RETRACTED: Dosimetric assessment of IMRT treatment planning for unilateral breast cancer patient using Octavius phantom detector

被引:16
作者
Alwakeel, Afnan F. [1 ]
Al Musawi, Mustafa S. [1 ]
Alabedi, Haydar H. [2 ]
Mohammed, Husam Jasim [3 ]
机构
[1] Mustansiriyah Univ, Coll Med, Med Phys Unit, Baghdad, Iraq
[2] Baghdad Univ, Coll Med, Surg Dept, Baghdad, Iraq
[3] Imam Jaafar Al Sadiq Univ, Dept Business Adm, Coll Adm Sci & Financial, Baghdad, Iraq
关键词
IMRT; Octavius 2D array; Threshold; Gamma passing rate; Global;
D O I
10.1007/s13204-021-01836-1
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
The goal of this research was to study the widely used criteria in Iraq and the suggested criteria from the American association of physics in medicine (AAPM). The breast cancer patients treated at Al-Andalus private hospital underwent a CT simulation. The data were exported to Monaco 5.1 treatment planning system for planning purposes using the intensity-modulated radiation therapy technique with X-ray energies of 6 or 10 MV. The planning was transferred to Agility linear accelerator for irradiation to Octavius-4D phantom detector for pre-treatment verification. The results show that there is no significant difference between the prescribed and measured point dose. In addition, the dose difference is acceptable and there was no significant correlation between the dose error and the result of gamma passing rates for all criteria. A high significant difference appeared between the 3%/2 mm and 3%/3 mm criteria for both local and global gamma passing rates at the two studied thresholds 5% and 10%. All the values of 10% thresholds were highly significant than 5% in local %GP but not significant for the global %GP. The number of segments and the total number of monitor units show an inverse relationship with gamma passing rates. The results clearly show that criteria and thresholds depend on the situation and health of the patient. This means the criteria are to be set at 3%/2 mm for plans which use high doses to organs at risk, while 3%/3 mm may be used for simpler plans with lower doses to organs at risk.
引用
收藏
页码:549 / 549
页数:8
相关论文
共 19 条
[1]  
Abdulbaqi AM, 2020, Ann Trop Med Public Health, V23, P1, DOI [10.36295/ASRO.2020.232126, DOI 10.36295/ASRO.2020.232126]
[2]  
Al Musawi MS., 2020, MAT TODAY PROC, DOI [10.1016/j.matpr.2020.10.900, DOI 10.1016/J.MATPR.2020.10.900]
[3]  
Al Musawi MS, 2016, 6 INT C PHOT ICP, P1
[4]   Efficient cooperative image transmission in one-way multi-hop sensor network (Publication with Expression of Concern) [J].
Al-Hayani, Bilal ;
Ilhan, Haci .
INTERNATIONAL JOURNAL OF ELECTRICAL ENGINEERING EDUCATION, 2020, 57 (04) :321-339
[5]  
Alani EA, 2020, PERIOD TCHE QUIM, V17, P128
[6]   Statistical variability and confidence intervals for planar dose QA pass rates [J].
Bailey, Daniel W. ;
Nelms, Benjamin E. ;
Attwood, Kristopher ;
Kumaraswamy, Lalith ;
Podgorsak, Matthew B. .
MEDICAL PHYSICS, 2011, 38 (11) :6053-6064
[7]   Patient delivery quality assurance for linac-based IMRT and helical tomotherapy using solid state detectors [J].
Chong, Ngot Swan ;
Lee, Jason J. S. ;
Kung, Wei-Hsiang ;
Chen, Chien-An ;
Hsieh, Chen-Hsi ;
Tien, Hui-Ju ;
Lin, Chi-Ta ;
Yeh, Hsin-Pei ;
Wu, Le-Jung ;
Huang, Hsien-Da ;
Shueng, Pei-Wei .
RADIATION MEASUREMENTS, 2011, 46 (12) :1993-1995
[8]  
DeVita V. T., 2012, Cancer: principles practice of oncology: annual advances in oncology
[9]   Patient QA systems for rotational radiation therapy: A comparative experimental study with intentional errors [J].
Fredh, Anna ;
Scherman, Jonas Bengtsson ;
Fog, Lotte S. ;
Af Rosenschold, Per Munck .
MEDICAL PHYSICS, 2013, 40 (03)
[10]  
Jubbier ON, 2021, 2 INT C AL KARKH U S