3D-printable lung phantom for distal falloff verification of proton Bragg peak

被引:8
作者
Koketsu, Junichi [1 ]
Kumada, Hiroaki [1 ,2 ]
Takada, Kenta [3 ]
Takei, Hideyuki [1 ,2 ]
Mori, Yutaro [1 ,2 ]
Kamizawa, Satoshi [1 ]
Hu, Yuchao [1 ]
Sakurai, Hideyuki [1 ,2 ]
Sakae, Takeji [1 ,2 ]
机构
[1] Univ Tsukuba Hosp, Proton Med Res Ctr, Tsukuba, Ibaraki, Japan
[2] Univ Tsukuba, Fac Med, Tsukuba, Ibaraki, Japan
[3] Gunma Prefectural Coll Hlth Sci, Dept Radiol Technol, Maebashi, Gunma, Japan
关键词
3D printer; lung phantom; measurement; Monte Carlo simulation; proton therapy; quality assurance; THERAPY; DEGRADATION; DENSITY; TUMORS; MODEL;
D O I
10.1002/acm2.12706
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In proton therapy, the Bragg peak of a proton beam reportedly deteriorates when passing though heterogeneous structures such as human lungs. Previous studies have used heterogeneous random voxel phantoms, in which soft tissues and air are randomly allotted to render the phantoms the same density as human lungs, for conducting Monte Carlo (MC) simulations. However, measurements of these phantoms are complicated owing to their difficult-to-manufacture shape. In the present study, we used Voronoi tessellation to design a phantom that can be manufactured, and prepared a Voronoi lung phantom for which both measurement and MC calculations are possible. Our aim was to evaluate the effectiveness of this phantom as a new lung phantom for investigating proton beam Bragg peak deterioration. For this purpose, we measured and calculated the percentage depth dose and the distal falloff widths (DFW) passing through the phantom. For the 155 MeV beam, the measured and calculated DFW values with the Voronoi lung phantom were 0.40 and 0.39 cm, respectively. For the 200 MeV beam, the measured and calculated DFW values with the Voronoi lung phantom were both 0.48 cm. Our results indicate that both the measurements and MC calculations exhibited high reproducibility with plastinated lung sample from human body in previous studies. We found that better results were obtained using the Voronoi lung phantom than using other previous phantoms. The designed phantom may contribute significantly to the improvement of measurement precision. This study suggests that the Voronoi lung phantom is useful for simulating the effects of the heterogeneous structure of lungs on proton beam deterioration.
引用
收藏
页码:86 / 94
页数:9
相关论文
共 19 条
[1]   IMAGING PLATE ILLUMINATES MANY FIELDS [J].
AMEMIYA, Y ;
MIYAHARA, J .
NATURE, 1988, 336 (6194) :89-90
[2]  
[Anonymous], ACTA ANAESTHESIOL SC
[3]  
Cignoni P., 2008, 6 EUR IT C SAL IT JU, DOI [DOI 10.2312/LOCALCHAPTEREVENTS/ITALCHAP/ITALIANCHAPCONF2008/129-136, 10.2312/LocalChapterEvents/ItalChap/ItalianChapConf2008/129-136]
[4]   Efficient and Flexible Sampling with Blue Noise Properties of Triangular Meshes [J].
Corsini, Massimiliano ;
Cignoni, Paolo ;
Scopigno, Roberto .
IEEE TRANSACTIONS ON VISUALIZATION AND COMPUTER GRAPHICS, 2012, 18 (06) :914-924
[5]   Centroidal Voronoi tessellations: Applications and algorithms [J].
Du, Q ;
Faber, V ;
Gunzburger, M .
SIAM REVIEW, 1999, 41 (04) :637-676
[6]   Uncertainties in planned dose due to the limited voxel size of the planning CT when treating lung tumors with proton therapy [J].
Espana, Samuel ;
Paganetti, Harald .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (13) :3843-3856
[7]   Deep inspiration breath-hold technique for lung tumors: The potential value of target immobilization and reduced lung density in dose escalation [J].
Hanley, J ;
Debois, MM ;
Mah, D ;
Mageras, GS ;
Raben, A ;
Rosenzweig, K ;
Mychalczak, B ;
Schwartz, LH ;
Gloeggler, PJ ;
Lutz, W ;
Ling, CC ;
Leibel, SA ;
Fuks, Z ;
Kutcher, GJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (03) :603-611
[8]   A pencil beam algorithm for proton dose calculations [J].
Hong, L ;
Goitein, M ;
Bucciolini, M ;
Comiskey, R ;
Gottschalk, B ;
Rosenthal, S ;
Serago, C ;
Urie, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (08) :1305-1330
[9]   An anatomically realistic lung model for Monte Carlo-based dose calculations [J].
Liang, Liang ;
Larsen, Edward W. ;
Chetty, Indrin J. .
MEDICAL PHYSICS, 2007, 34 (03) :1013-1025
[10]   Small-spot intensity-modulated proton therapy and volumetric-modulated arc therapies for patients with locally advanced non-small-cell lung cancer: A dosimetric comparative study [J].
Liu, Chenbin ;
Sio, Terence T. ;
Deng, Wei ;
Shan, Jie ;
Daniels, Thomas B. ;
Rule, William G. ;
Lara, Pedro R. ;
Korte, Shawn M. ;
Shen, Jiajian ;
Ding, Xiaoning ;
Schild, Steven E. ;
Bues, Martin ;
Liu, Wei .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (06) :140-148