Development of CO2 laser interferometer/polarimeter system in JT-60SA

被引:11
作者
Sasao, H. [1 ]
Kawano, Y. [2 ]
Arakawa, H. [1 ]
Sakuma, T. [1 ]
Imazawa, R. [1 ]
Kubo, H. [1 ]
Itami, K. [1 ]
机构
[1] Japan Atom Energy Agcy, 801-1 Mukoyama, Naka, Ibaraki 3110193, Japan
[2] Teikyo Univ, Fac Fukuoka Med Technol, 6-22 Misakimachi, Omuta, Fukuoka 8368505, Japan
来源
JOURNAL OF INSTRUMENTATION | 2016年 / 11卷
关键词
Detector alignment and calibration methods (lasers; sources; particle-beams); Plasma diagnostics - interferometry; spectroscopy and imaging; ELECTRON-DENSITY MEASUREMENT; TOKAMAK; LASER; DIAGNOSTICS;
D O I
10.1088/1748-0221/11/02/C02082
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Development of the CO2 interferometer/polarimeter system for line averaged electron density measurement in JT-60SA was progressed significantly. For the measurement, two viewing lines were selected, i.e. a tangential line and a vertical line. Two beam convertors were designed to optimize the beam size for long-distance transmission up to similar to 240 m. A CO2 laser spectral beam axis/profile monitoring system and a beam axis stabilizing system were newly developed. By use of blue light for fluorescence of targets of the monitoring system, instead of conventionally used ultra-violet light, monitoring stability was significantly improved. Practical characteristics of a high power RF discharge laser for JT-60SA such as wavelength stability and beam quality were investigated by the monitoring system. Beam axis deviation was controlled by the stabilizing system within 1% of the beam diameter. By a performance test of the RF discharge laser using a single-color table-top interferometer, a good linearity of phase difference is confirmed in whole of the fringe range by a plasma (similar to 6 fringes) in JT-60SA. Obtained resolution of the phase difference (statistical error/5 mu s) is within 10 degrees, which is promising to satisfy the target resolution of 1% for the line averaged electron density measurement.
引用
收藏
页数:12
相关论文
共 15 条
[1]  
[Anonymous], 2015, JT60SA RES PLAN 3 2, P9
[2]  
Arakawa H., COMMUNICATION
[3]   JET POLARI-INTERFEROMETER [J].
BRAITHWAITE, G ;
GOTTARDI, N ;
MAGYAR, G ;
OROURKE, J ;
RYAN, J ;
VERON, D .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (09) :2825-2834
[4]   REAL-TIME, VIBRATION-COMPENSATED CO2 INTERFEROMETER OPERATION ON THE DIII-D TOKAMAK [J].
CARLSTROM, TN ;
AHLGREN, DR ;
CROSBIE, J .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (07) :1063-1066
[5]   FREQUENCY-STABILIZED SINGLE-MODE CW 118.8-MU-M CH3OH WAVE-GUIDE LASER FOR LARGE TOKAMAK DIAGNOSTICS [J].
FUKUDA, T ;
NAGASHIMA, A .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (06) :1080-1085
[6]   AN INFRARED VIBRATION COMPENSATED INTERFEROMETER FOR THE MST EXPERIMENT [J].
INNOCENTE, P ;
MARTINI, S ;
ROCA, CF .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (10) :4999-5001
[7]  
International standard, 2005, ISO1111461
[8]  
Kawano K., 1994, Review of Laser Engineering, V22, P226, DOI 10.2184/lsj.22.226
[9]   Application of diamond window for infrared laser diagnostics in a tokamak device [J].
Kawano, Y ;
Chiba, S ;
Inoue, A .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (01) :279-280
[10]   Infrared laser polarimetry for electron density measurement in tokamak plasmas [J].
Kawano, Y ;
Chiba, S ;
Inoue, A .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (01) :1068-1072