P = 0.01). At last follow-up, biventricular sensing and capture threshold were 10 +/- 4 mV and 1.8 +/- 0.7 V, respectively, and there was no difference between over-the-wire and shaped leads. By multivariate analysis, mortality was associated with absence of spironolactone therapy but not procedural features. Conclusion: Nonthoracotomy CS lead implantation is feasible, with a success rate of about 90% and few adverse events. For the remaining 10%, a thoracotomy approach can be completed safely in these ill patients without increased risk for death.
引用
收藏
页码:964 / 970
页数:7
相关论文
共 24 条
[1]  
Abraham WT, 2002, NEW ENGL J MED, V346, P1902
[2]  
Achtelik M, 2000, PACE, V23, P1741
[3]   High failure rate for an epicardial implantable cardioverter-defibrillator lead: Implications for long-term follow-up of patients with an implantable cardioverter-defibrillator [J].
Brady, PA ;
Friedman, PA ;
Trusty, JM ;
Grice, S ;
Hammill, SC ;
Stanton, MS .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (03) :616-622
[4]   Echocardiographic evidence of a hemodynamic and clinical improvement in patients paced for heart failure [J].
Breithardt, OA ;
Stellbrink, C ;
Franke, A ;
Auricchio, A ;
Huvelle, E ;
Sack, S ;
Bakker, P ;
Hanrath, P .
AMERICAN JOURNAL OF CARDIOLOGY, 2000, 86 (9A) :133K-137K
[5]   Effects of multisite biventricular pacing in patients with heart failure and intraventricular conduction delay. [J].
Cazeau, S ;
Leclercq, C ;
Lavergne, T ;
Walker, S ;
Varma, C ;
Linde, C ;
Garrigue, S ;
Kappenberger, L ;
Haywood, GA ;
Santini, M ;
Bailleul, C ;
Daubert, JC .
NEW ENGLAND JOURNAL OF MEDICINE, 2001, 344 (12) :873-880
[6]   Initial clinical experience with ambulatory use of an implantable atrial defibrillator for conversion of atrial fibrillation [J].
Daoud, EG ;
Timmermans, C ;
Fellows, C ;
Hoyt, R ;
Lemery, R ;
Dawson, K ;
Ayers, GM .
CIRCULATION, 2000, 102 (12) :1407-1413
[7]   INCIDENCE, PRESENTATION, DIAGNOSIS, AND MANAGEMENT OF MALFUNCTIONING IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR RATE-SENSING LEADS [J].
DAOUD, EG ;
KIRSH, MM ;
BOLLING, SF ;
STRICKBERGER, SA ;
MORADY, F ;
KOU, WH .
AMERICAN HEART JOURNAL, 1994, 128 (05) :892-895
[8]   FUNCTIONAL ABNORMALITIES IN ISOLATED LEFT-BUNDLE BRANCH-BLOCK - THE EFFECT OF INTERVENTRICULAR ASYNCHRONY [J].
GRINES, CL ;
BASHORE, TM ;
BOUDOULAS, H ;
OLSON, S ;
SHAFER, P ;
WOOLEY, CF .
CIRCULATION, 1989, 79 (04) :845-853
[9]   Biventricular pacing decreases sympathetic activity compared with right ventricular pacing in patients with depressed ejection fraction [J].
Hamdan, MH ;
Zagrodzky, JD ;
Joglar, JA ;
Sheehan, CJ ;
Ramaswamy, K ;
Erdner, JF ;
Page, RL ;
Smith, ML .
CIRCULATION, 2000, 102 (09) :1027-1032
[10]   Left heart pacing - Experience with several types of coronary vein leads [J].
Hansky, B ;
Vogt, J ;
Gueldner, H ;
Lamp, B ;
Tenderich, G ;
Krater, L ;
Heintze, J ;
Minami, K ;
Horstkotte, D ;
Koerfer, R .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2002, 6 (01) :71-75