Study of an Additional Layer of Cement Mantle Hip Joints for Reducing Cracks

被引:5
作者
Jamari, J. [1 ]
Saputra, Eko [1 ,2 ]
Anwar, Iwan Budiwan [2 ,3 ]
van der Heide, Emile [2 ]
机构
[1] Diponegoro Univ, Engn Fac, Dept Mech Engn, Jl Prof Soedarto SH Tembalang, Semarang 50275, Indonesia
[2] Twente Univ, Engn Technol Fac, Lab Surface Technol & Tribol, Postbox 217, NL-7500 AE Enschede, Netherlands
[3] Prof Dr R Soeharso Orthopaed Hosp, Orthopaed & Traumatol Dept, Jl A Yani Pabelan, Surakarta 57162, Indonesia
关键词
additional layer; liner; cement mantle; hip joint; THICKNESS; MOVEMENT; MODEL;
D O I
10.3390/jfb10030040
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Failure of the cement mantle in total hip arthroplasty is not a simple phenomenon. Cracking, which can be caused by crack initiation and repeated loading, can cause loosening of the acetabular liner component. A previous study showed that addition of a metal layer between the liner and acetabular could reduce the stress at the contact surface of the cement mantle. This study elaborates on the performance of the additional layer. Several material properties of the layer were simulated using finite element analysis for maximum performance. A static contact analysis was used to simulate the stresses at the contact surface of the cement mantle. The results show that an additional layer of cobalt chrome produced the best performance.
引用
收藏
页数:8
相关论文
共 28 条
[1]   Finite element analysis of interfacial crack behaviour in cemented total hip arthroplasty [J].
Achour, T. ;
Tabeti, M. S. H. ;
Bouziane, M. M. ;
Benbarek, S. ;
Bouiadjra, B. Bachir ;
Mankour, A. .
COMPUTATIONAL MATERIALS SCIENCE, 2010, 47 (03) :672-677
[2]   Subject-specific finite element model of the pelvis: Development, validation and sensitivity studies [J].
Anderson, AE ;
Peters, CL ;
Tuttle, BD ;
Weiss, JA .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (03) :364-373
[3]  
[Anonymous], 2012, ABAQUS 6 12 DOC
[4]   Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[5]  
Eichmiller F. C., 2001, SOC BIOMATERIALS, P472
[6]   Cement-mantle thickness affects cement strains in total hip replacement [J].
Fisher, DA ;
Tsang, AC ;
Paydar, N ;
Milionis, S ;
Turner, CH .
JOURNAL OF BIOMECHANICS, 1997, 30 (11-12) :1173-1177
[7]  
Gunn Elizabeth, 2012, Biomatter, V2, P87, DOI 10.4161/biom.20709
[8]   Fatigue crack growth rate does not depend on mantle thickness: an idealized cemented stem construct under torsional loading [J].
Hertzler, J ;
Miller, MA ;
Mann, KA .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2002, 20 (04) :676-682
[9]  
Jacobs C.R., FATIGUE LIFETIME STU
[10]  
Jamari J, 2018, INT J ENG TECHNOL IN, V8, P99