Cadmium determination in biological samples by direct solid sampling flame atomic absorption spectrometry

被引:0
作者
Flores, TMD [1 ]
Paniz, JNG
Martins, AF
Dressler, VL
Müller, EI
da Costa, AB
机构
[1] Univ Fed Santa Maria, Dept Quim, BR-97105900 Santa Maria, RS, Brazil
[2] Univ Santa Cruz Sul, Dept Biol, BR-96815900 Santa Cruz Do Sul, RS, Brazil
关键词
direct solid analysis; flame atomic absorption spectrometry; cadmium determination; biological samples;
D O I
暂无
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A direct solid sampling flame atomic absorption spectrometric procedure for trace determination of cadmium in biological samples has been developed. Test samples (0.05-2.00 mg) were ground and weighed into small polyethylene vials, which were connected to the device for solid sample introduction into a conventional air/acetylene flame. Test samples were carried as a dry aerosol to a quartz cell, placed between the burner and the optical path, which had a perpendicular entrance and a slit in the upper part. The atomic vapor generated in the flame produced a transient signal that was totally integrated within 1 s. The effect of operating conditions and the extent of grinding on the analytical signal were evaluated. Background signals were always low and a characteristic mass of 0.29 ng Cd was obtained. Calibration was performed using different masses of solid certified reference materials. Results obtained for certified and in-house reference materials were typically within the 95% confidence interval of the certified and/or reference value, and the precision, expressed as relative standard deviation, was between 3.9 and 6.7%. The proposed system is simple and it might be adapted to conventional atomic absorption spectrometers allowing the determination of Cd in more than 80 test samples per hour, excluding weighing. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:2187 / 2193
页数:7
相关论文
共 28 条
[1]   Determination of lead by atomic absorption spectrometry using tube-in-flame atomisation and solid sampling [J].
Alvarado, J ;
Jaffe, R .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (11) :1297-1300
[2]   Direct analysis of Antarctic krill by slurry sampling:: Determination of copper, iron, manganese and zinc by flame atomic absorption spectrometry [J].
Alves, FL ;
Smichowski, P ;
Farías, S ;
Marrero, J ;
Arruda, MAZ .
JOURNAL OF THE BRAZILIAN CHEMICAL SOCIETY, 2000, 11 (04) :365-370
[4]  
Campos R.C., 1990, J. Braz. Chem. Soc, V1, P66, DOI [10.5935/0103-5053.19900014, DOI 10.5935/0103-5053.19900014]
[5]   COMBUSTION AND VOLATILIZATION OF SOLID SAMPLES FOR DIRECT ATOMIC-ABSORPTION SPECTROMETRY USING SILICA OR NICKEL TUBE FURNACE ATOMIZERS [J].
CAMPOS, RC ;
CURTIUS, AJ ;
BERNDT, H .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1990, 5 (08) :669-673
[6]   SIMPLE MATRIX MODIFICATION PROCEDURE TO ALLOW DIRECT DETERMINATION OF CADMIUM IN BLOOD BY FLAME MICRO-SAMPLING ATOMIC-ABSORPTION SPECTROPHOTOMETRY [J].
DELVES, HT .
ANALYST, 1977, 102 (1214) :403-405
[8]  
Flores ÉMD, 2001, SPECTROCHIM ACTA B, V56, P1875
[9]   Direct analysis of tantalum powders by electrothermal vaporization inductively coupled plasma atomic emission spectrometry [J].
Friese, KC ;
Krivan, V .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1999, 364 (1-2) :72-78
[10]   DETERMINATION OF TRACE-ELEMENTS IN TITANIUM(IV) OXIDE PIGMENTS BY ATOMIC-ABSORPTION SPECTROMETRY USING AN AQUEOUS SLURRY TECHNIQUE [J].
FULLER, CW .
ANALYST, 1976, 101 (1209) :961-965