Broadband, high dynamics and high resolution charge coupled device-based spectrometer in dynamic mode for multi-keV repetitive x-ray sources

被引:44
作者
Fourment, C. [1 ]
Arazam, N. [1 ]
Bonte, C. [1 ]
Caillaud, T. [2 ]
Descamps, D. [1 ]
Dorchies, F. [1 ]
Harmand, M. [1 ]
Hulin, S. [1 ]
Petit, S. [1 ]
Santos, J. J. [1 ]
机构
[1] Univ Bordeaux, CNRS, CEA, CELIA, F-33405 Talence, France
[2] CEA, CESTA, F-33114 Le Barp, France
关键词
LASER-SOLID INTERACTIONS; CCD; DIFFUSION; DETECTORS; CALIBRATION;
D O I
10.1063/1.3189004
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We present a new operating mode, using a charged coupled device as dispersionless spectrometer dedicated to repetitive x-ray sources in the multi-keV domain. This enables to get spectra with high statistics in a short acquisition time and a way compatible with the operation of other diagnostics requiring accumulation. Several reconstruction algorithms for the spreading events are discussed, and a near Fano-limited resolution is demonstrated by using single pixel events. In this case, a method to take into account partial canceling of the events is presented. Experimental characterization and detailed modeling of the detector are performed, which allow to determine absolute number of photon with +/- 35% accuracy. Characterization of the 5-25 keV x rays emitted by a short pulse laser-produced plasma is reported, as well as their dependency with the atomic number, the laser duration, and energy. (C) 2009 American Institute of Physics. [DOI: 10.1063/1.3189004]
引用
收藏
页数:10
相关论文
共 32 条
[1]   A novel method to estimate the thickness of the depletion layer of an X-ray CCD [J].
Awaki, H ;
Tachibana, K ;
Tamai, Y ;
Yamamoto, K ;
Kitamoto, S ;
Tsujimoto, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 495 (03) :232-239
[2]  
Beg FN, 1997, PHYS PLASMAS, V4, P447, DOI 10.1063/1.872103
[3]   Synchrotron calibration and response modelling of back-illuminated XMM-RGS CCDs [J].
Bootsma, TMV ;
van Zwet, EJ ;
Brinkman, AC ;
den Herder, JW ;
de Jong, L ;
de Korte, P ;
Olsthoorn, SM .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 439 (2-3) :575-581
[4]   Study of x-ray emission enhancement via a high-contrast femtosecond laser interacting with a solid foil [J].
Chen, L. M. ;
Kando, M. ;
Xu, M. H. ;
Li, Y. T. ;
Koga, J. ;
Chen, M. ;
Xu, H. ;
Yuan, X. H. ;
Dong, Q. L. ;
Sheng, Z. M. ;
Bulanov, S. V. ;
Kato, Y. ;
Zhang, J. ;
Tajima, T. .
PHYSICAL REVIEW LETTERS, 2008, 100 (04)
[5]   Hot electron generation via vacuum heating process in femtosecond laser-solid interactions [J].
Chen, LM ;
Zhang, J ;
Dong, QL ;
Teng, H ;
Liang, TJ ;
Zhao, LZ ;
Wei, ZY .
PHYSICS OF PLASMAS, 2001, 8 (06) :2925-2929
[6]   On-board event processing algorithms for a CCD-based space borne X-ray spectrometer [J].
Chun, HJ ;
Bowles, JA ;
BranduardiRaymont, G ;
Gowen, RA .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 376 (02) :254-262
[7]   Observation of subpicosecond x-ray emission from laser-cluster interaction [J].
Dorchies, F. ;
Blasco, F. ;
Bonte, C. ;
Caillaud, T. ;
Fourment, C. ;
Peyrusse, O. .
PHYSICAL REVIEW LETTERS, 2008, 100 (20)
[8]   High-power 1 kHz laser-plasma x-ray source for ultrafast x-ray absorption near-edge spectroscopy in the keV range [J].
Dorchies, F. ;
Harmand, M. ;
Descamps, D. ;
Fourment, C. ;
Hulin, S. ;
Petit, S. ;
Peyrusse, O. ;
Santos, J. J. .
APPLIED PHYSICS LETTERS, 2008, 93 (12)
[9]   THEORY OF HOT-ELECTRON SPECTRA AT HIGH LASER INTENSITY [J].
FORSLUND, DW ;
KINDEL, JM ;
LEE, K .
PHYSICAL REVIEW LETTERS, 1977, 39 (05) :284-288
[10]  
Fraser GW., 1989, XRAY DETECTORS ASTRO