Do volar locking plates fit the volar cortex of the distal radius?

被引:7
作者
Kwak, D-S. [1 ]
Lee, J. -Y. [2 ]
Im, J. -H. [2 ]
Song, H. -J. [2 ]
Park, D. [2 ]
机构
[1] Catholic Univ Korea, Catholic Inst Appl Anat, Dept Anat, Seoul, South Korea
[2] Catholic Univ Korea, Dept Orthoped Surg, Seoul, South Korea
关键词
Distal radius fractures; volar locking plate; surface anatomy; volar cortical angle; POLLICIS LONGUS TENDON; INTERNAL-FIXATION; EXTERNAL FIXATION; FRACTURES; RUPTURES; ANGLE;
D O I
10.1177/1753193416676704
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The purpose of this study was to measure the volar cortical angles of the commercially available volar locking plates and evaluate how well they fit the distal radius. We measured the volar cortical angles of the radial and intermediate columns of eight volar locking plates and compared them with the volar cortical angles of 90 cadaver distal radii. The mean radial and intermediate column volar cortical angles of the wider plates were significantly larger than those of narrower plates. The mean radial and intermediate column volar cortical angles were significantly larger in wider radii than narrower radii. The width of the distal radius did not correlate well with the mean volar cortical angles of the radial and intermediate columns.
引用
收藏
页码:266 / 270
页数:5
相关论文
共 21 条
[1]  
Adham Mehdi N, 2009, Hand (N Y), V4, P406, DOI 10.1007/s11552-009-9180-0
[2]   Distal radius measurements and efficacy of fixed-angle locking volar plates [J].
Agir, Ismail ;
Aytekin, Mahmut Nedim ;
Kucukdurmaz, Fatih ;
Basci, Onur ;
Tetik, Cihangir .
TURKISH JOURNAL OF MEDICAL SCIENCES, 2014, 44 (01) :36-41
[3]   Delayed ruptures of the extensor tendon secondary to the use of volar locking compression plates for distal radial fractures [J].
Al-Rashid, M. ;
Theivendran, K. ;
Craigen, M. A. C. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2006, 88B (12) :1610-1612
[4]   Complications following internal fixation of unstable distal radius fracture with a palmar locking-plate [J].
Arora, Rohit ;
Lutz, Martin ;
Hennerbichler, Alfred ;
Krappinger, Dietinar ;
Espen, David ;
Gabl, Markus .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2007, 21 (05) :316-322
[5]   Palmar cortical angle of the distal radius: A radiological study [J].
Bassi, RS ;
Krishnan, KM ;
Dhillon, SS ;
Deshmukh, SC .
JOURNAL OF HAND SURGERY-BRITISH AND EUROPEAN VOLUME, 2003, 28B (02) :163-164
[6]   Rupture of flexor pollicis longus tendon - A complication of volar plating of the distal radius [J].
Bell, JSP ;
Wollstein, R ;
Citron, ND .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (02) :225-226
[7]   Three-dimensional computed tomography scan evaluation of the pattern of erosion in type B glenoids [J].
Beuckelaers, Ellen ;
Jacxsens, Matthijs ;
Van Tongel, Alexander ;
De Wilde, Lieven F. .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2014, 23 (01) :109-116
[8]   FLEXOR POLLICIS LONGUS TENDON RUPTURES AFTER PALMAR PLATE FIXATION OF FRACTURES OF THE DISTAL RADIUS [J].
Casaletto, J. A. ;
Machin, D. ;
Leung, R. ;
Brown, D. J. .
JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2009, 34E (04) :471-474
[9]   Trends in the United States in the Treatment of Distal Radial Fractures in the Elderly [J].
Chung, Kevin C. ;
Shauver, Melissa J. ;
Birkmeyer, John D. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (08) :1868-1873
[10]   Distal volar radial plates: How anatomical are they? [J].
Evans, S. ;
Ramasamy, A. ;
Deshmukh, S. C. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2014, 100 (03) :293-295