Collimator and energy window optimization for 90Y bremsstrahlung SPECT imaging: A SIMIND Monte Carlo study

被引:22
作者
Roshan, Hoda Rezaei [1 ]
Mahmoudian, Babak [2 ]
Gharepapagh, Esmaeil [2 ]
Azarm, Ahmadreza [1 ]
Islamian, Jalil Pirayesh [1 ]
机构
[1] Tabriz Univ Med Sci, Fac Med, Dept Med Phys, Attar Neyshapouri St,Azadi Ave, Tabriz 5166614766, Iran
[2] Tabriz Univ Med Sci, Fac Med, Dept Radiol Radiotherapy & Nucl Med, Attar Neyshapouri St,Azadi Ave, Tabriz 5166614766, Iran
关键词
Bremsstrahlung imaging; Energy window; Image contrast; Medium-energy collimator; Monte Carlo simulation; Radioembolization SPECT; Yttrium-90; RADIOEMBOLIZATION; TUMORS;
D O I
10.1016/j.apradiso.2015.12.041
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Treatment efficacy of radioembolization using Yttrium-90 (Y-90) microspheres is assessed by the Y-90 bremsstrahlung single photon emission computed tomography (SPECT) imaging following radioembolization. The radioisotopic image has the potential of providing reliable activity map of Y-90 microspheres distribution. One of the main reasons of the poor image quality in Y-90 bremsstrahlung SPECT imaging is the continuous and broad energy spectrum of the related bremsstrahlung photons. Furthermore, collimator geometry plays an impressive role in the spatial resolution, sensitivity and image contrast. Due to the relatively poor quality of the Y-90 bremsstrahlung SPECT images, we intend to optimize the medium-energy (ME) parallel-hole collimator and energy window. The Siemens e.cam gamma camera equipped with a ME collimator and a voxelized phantom was simulated by the Simulating Medical Imaging Nuclear Detectors (SIMIND) program. We used the SIMIND Monte Carlo program to generate the Y-90 bremsstrahlung SPECT projection of the digital Jaszczak phantom. The phantom consist of the six hot spheres ranging from 9.5 to 31.8 mm in diameter, which are used to evaluate the image contrast. In order to assess the effect of the energy window on the image contrast, three energy windows ranging from 60 to 160 KeV, 160 to 400 KeV, and 60 to 400 KeV were set on a Y-90 bremsstrahlung spectrum. As well, the effect of the hole diameter of a ME collimator on the image contrast and bremsstrahlung spectrum were investigated. For the fixed collimator and septa thickness values (3.28 cm and 1.14 mm, respectively), a hole diameter range (2.35-3.3 mm) was chosen based on the appropriate balance between the spatial resolution and sensitivity. The optimal energy window for Y-90 bremsstrahlung SPECT imaging was extended energy window from 60 to 400 KeV. Besides, The optimal value of the hole diameter of ME collimator was obtained 33 mm. Geometry of the ME parallel-hole collimator and energy window are indeed important indicators of the image quality in Y-90 bremsstrahlung imaging. The obtained optimal ME collimator and optimal energy window have the potential to improve the image contrast of Y-90 bremsstrahlung images. Subsequently, high quality Y-90 bremsstrahlung images can provide reliable estimate of the Y-90 microsphere activity distribution after radioembolization. (C) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:124 / 128
页数:5
相关论文
共 18 条
[11]   Optimization of energy window for 90Y bremsstrahlung SPECT imaging for detection tasks using the ideal observer with model-mismatch [J].
Rong, Xing ;
Ghaly, Michael ;
Frey, Eric C. .
MEDICAL PHYSICS, 2013, 40 (06)
[12]   A method for energy window optimization for quantitative tasks that includes the effects of model-mismatch on bias: application to Y-90 bremsstrahlung SPECT imaging [J].
Rong, Xing ;
Du, Yong ;
Frey, Eric C. .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (12) :3711-3725
[13]   Development and evaluation of an improved quantitative 90Y bremsstrahlung SPECT method [J].
Rong, Xing ;
Du, Yong ;
Ljungberg, Michael ;
Rault, Erwann ;
Vandenberghe, Stefaan ;
Frey, Eric C. .
MEDICAL PHYSICS, 2012, 39 (05) :2346-2358
[14]  
Roshan Hoda Rezaei, 2015, World J Nucl Med, V14, P75, DOI 10.4103/1450-1147.157120
[15]   Optimizing pinhole and parallel hole collimation for scintimammography with compact pixellated detectors [J].
Smith, MF ;
Majewski, S ;
Weisenberger, AG .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (03) :321-326
[16]   SIMIND Monte Carlo simulation of a single photon emission CT [J].
Toossi, M. T. Bahreyni ;
Islamian, J. Pirayesh ;
Momennezhad, M. ;
Ljungberg, M. ;
Naseri, S. H. .
JOURNAL OF MEDICAL PHYSICS, 2010, 35 (01) :42-47
[17]   Optimal design of Anger camera for bremsstrahlung imaging: Monte Carlo evaluation [J].
Walrand, Stephan ;
Hesse, Michel ;
Wojcik, Randy ;
Lhommel, Renaud ;
Jamar, Francois .
FRONTIERS IN ONCOLOGY, 2014, 4
[18]  
Zeng H., 2014, MED CHEM, V4, P481