Precision of NIDEK OPD-scan measurements

被引:20
作者
Holzer, Mike P. [1 ]
Goebels, Susanne [1 ]
Auffarth, Gerd U. [1 ]
机构
[1] Univ Heidelberg, Dept Ophthalmol, IVCRC, D-69120 Heidelberg, Germany
关键词
D O I
10.3928/1081-597X-20061102-02
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To evaluate the repeatability of wavefront measurements using the NIDEK OPD-Scan. METHODS: A total of 179 eyes from 90 healthy volunteers (57 women and 33 men) with no corneal or lenticular pathology and normal visual acuity were enrolled in this study. Mean patient age was 39 years (range: 17 to 85 years). All patients underwent four consecutive measurements by one examiner with the NIDEK OPD-Scan. Total, corneal, and internal wavefront errors were measured and calculated with the device, using slit retinoscopy. Repeatability of the measurements was evaluated for spherical aberration, coma, and trefoil. RESULTS: The repeatability test revealed a good result for all three higher order aberrations evaluated. The best repeatability values were found for total aberrations, followed by internal and corneal aberrations. CONCLUSIONS: The NIDEK OPD-Scan has good precision in the wavefront measurement of total, corneal, and internal optical aberrations.
引用
收藏
页码:S1021 / S1023
页数:3
相关论文
共 11 条
[1]  
Canals M, 2004, J REFRACT SURG, V20, P106
[2]   Test-retest reliability of clinical Shack-Hartmann measurements [J].
Cheng, X ;
Himebaugh, NL ;
Kollbaum, PS ;
Thibos, LN ;
Bradley, A .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2004, 45 (01) :351-360
[3]   A comparative study of the performance of different corneal topographers on children with respect to orthokeratology practice [J].
Chui, WS ;
Cho, P .
OPTOMETRY AND VISION SCIENCE, 2005, 82 (05) :420-427
[4]   Corneal flap thickness and topography changes induced by flap creation during laser in situ keratomileusis [J].
Güell, JL ;
Velasco, F ;
Roberts, C ;
Sisquella, MT ;
Mahmoud, A .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (01) :115-119
[5]   Impact of a modified optic design on visual function: Clinical comparative study [J].
Mester, U ;
Dillinger, P ;
Anterist, N .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (04) :652-660
[6]  
Mrochen M, 2003, J REFRACT SURG, V19, P104
[7]  
Pop M, 2005, J REFRACT SURG, V21, pS636
[8]   Reliability of Orbscan II topography measurements in relation to refractive status [J].
Rabsilber, TM ;
Bccker, KA ;
Auffarth, GU .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (08) :1607-1613
[9]  
Rodríguez P, 2004, J REFRACT SURG, V20, P810
[10]   Clinical comparison of 6 aberrometers Part 2: Statistical comparison in a test group [J].
Rozema, JJ ;
Van Dyck, DEM ;
Tassignon, MJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (01) :33-44