A theoretical study on phase-contrast mammography with Thomson-scattering x-ray sources

被引:13
作者
De Caro, Liberato [1 ,2 ]
Giannini, Cinzia [1 ,2 ]
Bellotti, Roberto [2 ,3 ]
Tangaro, Sabina [2 ]
机构
[1] CNR, IC, I-70125 Bari, Italy
[2] Ist Nazl Fis Nucl, Sez Bari, I-70126 Bari, Italy
[3] Univ Bari, Dipartimento Interateneo Fis M Merlin, I-70126 Bari, Italy
关键词
phase-contrast imaging; mammography; Thomson scattering; SYNCHROTRON-RADIATION; RESOLUTION; VISUALIZATION; COHERENCE; CT;
D O I
10.1118/1.3213086
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The x-ray transmitted beam from any material/tissue depends on the complex refractive index (n=1-delta+i beta), where delta is responsible for the phase shift and beta is for the beam attenuation. Although for human tissues, the delta cross section is about 1000 times greater than the beta ones in the x-ray energy range from 10 to 150 keV, the gain in breast tumor visualization of phase-contrast mammography (PCM) with respect to absorption contact imaging (AI) is limited by the maximum dose that can be delivered to the patient. Moreover, in-line PC imaging (PCI) is the simplest experimental mode among all available x-ray PCI techniques since no optics are needed. The latter is a fundamental requirement in order to transfer the results of laboratory research into hospitals. Alternative to synchrotron radiation sources, the implementation of relativistic Thomson-scattering (TS) x-ray sources is particularly suitable for hospital use because of their high peak brightness within a relatively compact and affordable system. In this work, the possibility to realize PCM using a TS source in a hospital environment is studied, accounting for the effect of a finite deliverable dose on the PC visibility enhancement with respect to AI. Methods: The contrast-to-noise ratio of tumor-tissue lesions in PCM has been studied on the bases of a recent theoretical model, describing image contrast formation by means of both wave-optical theory and the mutual coherence formalism. The latter is used to describe the evolution, during wave propagation, of the coherence of the wave field emitted by a TS source. The contrast-to-noise ratio for both PCI and AI has been analyzed in terms of tumor size, beam energy, detector, and source distances, studying optimal conditions for performing PCM. Regarding other relevant factors which could influence "tumor" visibility, the authors have assumed simplified conditions such as a spherical shape description of the tumor inclusion, a constant surrounding background, ideal conditions for the calculus of the contrast-to-noise ratio. Results: The results show the possibility to enhance with PCI the signal-to-noise ratio for features in the submillimeter scale. This finding could give PCM a great advantage with respect to AI, opening the possibility to decrease the number of wrong diagnoses before histological exams. The results agree with experimental evidences obtained by Dreossi et al. [D. Dreossi et al., "The mammography project at the SYRMEP beamline," Eur. J. Radiol. 68, S58-S62 (2008)] with real PCM using synchrotron radiation. Conclusions: The challenging characteristics of TS sources, suitable for PCM, should be fluxes of at least 10(11) photons/s emitted by very small sources of about 10 mu m, together with moderate energy spreads (less than 10%) in order to realize both sufficient spatial coherence and enough fluence on the patient, collecting images in reasonable exposure times. These fluxes will be achieved by next generation TS sources which are already under development. (C) 2009 American Association of Physicists in Medicine. [DOI :10.1118/1.3213086]
引用
收藏
页码:4644 / 4653
页数:10
相关论文
共 47 条
[1]   Mammography with synchrotron radiation: Phase-detection techniques [J].
Arfelli, F ;
Bonvicini, V ;
Bravin, A ;
Cantatore, G ;
Castelli, E ;
Dalla Palma, L ;
Di Michiel, M ;
Fabrizioli, M ;
Longo, R ;
Menk, RH ;
Olivo, A ;
Pani, S ;
Pontoni, D ;
Poropat, P ;
Prest, M ;
Rashevsky, A ;
Ratti, M ;
Rigon, L ;
Tromba, G ;
Vacchi, A ;
Vallazza, E ;
Zanconati, F .
RADIOLOGY, 2000, 215 (01) :286-293
[2]   Hard X-ray phase-contrast imaging with the Compact Light Source based on inverse Compton X-rays [J].
Bech, Martin ;
Bunk, Oliver ;
David, Christian ;
Ruth, Ronald ;
Rifkin, Jeff ;
Loewen, Rod ;
Feidenhans'l, Robert ;
Pfeiffer, Franz .
JOURNAL OF SYNCHROTRON RADIATION, 2009, 16 :43-47
[3]   Glandular breast dose for monoenergetic and high-energy X-ray beams: Monte Carlo assessment [J].
Boone, JM .
RADIOLOGY, 1999, 213 (01) :23-37
[4]   High-resolution CT by diffraction-enhanced x-ray imaging:: mapping of breast tissue samples and comparison with their histo-pathology [J].
Bravin, Alberto ;
Keyrilainen, Jani ;
Fernandez, Manuel ;
Fiedler, Stefan ;
Nemoz, Christian ;
Karjalainen-Lindsberg, Marja-Liisa ;
Tenhunen, Mikko ;
Virkkunen, Pekka ;
Leidenius, Marjut ;
von Smitten, Karl ;
Sipila, Petri ;
Suortti, Pekka .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (08) :2197-2211
[5]   Three-dimensional time and frequency-domain theory of femtosecond x-ray pulse generation through Thomson scattering [J].
Brown, WJ ;
Hartemann, FV .
PHYSICAL REVIEW SPECIAL TOPICS-ACCELERATORS AND BEAMS, 2004, 7 (06) :35-54
[6]   Experimental characterization of an ultrafast Thomson scattering x-ray source with three-dimensional time and frequency-domain analysis [J].
Brown, WJ ;
Anderson, SG ;
Barty, CPJ ;
Betts, SM ;
Booth, R ;
Crane, JK ;
Cross, RR ;
Fittinghoff, DN ;
Gibson, DJ ;
Hartemann, FV ;
Hartouni, EP ;
Kuba, J ;
Le Sage, GP ;
Slaughter, DR ;
Tremaine, AM ;
Wootton, AJ ;
Springer, PT ;
Rosenzweig, JB .
PHYSICAL REVIEW SPECIAL TOPICS-ACCELERATORS AND BEAMS, 2004, 7 (06) :23-34
[7]   SPATIAL COHERENCE OF SYNCHROTRON-RADIATION [J].
COISSON, R .
APPLIED OPTICS, 1995, 34 (05) :904-908
[8]   In-line phase-contrast imaging for strong absorbing objects [J].
De Caro, Liberato ;
Cedola, Alessia ;
Giannini, Cinzia ;
Bukreeva, Inna ;
Lagomarsino, Stefano .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (22) :6619-6637
[9]   The mammography project at the SYRMEP beamline [J].
Dreossi, D. ;
Abrami, A. ;
Arfelli, F. ;
Bregant, R. ;
Casarin, K. ;
Chenda, V. ;
Cova, M. A. ;
Longo, R. ;
Menk, R. -H. ;
Quai, E. ;
Quaia, E. ;
Rigon, L. ;
Rokvic, T. ;
Sanabor, D. ;
Tonutti, M. ;
Tromba, G. ;
Vascotto, A. ;
Zanconati, F. ;
Castelli, E. .
EUROPEAN JOURNAL OF RADIOLOGY, 2008, 68 (03) :S58-S62
[10]  
EUREF, 2003, ADD DIG MAMM EUR GUI