The Effective Dynamic Ranges of Standard Automated Perimetry Sizes III and V and Motion and Matrix Perimetry

被引:57
作者
Wall, Michael [1 ,2 ]
Woodward, Kimberly R. [2 ]
Doyle, Carrie K. [2 ]
Zamba, Gideon [3 ]
机构
[1] Univ Iowa, Coll Med, Vet Adm Hosp, Dept Neurol, Iowa City, IA 52242 USA
[2] Univ Iowa, Coll Med, Vet Adm Hosp, Dept Ophthalmol, Iowa City, IA 52242 USA
[3] Univ Iowa, Dept Biostat, Coll Publ Hlth, Iowa City, IA 52242 USA
关键词
GLAUCOMA;
D O I
10.1001/archophthalmol.2010.71
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objectives: To establish the associations between threshold estimates of 4 perimetric tests and to define and compare the tests' effective dynamic ranges. Methods: We examined 152 patients with glaucoma and 80 controls using standard automated perimetry (SAP) with stimulus size III, SAP with size V, and motion and matrix perimetry. We explored the intertest associations using principal-components analysis. We defined the effective dynamic range bottom using the frequency of 0-dB trials on retest. We defined the upper effective dynamic range as a value above which fewer than 0.5% of the values fall in the controls. We also calculated the number of discriminable steps from normal to the floor value of the perimeter. Results: The association between SAP III and V was approximately linear up to a sensitivity of about 20 dB on both tests and with motion and matrix perimetry up to about 25 dB from 0 dB. While the upper bounds were similar among the tests, size V SAP had a lower floor and more discriminable steps. Conclusions: The effective dynamic range of SAP III is substantially less than its physically tested limits. Size V stimuli have a greater effective dynamic range than size III by about 1 log unit and have about twice as many discriminable steps.
引用
收藏
页码:570 / 576
页数:7
相关论文
共 10 条
[1]   Characteristics of the normative database for the Humphrey Matrix perimeter [J].
Anderson, AJ ;
Johnson, CA ;
Fingeret, M ;
Keltner, JL ;
Spry, PGD ;
Wall, M ;
Werner, JS .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46 (04) :1540-1548
[2]  
Anderson Andrew J, 2003, Ophthalmol Clin North Am, V16, P213, DOI 10.1016/S0896-1549(03)00011-7
[3]  
Artes PH, 2002, INVEST OPHTH VIS SCI, V43, P2654
[4]   TEST-RETEST VARIABILITY IN GLAUCOMATOUS VISUAL-FIELDS [J].
HEIJL, A ;
LINDGREN, A ;
LINDGREN, G .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1989, 108 (02) :130-135
[5]  
Heijl A, 1989, PERIMETRY UPDATE 199, P313
[6]   Structure versus function in glaucoma: An application of a linear model [J].
Hood, Donald C. ;
Anderson, Susan C. ;
Wall, Michael ;
Kardon, Randy H. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2007, 48 (08) :3662-3668
[7]   PSYCHOPHYSICAL ISOLATION OF MOVEMENT SENSITIVITY BY REMOVAL OF FAMILIAR POSITION CUES [J].
NAKAYAMA, K ;
TYLER, CW .
VISION RESEARCH, 1981, 21 (04) :427-433
[8]   RANDOM-DOT MOTION PERIMETRY IN PATIENTS WITH GLAUCOMA AND IN NORMAL SUBJECTS [J].
WALL, M ;
KETOFF, KM .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1995, 120 (05) :587-596
[9]   USING MOTION PERIMETRY TO DETECT VISUAL-FIELD DEFECTS IN PATIENTS WITH IDIOPATHIC INTRACRANIAL HYPERTENSION - A COMPARISON WITH CONVENTIONAL AUTOMATED PERIMETRY [J].
WALL, M ;
MONTGOMERY, EB .
NEUROLOGY, 1995, 45 (06) :1169-1175
[10]   Logics and logistics of community intervention against osteoporosis: An evidence basis [J].
Waller J. ;
Angbratt M. ;
Blomberg C. ;
Kronhed A.-C. ;
Larsson L. ;
Löfman O. ;
Möller M. ;
Toss G. ;
Foldevi M. ;
Trell E. .
Journal of Medical Systems, 1997, 21 (1) :33-47