Evaluation of the Mechanical Reliability of Different Implant-Abutment Connections for Single Crowns According to the ISO 14801 Fatigue Test Protocol

被引:5
作者
Thome, Geninho [1 ]
Torres Gomes, Andre Felipe [1 ]
de Moura, Marcos Boaventura [2 ]
Beng, Christian Felippi [3 ]
Trojan, Larissa Carvalho [1 ]
机构
[1] Latin Amer Inst Res & Dent Educ ILAPEO, Dept Implantol, Curitiba, Parana, Brazil
[2] Univ Fed Uberlandia, Sch Dent, Dept Occlus Fixed Prosthodont & Dent Mat, Av Para 1720, BR-38405902 Uberlandia, MG, Brazil
[3] Univ Fed Parana, Curitiba, Parana, Brazil
关键词
abutment; biomechanics; compressive loading; dental implants; COMPLICATION RATES; OBSERVATION PERIOD; FRACTURE STRENGTH; CANTILEVER; INTERFACE; SURVIVAL; FAILURE; SYSTEM;
D O I
10.11607/jomi.7965
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The aim of this in vitro study was to evaluate the mechanical behavior regarding dynamic fatigue of different implant-abutment connections and the unitary indication of abutments for all regions of the mouth. Materials and Methods: This experimental study developed according to international standards (ISO 14801:2007) was performed using five types of implants and abutments: G1-external hex smart implant and 17-degree universal abutment (EHS); G2-cortical external hex implant and 17-degree universal abutment (EHTi); G3-internal hex implant and 30-degree universal abutment (IH); G4-Morse taper implant (11.5 degrees) and 17-degree universal abutment (MT11.5); and G5-Morse taper implant (16 degrees) and 30-degree universal abutment (MT16). A 15-Hz cyclic loading was applied to the specimens with the maximum number of cycles set at 5 x 10(6). Success was defined when three samples supported 5 million cycles without failure. The maximum load supported from each group after dynamic loading was recorded. The Spearman correlation and the Lowess method were used to analyze the correlation between the number of cycles and the applied load, and the Kruskal-Wallis and Nemenyi tests were used for comparison between the abutments when reaching 5 million cycles. Results: There was a negative correlation (r < 0.00) and significant difference (P < .05) between the number of cycles and the load for each type of implant and abutment. The load values supported by each group after cyclic loading to achieve 5 million cycles were as follows: EHS, 225 N; EHTi, 215 N; IH, 220 N; MT11.5, 210 N; and MT16, 240 N. The MT16 implant-abutment assembly presented a significantly higher load (P = .024) than the MT11.5 implant-abutment assembly. Conclusion: All implant-abutment connections investigated in this study resisted average occlusal force values reported as acceptable in the literature and may be indicated for any region of the mouth.
引用
收藏
页码:47 / 54
页数:8
相关论文
共 29 条
[1]   A systematic review of the survival and complication rates of implant supported fixed dental prostheses with cantilever extensions after an observation period of at least 5 years [J].
Aglietta, Marco ;
Siciliano, Vincenzo Iorio ;
Zwahlen, Marcel ;
Braegger, Urs ;
Pjetursson, Biarni E. ;
Lang, Niklaus P. ;
Salvi, Giovanni E. .
CLINICAL ORAL IMPLANTS RESEARCH, 2009, 20 (05) :441-451
[2]   Mechanics of the tapered interference fit in dental implants [J].
Bozkaya, D ;
Müftü, S .
JOURNAL OF BIOMECHANICS, 2003, 36 (11) :1649-1658
[3]   A retrospective study on clinical and radiological outcomes of oral implants in patients followed up for a minimum of 20 years [J].
Chrcanovic, Bruno Ramos ;
Kisch, Jeno ;
Albrektsson, Tomas ;
Wennerberg, Ann .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2018, 20 (02) :199-207
[4]   Fracture strength of implant abutments after fatigue testing: A systematic review and a meta-analysis [J].
Coray, Rafaela ;
Zeltner, Marco ;
Ozcan, Mutlu .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2016, 62 :333-346
[5]   Biomechanical behavior of three different types of internal tapered connections after cyclic and static loading tests: experimental in vitro [J].
de Moura, Marcos Boaventura ;
Loureiro, Karine Regina Tolesano ;
Lima, Livia Bonjardim ;
Felippi, Christian ;
Simamoto Junior, Paulo Cezar .
INTERNATIONAL JOURNAL OF IMPLANT DENTISTRY, 2020, 6 (01)
[6]   Influence of Screw Surface Treatment on Retention of Implant-Supported Fixed Partial Dentures [J].
de Moura, Marcos Boaventura ;
Rodrigues, Renata Borges ;
Pinto, Leandro Moreira ;
de Araujo, Cleudmar Amaral ;
Novais, Veridiana Resende ;
Simamoto Junior, Paulo Cezar .
JOURNAL OF ORAL IMPLANTOLOGY, 2017, 43 (04) :254-260
[7]   Maximal bite forces in healthy young adults as predicted by surface electromyography [J].
Ferrario, VF ;
Sforza, C ;
Zanotti, G ;
Tartaglia, GM .
JOURNAL OF DENTISTRY, 2004, 32 (06) :451-457
[8]   Biting and chewing in overdentures, full dentures, and natural dentitions [J].
Fontijn-Tekamp, FA ;
Slagter, AP ;
Van Der Bilt, A ;
Van 'T Hof, MA ;
Witter, DJ ;
Kalk, W ;
Jansen, JA .
JOURNAL OF DENTAL RESEARCH, 2000, 79 (07) :1519-1524
[9]   Micromotion and dynamic fatigue properties of the denial implant-abutment interface [J].
Gratton, DG ;
Aquilino, SA ;
Stanford, CM .
JOURNAL OF PROSTHETIC DENTISTRY, 2001, 85 (01) :47-52
[10]   Influence of Different Screw Torque Levels on the Biomechanical Behavior of Tapered Prosthetic Abutments [J].
Herbst, Paulo Eduardo ;
de Carvalho, Eduardo Bortolas ;
Salatti, Rafael C. ;
Valgas, Laiz ;
Tiossi, Rodrigo .
INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2018, 33 (03) :536-540