Figulla ASD Occluder versus Amplatzer Septal Occluder: A Comparative Study on Validation of a Novel Device for Percutaneous Closure of Atrial Septal Defects

被引:45
作者
Pac, Aysenur [1 ]
Polat, Tugcin Bora [1 ]
Cetin, Ilker [1 ]
Oflaz, Mehmet Burhan [1 ]
Balli, Sevket [1 ]
机构
[1] TurkiyeYuksek Ihtisas Educ & Res Hosp, Clin Pediat Cardiol, TR-06100 Ankara, Turkey
关键词
TRANSCATHETER CLOSURE; NONOPERATIVE CLOSURE; OCCLUSION; COMPLICATIONS; EXPERIENCE;
D O I
10.1111/j.1540-8183.2009.00497.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Interventions: Between December 2005 and February 2009, 75 patients diagnosed with secundum atrial septal defects underwent transcatheter closure. The FSO device was used in 33 patients, and the ASO was used in 42. Results: Patient characteristics, stretch size of the defect, device left disc size, procedure, and fluoroscopy time were similar between the groups. However, the difference between device waist size and stretched diameter of the defect was significantly higher, and device delivery sheath was significantly larger in FSO group and device left disc size was significantly lower in the FSO group. In all subjects, the residual shunt was small to trivial during follow-up and the reduction in prevalence of residual shunt with time was similar in both groups (P = 0.68). We found no differences in complication rate between the two devices; however, device embolization to the pulmonary bifurcation in one patient was recorded as major complication in FSO device group. Conclusions: Both devices are clinically safe and effective in ASD closure. FSO device has similar outcomes when compared to ASO device. Difficulties in selecting the correct device size in larger defects and larger venous sheath requirement need to be evaluated in further studies. (J Interven Cardiol 2009;22:489-495).
引用
收藏
页码:489 / 495
页数:7
相关论文
共 50 条
[41]   Tips and tricks to prevent prolapse of the amplatzer septal occluder through Large atrial septal defects [J].
Wahab, Hussien Abdul ;
Almossawy, Ali ;
Al Bitar, Ibn ;
Hijazi, Ziyad M. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2011, 78 (07) :1041-1044
[42]   Failed closure of a ventricular septal defect with an Amplatzer occluder [J].
Wippermann, Jens ;
Hoppe, Uta C. ;
Sreeram, Narayanswami ;
Wahlers, Thorsten .
INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2010, 10 (02) :320-321
[43]   Transcatheter closure of defects within the oval fossa using the Amplatzer® Septal Occluder [J].
Latiff, HA ;
Alwi, M ;
Samion, H ;
Kandhavel, G .
CARDIOLOGY IN THE YOUNG, 2002, 12 (03) :224-228
[44]   First experience with the new GORE® septal occluder for the closure of multiple atrial septal defects [J].
Freixa, Xavier ;
Garceau, Patrick ;
Asgar, Anita W. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 81 (07) :1238-1242
[45]   Percutaneous Closure of Patent Foramen Ovale and Secundum Atrial Septal Defects with the GORE® CARDIOFORM Septal Occluder: Incidence and Implications of Device Wire Frame Fracture [J].
Castaldi, Biagio ;
Cabrelle, Giulio ;
Padalino, Massimo ;
Vida, Vladimiro ;
Milanesi, Ornella ;
Di Salvo, Giovanni .
CONGENITAL HEART DISEASE, 2020, 15 (05) :347-360
[46]   Two cases of late embolization of Amplatzer septal occluder devices to the pulmonary artery following closure of secundum atrial septal defects [J].
Mashman, WE ;
King, SB ;
Jacobs, WC ;
Ballard, WL .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2005, 65 (04) :588-592
[47]   Coronary Sinus Defect Following Transcatheter Closure of ASD Using Amplatzer Septal Occluder: Potential Erosion by the Device [J].
Nijres, Bassel Mohammad ;
AL-Kubaisi, Maytham ;
Bokowski, John ;
Abdulla, Ra-id ;
Awad, Sawsan .
PEDIATRIC CARDIOLOGY, 2017, 38 (05) :1084-1086
[48]   Use of the GORE® CARDIOFORM Septal Occluder for percutaneous closure of secundum atrial septal defects: Results of the multicenter US IDE trial [J].
Gillespie, Matthew J. ;
Javois, Alexander J. ;
Moore, Phillip ;
Forbes, Thomas ;
Paolillo, Joseph A. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2020, 95 (07) :1296-1304
[49]   Erosion of an Amplatzer Septal Occluder Device Into the Aortic Root [J].
Kamouh, Abdallah ;
Osman, Mohammed Najeeb ;
Rosenthal, Noah ;
Blitz, Arie .
ANNALS OF THORACIC SURGERY, 2011, 91 (05) :1608-1610
[50]   Impact of the Amplatzer Atrial Septal Occluder Device on Left Ventricular Function in Pediatric Patients [J].
Castaldi, Biagio ;
Santoro, Giuseppe ;
Di Salvo, Giovanni ;
Gaio, Gianpiero ;
Palladino, Maria Teresa ;
D'Aiello, Fabio ;
Iacono, Carola ;
Pacileo, Giuseppe ;
Calabro, Raffaele ;
Russo, Maria Giovanna .
PEDIATRIC CARDIOLOGY, 2013, 34 (07) :1645-1651