Dibenzo-18-crown-6-based carbon paste sensors for the nanomolar potentiometric determination of daclatasvir dihydrochloride: An anti-HCV drug and a potential candidate for treatment of SARS-CoV-2

被引:13
作者
Ahmed, Yomna M. [1 ]
Badawy, Sayed S. [1 ]
Abdel-Haleem, Fatehy M. [1 ,2 ]
机构
[1] Cairo Univ, Fac Sci, Chem Dept, Giza, Egypt
[2] Cairo Univ, Ctr Hazards Mitigat Environm Studies & Res CHMESR, Giza, Egypt
关键词
Daclatasvir; Anti-HCV; SARS-CoV-2; Carbon paste electrodes; Dibenzo-18-crown-6; Potentiometry; ION-SELECTIVE ELECTRODES; HEPATITIS-C; ANIONIC SITES; COEFFICIENTS; MEMBRANE; HYDROCHLORIDE; ADDITIVES; CATIONS;
D O I
10.1016/j.microc.2022.107276
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Daclatasvir dihydrochloride (DAC) is an anti-hepatitis C virus (HCV) drug that has recently proven to be a promising candidate for the treatment of SARS-CoV-2. Still, there is a lack of sensitive potentiometric methods for its determination. In this work, carbon paste sensors based on dibenzo-18-crown-6 (DB18C6) were fabricated and optimized for the sensitive and selective potentiometric determination of DAC in Daclavirocyrl (R) tablets, serum, and urine samples. The best performance was obtained by two sensors referred to as sensor I and sensor II. Both sensors exhibited a wide linear response range of 5x10- 91x10- 3mol/L, and Nernstian slopes of 29.8 & PLUSMN; 1.18 and 29.5 +/- 1.00 mV/decade, with limits of detection, 4.8x10(- 9) and 3.2x10(-9) mol/L, for the sensors I and II, respectively. Sensors I and II displayed fast response times of 5-8 and 5-6 s, respectively, with great reversibility and no memory effect. Moreover, the sensors exhibited a lifetime of 16 days. For the study of sensors morphology and elucidation of the interaction mechanism, the scanning electron microscope (SEM), Fourier transform infrared spectroscopy (FTIR), and nuclear magnetic resonance (1H NMR) techniques were performed. A selectivity study was performed, and the proposed sensors exhibited good discrimination between DAC and potentially coexisting interferents with sensor II displaying better selectivity. Finally, sensor II was successfully applied for the determination of DAC in the above-mentioned samples, with recovery values ranging from 99.25 to 101.42%, and relative standard deviation (RSD) values ranging from 0.79 to 1.53% which reflected the high accuracy and precision.
引用
收藏
页数:11
相关论文
共 64 条
[1]   Molecularly imprinted polymer-based bulk optode for the determination of itopride hydrochloride in physiological fluids [J].
Abdel-Haleem, F. M. ;
Madbouly, Adel ;
El Nashar, R. M. ;
Abdel-Ghani, N. T. .
BIOSENSORS & BIOELECTRONICS, 2016, 85 :740-742
[2]   Polyvinyl Chloride Modified Carbon Paste Electrodes for Sensitive Determination of Levofloxacin Drug in Serum, Urine, and Pharmaceutical Formulations [J].
Abdel-Haleem, Fatehy M. ;
Mahmoud, Sonia ;
Abdel-Ghani, Nour Eldin T. ;
El Nashar, Rasha Mohamed ;
Bechelany, Mikhael ;
Barhoum, Ahmed .
SENSORS, 2021, 21 (09)
[3]   Carbon-based Nanosensors for Salicylate Determination in Pharmaceutical Preparations [J].
Abdel-Haleem, Fatehy M. ;
Salah, Azza ;
Rizk, Mahmoud S. ;
Moustafa, Hussein ;
Bechelany, Mikhael ;
Barhoum, Ahmed .
ELECTROANALYSIS, 2019, 31 (04) :778-789
[4]   PVC membrane, coated-wire, and carbon-paste ion-selective electrodes for potentiometric determination of galantamine hydrobromide in physiological fluids [J].
Abdel-Haleem, Fatehy M. ;
Saad, Mohamed ;
Barhoum, Ahmed ;
Bechelany, Mikhael ;
Rizk, Mahmoud S. .
MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2018, 89 :140-148
[5]   Development of new potentiometric sensors for the determination of proguanil hydrochloride in serum and urine [J].
Abdel-Haleem, Fatehy M. ;
Saad, Mohamed ;
Rizk, Mohamed S. .
CHINESE CHEMICAL LETTERS, 2016, 27 (06) :857-863
[6]   Synthesis, characterization, and complexation of tetraarylborates with aromatic cations and their use in chemical sensors [J].
Alaviuhkola, T ;
Bobacka, J ;
Nissinen, M ;
Rissanen, K ;
Ivaska, A ;
Pursiainen, J .
CHEMISTRY-A EUROPEAN JOURNAL, 2005, 11 (07) :2071-2080
[7]   Applying different spectroscopic techniques for the selective determination of daclatasvir using merbromin as a probe: Applications on pharmaceutical analysis [J].
Almahri, Albandary ;
Abdel-Lateef, Mohamed A. .
LUMINESCENCE, 2021, 36 (06) :1544-1552
[8]   Cationic or anionic sites?: Selectivity optimization of ion-selective electrodes based on charged ionophores [J].
Amemiya, S ;
Bühlmann, P ;
Pretsch, E ;
Rusterholz, B ;
Umezawa, Y .
ANALYTICAL CHEMISTRY, 2000, 72 (07) :1618-1631
[9]   LIPOPHILIC SALTS AS MEMBRANE ADDITIVES AND THEIR INFLUENCE ON THE PROPERTIES OF MACROELECTRODES AND MICROELECTRODES BASED ON NEUTRAL CARRIERS [J].
AMMANN, D ;
PRETSCH, E ;
SIMON, W ;
LINDNER, E ;
BEZEGH, A ;
PUNGOR, E .
ANALYTICA CHIMICA ACTA, 1985, 171 (MAY) :119-129
[10]   Determination of unbiased selectivity coefficients of neutral carrier-based cation-selective electrodes [J].
Bakker, E .