Heartmate 3 fully magnetically levitated left ventricular assist device for the treatment of advanced heart failure-1 year results from the Ce mark trial

被引:84
作者
Krabatsch, Thomas [1 ]
Netuka, Ivan [2 ]
Schmitto, Jan D. [3 ]
Zimpfer, Daniel [4 ]
Garbade, Jens [5 ]
Rao, Vivek [6 ]
Morshuis, Michiel [7 ]
Beyersdorf, Friedhelm [8 ]
Marasco, Silvana [9 ]
Damme, Laura [10 ]
Pya, Yuriy [11 ]
机构
[1] Deutsch Herzzentrum Berlin, German Heart Ctr, Augustenburger Pl 1, D-13353 Berlin, Germany
[2] Inst Clin & Expt Med, Videnska 1958-9, Prague 4, Czech Republic
[3] Hannover Med Sch, Carl Neuberg Str 1, D-30625 Hannover, Germany
[4] Univ Vienna, Waehringer Guertel 18-20, A-1090 Vienna, Austria
[5] Heart Ctr Leipzig, Struempellstr 39, D-04289 Leipzig, Germany
[6] Toronto Gen Hosp, 4N-464 200 Elizabeth St, Toronto, ON M5G 2C4, Canada
[7] Herz & Diabet Zentrum NRW, D-32545 Bad Oeynhausen, Germany
[8] Univ Heart Ctr Freiburg Bad Krozingen, Hugstetterstr 55, D-79106 Freiburg, Germany
[9] Alfred Hosp, 55 Commercial Rd, Melbourne, Vic 3181, Australia
[10] St Jude Med, 11 Vincilaan, B-1935 Zaventem, Belgium
[11] Natl Res Cardiac Surg Ctr, 38 Turan St, Astana 010000, Kazakhstan
来源
JOURNAL OF CARDIOTHORACIC SURGERY | 2017年 / 12卷
关键词
HeartMate; 3; Magnetic levitation; Heart failure; LVAS; MECHANICAL CIRCULATORY SUPPORT; DRUG ADMINISTRATION-APPROVAL; VON-WILLEBRAND SYNDROME; ACCESS PROTOCOL TRIAL; PUMP THROMBOSIS; INTERAGENCY REGISTRY; CLINICAL-EXPERIENCE; INTERMACS REGISTRY; SYSTEM; TRANSPLANTATION;
D O I
10.1186/s13019-017-0587-3
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The HeartMate 3 Left Ventricular Assist System (LVAS) (St. Jude Medical Inc., St Paul, MN) with full magnetic levitation allows for wide and consistent blood flow paths and an artificial pulse designed for enhanced hemocompatibility. The HeartMate 3 received market approval in the European Union in 2015 following completion of a multicenter study. After reaching the 6-month study endpoint, patients continue to be followed for 2 years with the 1-year results presented herein. Methods: A prospective, non-randomized study included adults with advanced heart failure and ejection fraction (EF) <= 25%, cardiac index (CI) <= 2.2 L/min/m(2) while not on inotropes, or inotrope dependent, or on optimal medical management for 45/60 days. Results: Fifty patients-54% bridge to transplant (BTT) and 46% destination therapy (DT)-were enrolled and implanted with the HeartMate 3. At baseline, 92% of the patients were INTERMACS profiles 2-4, with cardiac index 1.8 +/- 0.5 L/min/m(2) and 58% were supported with inotropes. At 1 year, 74% of the patients remain on support, 18% expired, 6% transplanted, and 2% explanted. The adverse events include 12% gastrointestinal bleeding, 16% driveline infections, 18% strokes, and 2% outflow graft thrombosis. There was no hemolysis, pump thrombosis or pump malfunction through 1 year. The six-minute walk test distance increased from a mean of 273 m to 371 m (P < 0.0001). EQ-5D quality-of-life score increased from a mean of 52.7 to 70.8 (P = 0.0006). Conclusions: The 1-year HeartMate 3 LVAS results show survival and adverse-event profile are similar to other approved devices, with no pump thrombosis or pump failure. Patient's functional status and quality of life significantly improved over time.
引用
收藏
页数:8
相关论文
共 38 条
[1]   Design Rationale and Preclinical Evaluation of the HeartMate 3 Left Ventricular Assist System for Hemocompatibility [J].
Bourque, Kevin ;
Cotter, Christopher ;
Dague, Charles ;
Harjes, Daniel ;
Dur, Onur ;
Duhamel, Julien ;
Spink, Kaitlyn ;
Walsh, Kelly ;
Burke, Edward .
ASAIO JOURNAL, 2016, 62 (04) :375-383
[2]   Initial Clinical Experience With the HeartWare Left Ventricular Assist System: A Single-Center Report [J].
Dell'Aquila, Angelo Maria ;
Schneider, Stefan Rudolf Bertram ;
Schlarb, Dominik ;
Redwan, Bassam ;
Sindermann, Juergen R. ;
Ellger, Bjoern ;
Stypmann, Joerg ;
Tjan, Tony D. T. ;
Scheid, Hans H. ;
Hoffmeier, Andreas .
ANNALS OF THORACIC SURGERY, 2013, 95 (01) :170-178
[3]   Gastrointestinal Bleeding in Recipients of the HeartWare Ventricular Assist System [J].
Goldstein, Daniel J. ;
Aaronson, Keith D. ;
Tatooles, Antone J. ;
Silvestry, Scott C. ;
Jeevanandam, Valluvan ;
Gordon, Robert ;
Hathaway, David R. ;
Najarian, Kevin B. ;
Slaughter, Mark S. .
JACC-HEART FAILURE, 2015, 3 (04) :303-313
[4]   Clinical trial design and rationale of the Multicenter Study of MagLev Technology in Patients Undergoing Mechanical Circulatory Support Therapy With HeartMate 3 (MOMENTUM 3) investigational device exemption clinical study protocol [J].
Heatley, Gerald ;
Sood, Poornima ;
Goldstein, Daniel ;
Uriel, Nir ;
Cleveland, Joseph ;
Middlebrook, Don ;
Mehra, Mandeep R. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2016, 35 (04) :528-536
[5]   Gastrointestinal Bleeding during Continuous-Flow Left Ventricular Assist Device Support is Associated with Lower Rates of Cardiac Transplantation [J].
Holley, Christopher T. ;
Harvey, Laura ;
Roy, Samit S. ;
Cogswell, Rebecca ;
Eckman, Peter ;
Liao, Kenneth ;
John, Ranjit .
ASAIO JOURNAL, 2015, 61 (06) :635-639
[6]   Continuous Flow Left Ventricular Assist Device Outcomes in Commercial Use Compared With the Prior Clinical Trial [J].
John, Ranjit ;
Naka, Yoshifumi ;
Smedira, Nicholas G. ;
Starling, Randall ;
Jorde, Ulrich ;
Eckman, Peter ;
Farrar, David J. ;
Pagani, Francis D. .
ANNALS OF THORACIC SURGERY, 2011, 92 (04) :1406-1412
[7]   Results of the Destination Therapy Post-Food and Drug Administration Approval Study With a Continuous Flow Left Ventricular Assist Device A Prospective Study Using the INTERMACS Registry (Interagency Registry for Mechanically Assisted Circulatory Support) [J].
Jorde, Ulrich P. ;
Kushwaha, Sudhir S. ;
Tatooles, Antone J. ;
Naka, Yoshifumi ;
Bhat, Geetha ;
Long, James W. ;
Horstmanshof, Douglas A. ;
Kormos, Robert L. ;
Teuteberg, Jeffrey J. ;
Slaughter, Mark S. ;
Birks, Emma J. ;
Farrar, David J. ;
Park, Soon J. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2014, 63 (17) :1751-1757
[8]   First Annual IMACS Report A global International Society for Heart and Lung Transplantation Registry for Mechanical Circulatory Support [J].
Kirklin, James K. ;
Cantor, Ryan ;
Mohacsi, Paul ;
Gummert, Jan ;
Theo De By ;
Hannan, Margaret M. ;
Kormos, Robert L. ;
Schueler, Stephan ;
Lund, Lars H. ;
Nakatani, Takeshi ;
Taylor, Rhiannon ;
Lannon, Jenny .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2016, 35 (04) :407-412
[9]   Pump thrombosis in the Thoratec HeartMate II device: An update analysis of the INTERMACS Registry [J].
Kirklin, James K. ;
Naftel, David C. ;
Pagani, Francis D. ;
Kormos, Robert L. ;
Myers, Susan ;
Acker, Michael. A. ;
Rogers, Joseph ;
Slaughter, Mark S. ;
Stevenson, Lynne W. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2015, 34 (12) :1515-1526
[10]   Device Exchange in HeartMate II Recipients: Long-Term Outcomes and Risk of Thrombosis Recurrence [J].
Levin, Allison P. ;
Uriel, Nir ;
Takayama, Hiroo ;
Mody, Kanika P. ;
Ota, Takeyoshi ;
Yuzefpolskaya, Melana ;
Colombo, Paolo C. ;
Garan, Arthur R. ;
Dionizovik-Dimanovski, Marija ;
Sladen, Robert N. ;
Naka, Yoshifumi ;
Jorde, Ulrich P. .
ASAIO JOURNAL, 2015, 61 (02) :144-149