A Comparison of Experimental Aneurysm Occlusion Determination by Angiography, Scanning Electron Microscopy, MICROFIL® Perfusion, and Histology

被引:4
作者
Cruise, Gregory M. [1 ]
Rivera, Elias A. [2 ]
Jones, Russell M. [3 ]
Vestal, Maria [1 ]
Virmani, Renu [3 ]
Ding, Yong-Hong [4 ]
Kallmes, David F. [4 ]
机构
[1] MicroVent Terumo, Aliso Viejo, CA 92656 USA
[2] Tengion Inc, Winston Salem, NC 27103 USA
[3] CV Path, Gaithersburg, MD 20878 USA
[4] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
关键词
embolization; animal model; histology; hydrogel; platinum coils; GUGLIELMI DETACHABLE COILS; INTRACRANIAL ANEURYSMS; PLATINUM COILS; ENDOVASCULAR TREATMENT; CEREBRAL ANEURYSMS; EMBOLIZATION; HYDROCOIL; PACKING; CREATION; SYSTEM;
D O I
10.1002/jbm.b.31443
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In clinical practice, occlusion of embolized, intracerebral aneurysms is evaluated using angiography. Standard, two-dimensional digital subtract angiography (DSA) is unable to quantify irregular aneurysm remnants, and even three-dimensional rotational angiography cannot quantify the degree of occlusion. To better understand occlusion at the aneurysm neck, we compared angiographic results with MICROFILO (R) perfusion, histology, and scanning electron microscopy (SEM) results in 20 elastase-induced saccular aneurysms in rabbits. Aneurysms were embolized with HydroCoil devices (n = 12) or platinum coils (n = 8). Aneurysm follow-up occurred at 2 (n = 10) and 6 (n = 10) weeks. Aneurysm occlusion was evaluated using DSA, MICROFIL perfusion, histological ground sections, and SEM. Groups were compared statistically using ANOVA and chi(2) tests. The MICROFIL perfusion results were not concordant with the angiographic results for the HydroCoil and platinum coil groups. Both increased and decreased occlusion was observed on the MICROFIL-perfused aneurysms when compared with angiography. The histologic occlusion results of the HydroCoil group were concordant with the angiographic results; however, unoccluded areas not visible on angiography were routinely observed on the ground sections in the platinum coil group. SEM imaging of the aneurysm neck consistently showed decreased occlusion than angiographic results for both the HydroCoil and platinum coil groups. Although histology and MICROFIL-perfusion analyses provided additional details of aneurysm occlusion when compared with angiography, complete visualization of the entire neck of the aneurysm and accurate assessment of aneurysm occlusion was possible only with SEM. (C) 2009 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 91B: 669-678, 2009
引用
收藏
页码:669 / 678
页数:10
相关论文
共 25 条
[1]   Creation of saccular aneurysms in the rabbit: A model suitable for testing endovascular devices [J].
Altes, TA ;
Cloft, HJ ;
Short, JG ;
DeGast, A ;
Do, HM ;
Helm, GA ;
Kallmes, DF .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (02) :349-354
[2]   Gross and microscopic histopathological findings in aneurysms of the human brain treated with Guglielmi detachable coils [J].
Bavinzski, G ;
Talazoglu, V ;
Killer, M ;
Richling, B ;
Gruber, A ;
Gross, CE ;
Plenk, H .
JOURNAL OF NEUROSURGERY, 1999, 91 (02) :284-293
[3]  
Cloft HJ, 2004, AM J NEURORADIOL, V25, P60
[4]   Hydrogel-coated and platinum coils for intracranial aneurysm embolization compared in three experimental models using computerized angiographic and histologic morphometry [J].
Cruise, Gregory M. ;
Shum, Julie C. ;
Plenk, Hanns, Jr. .
JOURNAL OF MATERIALS CHEMISTRY, 2007, 17 (38) :3965-3973
[5]   A prospective single-center analysis of the safety and efficacy of the HydroCoil embolization system for the treatment of intracranial aneurysms [J].
Deshaies, Eric M. ;
Adamo, Matthew A. ;
Boulos, Alan S. .
JOURNAL OF NEUROSURGERY, 2007, 106 (02) :226-233
[6]  
Ding YH, 2005, AM J NEURORADIOL, V26, P1757
[7]   HydroCoil as an adjuvant to bare platinum coil treatment of 100 cerebral aneurysms [J].
Fanning, Noel F. ;
Berentei, Zsolt ;
Brennan, Paul R. ;
Thornton, John .
NEURORADIOLOGY, 2007, 49 (02) :139-148
[8]   Factors affecting outcome after endovascular treatment of intracranial aneurysms [J].
Fiehler, Jens ;
Byrne, James V. .
CURRENT OPINION IN NEUROLOGY, 2009, 22 (01) :103-108
[9]  
Fujiwara NH, 2001, AM J NEURORADIOL, V22, P698
[10]   Embolization of intracranial aneurysms with hydrogel-coated coils versus inert platinum coils - Effects on packing density, coil length and quantity, procedure performance, cost, length of hospital stay, and durability of therapy [J].
Gaba, RC ;
Ansari, SA ;
Roy, SS ;
Marden, FA ;
Viana, MAG ;
Malisch, TW .
STROKE, 2006, 37 (06) :1443-1450