Design and fabrication of microfluidic chip with micro/nano structures

被引:13
作者
Chen, X. [1 ]
Cui, D. F. [1 ]
Zhang, L. L. [1 ]
机构
[1] Chinese Acad Sci, Inst Elect, State Key Lab Transducer Tech, Beijing 100190, Peoples R China
来源
2009 4TH IEEE INTERNATIONAL CONFERENCE ON NANO/MICRO ENGINEERED AND MOLECULAR SYSTEMS, VOLS 1 AND 2 | 2009年
关键词
microfluidic; plasma isolation; micro/nano structures; MEMS; SEPARATION; FLOW; ERYTHROCYTES; SUSPENSIONS; PARTICLES; PLASMA; BLOOD;
D O I
10.1109/NEMS.2009.5068535
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this manuscript we designed and fabricated two types of microfluidic chips for plasma isolation from whole blood by using micro/nano structures. One type of the microchips was designed based on the principle of crossflow filtration, while the other was based on plasma skimming effect. The microchips were fabricated by MEMS technology, which were made of a silicon wafer and a PDMS-glass cover. The fabrication parameters were demonstrated and optimized for the both types of chips. These microfluidic chips used for plasma isolation has the potential to integrate into micro total analytical system (mu TAS) with the advantages of low cost, short analytical time, disposability, low reagent and sample consumption, which will create a microanalysis system for point-of-care diagnostics.
引用
收藏
页码:98 / 101
页数:4
相关论文
共 14 条
[1]  
Blattert C, 2004, P ANN INT IEEE EMBS, V26, P2627
[2]   A planar microfabricated fluid [J].
Brody, JP ;
Osborn, TD ;
Forster, FK ;
Yager, P .
SENSORS AND ACTUATORS A-PHYSICAL, 1996, 54 (1-3) :704-708
[3]   Sample filtration, concentration, and separation integrated on microfluidic devices [J].
Broyles, BS ;
Jacobson, SC ;
Ramsey, JM .
ANALYTICAL CHEMISTRY, 2003, 75 (11) :2761-2767
[4]   Fabrication of solid phase extraction deoxyribonucleic acid chips based on bio-microelectron-mechanical system technology [J].
Chen, X ;
Cui, DF ;
Liu, CC ;
Cai, HY .
CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2006, 34 (03) :433-436
[5]   Continuous flow microfluidic device for cell separation, cell lysis and DNA purification [J].
Chen, Xing ;
Cui, Dafu ;
Liu, Changchun ;
Li, Hui ;
Chen, Jian .
ANALYTICA CHIMICA ACTA, 2007, 584 (02) :237-243
[6]   Isolation of plasma from whole blood using planar microfilters for lab-on-a-chip applications [J].
Crowley, TA ;
Pizziconi, V .
LAB ON A CHIP, 2005, 5 (09) :922-929
[7]   Effects of particle concentration on the partitioning of suspensions at small divergent bifurcations [J].
Ditchfield, R ;
Olbricht, WL .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1996, 118 (03) :287-294
[8]   FLOW BEHAVIOR OF ERYTHROCYTES .2. PARTICLE MOTIONS IN CONCENTRATED SUSPENSIONS OF GHOST CELLS [J].
GOLDSMITH, HL ;
MARLOW, JC .
JOURNAL OF COLLOID AND INTERFACE SCIENCE, 1979, 71 (02) :383-407
[9]   Scaling and the design of miniaturized chemical-analysis systems [J].
Janasek, Dirk ;
Franzke, Joachim ;
Manz, Andreas .
NATURE, 2006, 442 (7101) :374-380
[10]   Self-contained, fully integrated biochip for sample preparation, polymerase chain reaction amplification, and DNA microarray detection [J].
Liu, RH ;
Yang, JN ;
Lenigk, R ;
Bonanno, J ;
Grodzinski, P .
ANALYTICAL CHEMISTRY, 2004, 76 (07) :1824-1831