Elimination of residual aliasing artifact that resembles brain lesion on multi-oblique diffusion-weighted echo-planar imaging with parallel imaging using virtual coil acquisition

被引:7
作者
Liu, Xiaoxi [1 ]
Hui, Edward S. [1 ,2 ]
Chang, Hing-Chiu [1 ]
机构
[1] Univ Hong Kong, Li Ka Shing Fac Med, Dept Diagnost Radiol, Hong Kong, Peoples R China
[2] Univ Hong Kong, State Key Lab Brain & Cognit Sci, Hong Kong, Peoples R China
关键词
diffusion-weighted imaging; parallel imaging; artifact removal; multiplexed sensitivity encoding; echo-planar imaging; virtual coil acquisition; PHASE CORRECTION; GEOMETRIC DISTORTION; RECONSTRUCTION; SENSE;
D O I
10.1002/jmri.26966
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background Single-shot diffusion-weighted echo-planar imaging (ssDW-EPI) acquired with parallel imaging and a multi-oblique scan plane may suffer from residual aliasing artifacts, resembling lesions on the calculated apparent diffusion coefficient (ADC) map. Purpose To combine ssDW-EPI and virtual coil acquisition and develop a self-reference reconstruction method to eliminate the residual aliasing artifact on multi-oblique ssDW-EPI sequence with parallel imaging and multiple signal averaging. Study Type Prospective. Subjects Three healthy subjects and 50 stroke patients. Field Strength/Sequence Conventional ssDW-EPI with parallel imaging, and proposed ssDW-EPI with virtual coil acquisition at 1.5T. Assessment The efficacy of the proposed method was evaluated in 50 stroke patients by comparing the ssDW-EPI with conventional parallel imaging reconstructions. The extent of residual aliasing artifacts were rated on a 5-point Likert scale by three independent raters. Only the data without residual aliasing artifacts on conventional ssDW-EPI were included for the assessment of signal-to-noise ratio (SNR), ghost-to-signal ratio (GSR), and ADC. Statistical Tests The interobserver agreements for examining residual aliasing artifacts were measured by the intraclass correlation coefficient (ICC). A two-sample t-test was performed for comparing SNR, GSR, and ADC. Results There was a perfect agreement (ICC = 1.00) in the examination of residual aliasing artifacts on images obtained using the proposed method. The incidence rates of the residual aliasing artifact on the ADC maps obtained from the scanner console and proposed method were 60% (ie, 30 out of 50) and 0%, respectively. The proposed method offers significantly lower GSR than conventional parallel imaging reconstruction (P < 0.001). There was no significant difference in SNR (P = 0.20-0.51) and ADC values (P = 0.20-0.94) between conventional parallel imaging reconstructions and the proposed method. Data Conclusion It appears that our method could effectively eliminate artifacts and significantly improve the GSR of b = 0 T2WI and b > 0 DWI, as well as permit ADC measurement consistent with conventional techniques. Our method may be beneficial to clinical assessment of the brain that utilizes multi-oblique ssDW-EPI. Technical Efficacy Stage: 1 J. Magn. Reson. Imaging 2019.
引用
收藏
页码:1442 / 1453
页数:12
相关论文
共 29 条
[1]   A NEW PHASE CORRECTION METHOD IN NMR IMAGING BASED ON AUTOCORRELATION AND HISTOGRAM ANALYSIS [J].
AHN, CB ;
CHO, ZH .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1987, 6 (01) :32-36
[2]  
Atlas S.W., 2009, MAGNETIC RESONANCE I
[3]   Diffusion weighted imaging: Technique and applications [J].
Baliyan, Vinit ;
Das, Chandan J. ;
Sharma, Raju ;
Gupta, Arun Kumar .
WORLD JOURNAL OF RADIOLOGY, 2016, 8 (09) :785-798
[4]   Diffusion tensor imaging using single-shot SENSE-EPI [J].
Bammer, R ;
Auer, M ;
Keeling, SL ;
Augustin, M ;
Stables, LA ;
Prokesch, RW ;
Stollberger, R ;
Moseley, ME ;
Fazekas, F .
MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (01) :128-136
[5]   Motion robust GRAPPA for echo-planar imaging [J].
Baron, Corey A. ;
Beaulieu, Christian .
MAGNETIC RESONANCE IN MEDICINE, 2016, 75 (03) :1166-1174
[6]   Virtual Coil Concept for Improved Parallel MRI Employing Conjugate Symmetric Signals [J].
Blaimer, Martin ;
Gutberlet, Marcel ;
Kellman, Peter ;
Breuer, Felix A. ;
Koestler, Herbert ;
Griswold, Mark A. .
MAGNETIC RESONANCE IN MEDICINE, 2009, 61 (01) :93-102
[7]   IMAGE-RECONSTRUCTION FOR ECHO PLANAR IMAGING WITH NONEQUIDISTANT KAPPA-SPACE SAMPLING [J].
BRUDER, H ;
FISCHER, H ;
REINFELDER, HE ;
SCHMITT, F .
MAGNETIC RESONANCE IN MEDICINE, 1992, 23 (02) :311-323
[8]   Joint correction of Nyquist artifact and minuscule motion-induced aliasing artifact in interleaved diffusion weighted EPI data using a composite two-dimensional phase correction procedure [J].
Chang, Hing-Chiu ;
Chen, Nan-Kuei .
MAGNETIC RESONANCE IMAGING, 2016, 34 (07) :974-979
[9]   Interleaved Diffusion-Weighted EPI Improved by Adaptive Partial-Fourier and Multiband Multiplexed Sensitivity-Encoding Reconstruction [J].
Chang, Hing-Chiu ;
Guhaniyogi, Shayan ;
Chen, Nan-kuei .
MAGNETIC RESONANCE IN MEDICINE, 2015, 73 (05) :1872-1884
[10]   A robust multi-shot scan strategy for high-resolution diffusion weighted MRI enabled by multiplexed sensitivity-encoding (MUSE) [J].
Chen, Nan-kuei ;
Guidon, Arnaud ;
Chang, Hing-Chiu ;
Song, Allen W. .
NEUROIMAGE, 2013, 72 :41-47